Vi self study exam preparatory note part 4 section 2

Page 1

ASNT Level III- Visual & Optical Testing My Pre-exam Preparatory Self Study Notes Reading 4 Section 2 2014-August

Charlie Chong/ Fion Zhang


For my coming ASNT Level III VT Examination 2014-August

Charlie Chong/ Fion Zhang


Reading 4 ASNT Nondestructive Handbook Volume 8 Visual & Optical testing- Section 2 For my coming ASNT Level III VT Examination 2014-August

Charlie Chong/ Fion Zhang


Fion Zhang 2014/August/15

Charlie Chong/ Fion Zhang


SECTION 2 THE PHYSICS OF LIGHT

Charlie Chong/ Fion Zhang


SECTION 2: The Physics of Light PART 1: THE PHYSICS OF LIGHT 1.1 1.2 1.3 1.4 1.5 1.6

Radiant Energy Theories Light and the Energy Spectrum Blackbody Radiation Atomic Structure and Radiation Luminous Efficiency of Radiant Energy Luminous Efficiency of Light Sources

Charlie Chong/ Fion Zhang


SECTION 2: The Physics of Light PART 2: Measurement of the properties of light 2.1 2.2 2.3 2.4 2.5 2.6 2.7 2.8 2.9 2.10 2.11 2.12 2.13 2.14

Photovoltaic Cells Photoconductor Cells Photoelectric Tubes Photodiodes and Phototransistors Photometry Principles of Photometry Photometers Photovoltaic Cell Meters Meters Using Photomultiplier Tubes Equivalent Sphere Illumination Photometers Reflectometers Radiometers Spectrophotometers Types of Photometers

Charlie Chong/ Fion Zhang


PART 1: THE PHYSICS OF LIGHT 1.0

General:

Light can be defined as radiant energy capable of exciting the human retina and creating a visual sensation. From the viewpoint of physics, light is defined as that portion of the electromagnetic spectrum with wavelengths between 380 and 770 nm. Visually, there is some variation in these limits among individuals. Radiant energy at the proper wavelength makes visible anything from which it is emitted or reflected in sufficient quantity to activate the receptors in the eye. The quantity of such radiant energy may be evaluated in many ways, including: radiant flux (measured in joules per second or in watts) and luminous flux (measured in lumens).

Charlie Chong/ Fion Zhang


1.1

Radiant Energy Theories

Several theories describing radiant energy have been proposed and the text below briefly discusses the primary theories. Corpuscular Theory The corpuscular theory was advocated by Sir Isaac Newton and is based on the following premises. 1. Luminous bodies emit radiant energy in particles. 2. These particles are intermittently ejected in straight lines. 3. The particles act on the retina of the eye, stimulating the optic nerves to produce the sensation of light.

Charlie Chong/ Fion Zhang


Wave Theory The wave theory of radiant energy was advocated by Christian Huygens and is based on these premises. 1. Light results from the molecular vibration in luminous material. 2. The vibrations are transmitted through the ether in wavelike movements (comparable to ripples in water). 3. The vibrations act on the retina of the eye, stimulating the optic nerves to produce visual sensation. The velocity of a wave is the product of its wavelength and its frequency.

Charlie Chong/ Fion Zhang


Electromagnetic Theory The electromagnetic theory was advanced by James Clerk Maxwell and is based on these premises. 1. Luminous bodies emit light in the form of radiant energy. 2. This radiant energy is propagated in the form of electromagnetic waves. 3. The electromagnetic waves act on the retina of the eye, stimulating the optic nerves to produce the sensation of light.

Charlie Chong/ Fion Zhang


Quantum Theory The quantum theory is an updated version of the corpuscular theory. It was advanced by Planck and is based on these premises. 1. Energy is emitted and absorbed in discrete quanta (photons). 2. The energy in each quantum is hv. The term h is known as Planck's constant and is equal to 6.626 x 10-34 joule•second. The term v is the frequency in hertz.

Charlie Chong/ Fion Zhang


Unified Theory The unified theory of radiant energy was proposed by De Broglie and Heisenberg and is based on the premise that every moving element of mass is associated with a wave whose length is given by: 位 = h / mv (Eq. 1) Where: 位 h m v

= wavelength of the wave motion (meters); = Planck's constant or 6.626 x 10-34 Joule.second; = Mass in Kg = velocity of the particle (meters per second).

It is impossible to determine all of the properties that are distinctive of a wave or a particle simultaneously, since the energy to do so changes one of the properties being determined.

Charlie Chong/ Fion Zhang


The quantum theory and the electromagnetic wave theory provide an explanation of radiant energy that is appropriate for the purposes of nondestructive testing. Whether it behaves like a wave or like a particle, light is radiation produced by atomic or molecular processes. That is, in an incandescent body, a gas discharge or a solid state device, light is produced when excited electrons have just reverted to more stable positions in their respective atoms, thereby releasing energy.

Charlie Chong/ Fion Zhang


1.2

Light and the Energy Spectrum

The wave theory permits a convenient representation of radiant energy in an arrangement based on the energy's wavelength or frequency. This arrangement is called a spectrum and is useful for indicating the relationship between various radiant energy wavelength regions. Such a representation should not be taken to mean that each region of the spectrum is physically divided from the others- actually there is a small but discrete transition from one region to the next. The general limits of the radiant energy spectrum extend over a range of wavelengths varying from 10-16 to over 105 m. Radiant energy in the visible spectrum has wavelengths between 380 x 10-9 and 770 x 10-9 m. In the SI system, the nanometer nm (10-9 m) and the micrometer Îźm (10-6 m) are the commonly used units of wavelength in the visible region. In the cgs system, the angstrom A (10-10 m) was used to denote wavelength.

Charlie Chong/ Fion Zhang


All forms of radiant energy are transmitted at the same speed in a vacuum: 299,793 km•s -1 (186,282 mil•s-1). However, each form of energy differs in wavelength and therefore in frequency. The wavelength and velocity may be altered by the medium through which it passes but the frequency is fixed independently of the medium. Equation 2 shows the relationship between radiation velocity, frequency and wavelength.

v = λ v/ n

(Eq.2)

Where: v = velocity of waves in the medium (meters per second); n = the medium's index of refraction; λ = wavelength in a vacuum (meters); and v = frequency (hertz).

Charlie Chong/ Fion Zhang


Keywords:  The velocity of light change in different medium with different refractive index  The wavelength and velocity may be altered by the medium through which it passes but the frequency is fixed independently of the medium. When light travelling in different medium  Velocity change  Wavelength change  Frequency always remain the same in all mediums

Charlie Chong/ Fion Zhang


Planck and the Quanta In 1900, Max Planck was working on the problem of how the radiation an object emits is related to its temperature. He came up with a formula that agreed very closely with experimental data, but the formula only made sense if he assumed that the energy of a vibrating molecule was quantized--that is, it could only take on certain values. The energy would have to be proportional to the frequency of vibration, and it seemed to come in little "chunks" of the frequency multiplied by a certain constant. This constant came to be known as Planck's constant, or h, and it has the value 6.626x10-34 J x s http://web2.uwindsor.ca/courses/physics/high_schools/2005/Photoelectric_effect/planck.html

Charlie Chong/ Fion Zhang


Table 1 gives the speed of light in different media for a frequency corresponding to a wavelength of 589 nm in air. TABLE 1. Speed of light for a wavelength of 589 nanometers (sodium D-lines1

Medium

Speed 106 meters per second

Vacuum Air 100 kilopascals [760 mm Hg] at 0째 C Crown glass Water

299.793 299.724

Charlie Chong/ Fion Zhang

198.223 224.915


1.3

Blackbody Radiation

The light from common sources, particularly light from incandescent lamps, is often compared with light from a theoretical source known as a blackbody. For equal area, a blackbody radiates more total power and more power at any wavelength than any other source operating at the same temperature. For experimental purposes, laboratory radiation sources have been built to approximate closely a blackbody. Designs of these sources are based on the fact that a hole in the wall of a closed chamber, small in size compared with the size of the enclosure, exhibits blackbody characteristics. This can be understood with the help of Fig. 1. At each reflection, some energy is absorbed. In time, all incoming energy is absorbed by the walls. Conversely, a blackbody can be a perfect radiator. If the interior walls of the blackbody are uniformly heated, the radiation which leaves the small opening will he that of a perfect radiator for a specific temperature and its emission energy and wavelength spectrum will be independent of the nature of the enclosure. From 1948 to 1979, the international reference standard for the unit of luminous intensity was taken to be the luminance of a blackbody operating at the temperature of freezing platinum. Charlie Chong/ Fion Zhang


FIGURE 1. Small aperture in an enclosure exhibits blackbody characteristics

Charlie Chong/ Fion Zhang


Black-body radiation is the type of electromagnetic radiation within or surrounding a body in thermodynamic equilibrium with its environment, or emitted by a black body (an opaque and non-reflective body) held at constant, uniform temperature. The radiation has a specific spectrum and intensity that depends only on the temperature of the body. The thermal radiation spontaneously emitted by many ordinary objects can be approximated as blackbody radiation. A perfectly insulated enclosure that is in thermal equilibrium internally contains black-body radiation and will emit it through a hole made in its wall, provided the hole is small enough to have negligible effect upon the equilibrium. A black-body at room temperature appears black, as most of the energy it radiates is infra-red and cannot be perceived by the human eye. Because of the specific colour responsiveness of the human eye, a black body, viewed in the dark at the lowest just faintly visible temperature, subjectively appears grey, even though its objective physical spectrum peaks in the red range. When it becomes a little hotter, it appears dull red. As its temperature increases further it eventually becomes blindingly brilliant blue-white.

Charlie Chong/ Fion Zhang


Although planets and stars are neither in thermal equilibrium with their surroundings nor perfect black bodies, black-body radiation is used as a first approximation for the energy they emit. Black holes are near-perfect black bodies, in the sense that they absorb all the radiation that falls on them. It has been proposed that they emit black-body radiation (called Hawking radiation), with a temperature that depends on the mass of the black hole. The term black body was introduced by Gustav Kirchhoff in 1860. When used as a compound adjective, the term is typically written as hyphenated, for example, black-body radiation, but sometimes also as one word, as in blackbody radiation. Black-body radiation is also called complete radiation or temperature radiation or thermal radiation. http://en.wikipedia.org/wiki/Black-body_radiation

Charlie Chong/ Fion Zhang


1.3.1

Planck Radiation Law

Data describing blackbody radiation curves have been obtained using a specially constructed and uniformly heated tube as the source. Planck, introducing the concept of discrete quanta of energy, developed an equation depicting these curves. It gives the spectral radiance of a blackbody as a function of the wavelength and temperature. Figure 2 shows the spectral radiance of a blackbody as a function of wavelength for several values of absolute temperature, plotted on a logarithmic scale.

Charlie Chong/ Fion Zhang


FIGURE 2. Blackbody radiation curves showing Wien displacement of peaks for operating temperatures between 500 and 20,000 K

Charlie Chong/ Fion Zhang


FIGURE 2. Blackbody radiation curves

Charlie Chong/ Fion Zhang


FIGURE 2. Blackbody radiation curves

Charlie Chong/ Fion Zhang


FIGURE 2. Blackbody radiation curves- Peak Shifts

Charlie Chong/ Fion Zhang


FIGURE 2. Blackbody radiation curves- Intensity & Peak Shifts

Charlie Chong/ Fion Zhang


The color (chromaticity) of black-body radiation depends on the temperature of the black body; the locus of such colors, shown here in CIE 1931 x,y space, is known as the Planckian locus. http://en.wikipedia.org/wiki/Black-body_radiation

Charlie Chong/ Fion Zhang


1.3.2

Wien Radiation Law

In the temperature range of incandescent filament lamps (2,000 to 3,400 K) and in the visible wavelength region (380 to 770 nm), a simplification of the Planck equation, known as the Wien radiation law, gives a good representation of the blackbody distribution of spectral radiance. The Wien displacement law gives the relationship between the wavelength at which a blackbody at temperature T in degrees Kelvin emits maximum power per unit wavelength and the temperature T. In fact the product of absolute temperature T and the peak wavelength is a constant. It gives the relationship between blackbody distributions at various temperatures only with this important limitation.

Charlie Chong/ Fion Zhang


Wien Radiation Law

Charlie Chong/ Fion Zhang


1.3.3

Stefan-Boltzmann Law

The Stefan-Boltzmann law is obtained by integrating Planck's expression for the spectral radiant excitance from zero to infinite wavelength. The law states that the total radiant power per unit area of a blackbody varies as the fourth power of the absolute temperature. The Stefan-Boltzmann law is explained in introductory physics texts. Note that this law applies to the total power (the whole spectrum) and cannot be used to estimate the power in the visible portion of the spectrum alone.

Charlie Chong/ Fion Zhang


1.3.4

Spectral Emissivity

No known radiator has the same emissive power as a blackbody. The ratio of a radiator's output at any wavelength to that of a blackbody at the same temperature and the same wavelength is known as the radiator's spectral emissivity 蔚(位). 1.3.5

Graybody Radiation

When the spectral emissivity is uniform for all wavelengths, the radiator is known as a graybody. No known radiator has a uniform spectral emissivity for all visible, infrared and ultraviolet wavelengths. In the visible region, a carbon filament exhibits very nearly uniform emissivity and is nearly a graybody.

Charlie Chong/ Fion Zhang


1.3.6

Selective Radiators

The emissivity of all known materials varies with wavelength. In Fig. 3, the radiation curves for a blackbody, a graybody and a selective radiator (tungsten), all operating at 3,000 K, are plotted on the same logarithmic scale to show the characteristic differences in output.

Charlie Chong/ Fion Zhang


FIGURE 3. Radiation curves for blackbody, graybody and selective radiators operating at 3,000 K

Charlie Chong/ Fion Zhang


FIGURE 3. Radiation curves for blackbody, graybody and selective radiators operating at 3,000 K ~ 6,000 K

http://www.webexhibits.org/causesofcolor/3.html

Charlie Chong/ Fion Zhang


1.3.7

Color Temperature

The radiation characteristics of a blackbody of unknown area may be specified with the aid of the radiation equations by modifying two quantities: the magnitude of the radiation at any given wavelength and the absolute temperature. The same type of specification may be used with reasonable accuracy in the visible region of the spectrum for tungsten filaments and other incandescent sources. However, the temperature used in the case of selective radiators is not that of the filament but a value called the color temperature. The color temperature of a selective radiator is that temperature at which a blackbody must be operated so that its output is the closest approximation to a perfect color match with the output of the selective radiator. While the match is never actually perfect, the small deviations that occur in the case of incandescent filaments are not of practical importance.

Charlie Chong/ Fion Zhang


The apertures between coils of the filaments used in many tungsten lamps act something like a blackbody because of the interreflections that occur at the inner surfaces of the helix formed by the coil. For this reason, the distribution from coiled filaments exhibits a combination of the characteristics of the straight filament and of a blackbody operating at the same temperature. The use of the color temperature method to deduce the spectral distribution from other than incandescent sources, even in the visible region, usually results in appreciable error. Color temperature values associated with light sources other than incandescent are known as correlated color temperatures and are not true color temperatures.

Charlie Chong/ Fion Zhang


1.4

Atomic Structure and Radiation

The atomic theories first proposed in 1913 have been expanded and confirmed by much experimental evidence. The atom consists of a central positively charged nucleus about which revolve negatively charged electrons. In the normal state, these electrons remain in specific orbits or energy levels and radiation is not emitted by the atom. The orbit described by a specific electron revolving about the nucleus is determined by the energy of the electron (there is a particular energy associated with each orbit). An atom's system of orbits or energy levels is characteristic of each element and remains stable until disturbed by external forces. The electrons of an atom can he divided into two classes. The first includes the inner shell electrons which are removed or excited only by high energy radiation. The second class includes the outer shell or valence electrons which cause chemical bonding into molecules. Valence electrons are readily excited by ultraviolet or visible radiation or by electron impact and can be removed completely with relative ease. The valence electrons of an atom in a solid when removed from their associated nucleus enter the so-called conduction band and give the material the property of electrical conductivity. Charlie Chong/ Fion Zhang


After absorption of sufficient energy by an atom, the valence electron is pushed to a higher energy level further from the nucleus. Eventually, the electron returns to the normal orbit or to an intermediate orbit. In so doing, the energy that the atom loses is emitted as a quantum of radiation and this is the source of light. The wavelength (or frequency) of the radiation is determined by Planck's equation: E1 – E2 = hv Where: E1 E2 h v

(Eq. 3)

= energy associated with the excited orbit; = energy associated with the normal orbit; = Planck's constant; and = frequency of the emitted radiation.

Charlie Chong/ Fion Zhang


Plank’s Equation

Charlie Chong/ Fion Zhang


This equation can he converted to a more practical form: 位 = 1239.76/ Vd Where: 位 = wavelength (nanometers); = the potential difference between two energy levels through which Vd the displaced electron has fallen in one transition (electron volts).

Charlie Chong/ Fion Zhang


1.5

Luminous Efficiency of Radiant Energy

Many apparent differences in intensity between radiant energy of various wavelengths are in fact differences in the ability of various sensing devices to detect them. The reception characteristics of the human eye have been subject to extensive investigations and the results may be summarized as follows. 1. The spectral response characteristic of the human eye varies between individuals, with time and with the age and health of an individual, to the extent that the selection of any individual to act as a standard observer is not scientifically feasible. 2. However, from the available data, a luminous efficiency curve has been selected to represent a typical human observer. This curve may be applied mathematically to the solution of photometric problems.

Charlie Chong/ Fion Zhang


The standard spectral luminous efficiency curve for photopic (light adapted) vision represents a typical characteristic, adopted arbitrarily to give unique solutions to photometric problems, from which the characteristics of any individual may be expected to vary. Some data indicate that most human observers are capable of experiencing a visual sensation on exposure to radiant energy of wavelengths longer than 770 nm, called infrared under most circumstances, provided the radiant energy reaches the eye at a sufficiently high rate. It also is known that ultraviolet radiation (wavelengths less than 380 nm) under most circumstances can be seen if it reaches the retina even at a moderate rate.

Charlie Chong/ Fion Zhang


Most observers yield only a slight response to ultraviolet radiation at the nearly visible wavelengths because the lens of the eye absorbs nearly all of it. Typically, human observers have a response that under normal lighting conditions extends from 380 to 770 manometers but some individuals have greater sensitivity at the blue and/or red ends of this range. Of course, at lower lighting levels even the average observer experiences a shift of the visible spectrum to shorter wavelengths and vice versa at higher lighting levels. The spectral range of visible response is therefore not static but greatly dependent on the lighting conditions.

Charlie Chong/ Fion Zhang


1.6

Luminous Efficiency of Light Sources

The luminous efficiency of a light source is defined as the ratio of the total luminous flux (lumens) to the total power input (watts or equivalent). The maximum luminous efficiency of an ideal white source (defined as a radiator with constant output over the visible spectrum and no radiation in other parts of the spectrum) is about 220 lm•W-1

Charlie Chong/ Fion Zhang


PART 2: MEASUREMENT OF THE PROPERTIES OF LIGHT 2.0

General:

The most widely used detector of light is the human eye. Other common, mechanical detectors are photovoltaic cells, photoconductive cells, photoelectric tubes, photodiodes, phototransistors and photographic film.

Charlie Chong/ Fion Zhang


2.1

Photovoltaic Cells

Photovoltaic cells typically include selenium barrier layer cells and silicon or gallium arsenide, photodiodes operated in the photovoltaic or unbiased mode. These devices depend on the generation of a current resulting from the absorption of a photon. The cell is comprised of 1. a p-type material, typically a metal plate coated with a semiconductor, such as selenium on iron; and 2. a semitransparent n-type material such as cadmium oxide. Unless there is an external circuit, electrons liberated from the semiconductor are trapped at the p-n junction after exposure to light. The device thereby converts radiant energy to electric energy, which can be used directly or amplified to drive a micro-ammeter (see Fig. 4). Photovoltaic cells can be filtered to correct their spectral response so that the micro-ammeter can be calibrated in units of illuminance. Factors such as response time, fatigue, temperature effects, linearity stability, noise and magnitude of current influence the choice of cell and circuit for a given application.

Charlie Chong/ Fion Zhang


FIGURE 4. Cross section of a barrier layer photovoltaic cell showing motion of photoelectrons through a micro-ammeter circuit

Charlie Chong/ Fion Zhang


FIGURE 4. Cross section of a barrier layer photovoltaic cell showing motion of photoelectrons through a micro-ammeter circuit

Charlie Chong/ Fion Zhang


FIGURE 4. Cross section of a barrier layer photovoltaic cell showing motion of photoelectrons through a micro-ammeter circuit

Charlie Chong/ Fion Zhang


FIGURE 4. Cross section of a barrier layer photovoltaic cell showing motion of photoelectrons through a micro-ammeter circuit

Charlie Chong/ Fion Zhang


2.2

Photoconductor Cells

Photoconductor cells depend on the resistance of the cell changing directly as a result of photon absorption. These detectors use materials such as cadmium sulfide, cadmium selenide and selenium. Cadmium sulfide and cadmium selenide are available in transparent resin or glass envelopes and are suitable for low illuminance levels less than 10-4 lx (10-5 ftc).

Charlie Chong/ Fion Zhang


Photoconductor Cells

Charlie Chong/ Fion Zhang


2.3

Photoelectric Tubes

The photoelectric effect is the emission of electrons from a surface bombarded by sufficiently energetic photons. If the surface is connected as a cathode in an electric field (see Fig. 5), the liberated electrons flow to the anode, creating a photoelectric current. An arrangement of this sort may be used as an illuminance meter and can be calibrated in lux or footcandles. The photoelectric current in vacuum varies directly with the illuminance level over a wide range (spectral distribution, polarization and cathode potential remain the same). In gas filled tubes, the response is linear only over a limited range. If the radiant energy is polarized, the photoelectric current varies as the orientation of the polarization is changed (except at normal incidence).

Charlie Chong/ Fion Zhang


FIGURE 5. By the photoelectric effect, electrons may be liberated from an Illuminated metal surface, flowing to an anode and creating an electric current that may be detected by a galvanometer (see Eq. 1 and 2)

Charlie Chong/ Fion Zhang


Photoelectric Effects

Charlie Chong/ Fion Zhang


Photoelectric Effects & Secondary radiation

Charlie Chong/ Fion Zhang


Photoelectric Effects

http://whs.wsd.wednet.edu/faculty/busse/mathhomepage/busseclasses/radiationphysics/lecturenotes/chapter12/chapter12.html

Charlie Chong/ Fion Zhang


2.4

Photodiodes and Phototransistors

Photodiodes or junction photocells are based on solid state p-n junctions that react to external stimuli such as light. Conversely, if properly constructed, they can emit radiant energy (light emitting diodes). In a photosensitive diode, the reverse saturation current of the junction increases in proportion to the illuminance. Such a diode can therefore be used as a sensitive detector of light and is particularly suitable for indicating extremely short pulses of radiation because of its very fast response time. Phototransistors operate in a manner similar to photodiodes but, because they provide an additional amplifier effect, they are many times more sensitive than simple photodiodes.

Charlie Chong/ Fion Zhang


Photodiode

Charlie Chong/ Fion Zhang


Schematic cross section of an integrated CMOS single-photon-counting avalanche diode (SPAD) device.2HV: High-voltage. p, n: Semiconductor materials.

http://spie.org/x93517.xml

Charlie Chong/ Fion Zhang


Photodiode

Charlie Chong/ Fion Zhang


Phototransistors

Charlie Chong/ Fion Zhang


Phototransistors

Charlie Chong/ Fion Zhang


2.5

Photometry

Progress in the sciences is often dependent on our ability to measure the physical quantities associated with the technology- each advance in measurement ability or accuracy provides a broadening of the science's knowledge base. The measurement of light and its properties is called photometry. The basic measuring instrument is known as a photometer. The earliest photometers depended on visual appraisal by the operator as the means of measurement and such meters are rarely used now. They have been replaced by nonvisual meters that are sensitive to light's physical properties, measuring radiant energy incident on a receiver, producing measurable electrical quantities. Physical photometers are more accurate and simpler to operate than their earlier counterparts.

Charlie Chong/ Fion Zhang


2.5.1

Observer Response Curves

Light measurements by physical photometers are useful only if they indicate reliably how the eye reacts to a certain stimulus. In other words, the photometer should be sensitive to the spectral power distribution of light in the same way that the eye is. Because of the substantial differences between individual eyes, standard observer response curves or eye sensitivity curves have been established. The sensitivity characteristics of a physical photometer should be equivalent to the standard observer. The required match is typically achieved by adding filters between the sensitive elements of the meter and the light source.

Charlie Chong/ Fion Zhang


2.5.2

Photopic and Scotopic Vision

The human eye contains two basic types of retinal receptors known as rods and cones. They differ not only in relative spectral response and other properties but by orders of magnitude in responsivity. The rods are the most sensitive and spectrally respond more to the blue and less to the red end of the spectrum. However, they do not actually give the sensation of color as the cones do. Luminance is measured in candelas (cd). When the eye has been subjected to a field luminance of more than 3.0 cd•m-2 (0.27 cd•ft -2) for more than a few minutes, the eye is said to he in a light adapted state in which only the cones are responsible for vision; the state is also known as photopic vision or fovea/ vision. At light levels five or more orders of magnitude below this, at or below 3.0 x 10-5 cd.m-2 (2.7 x 10-6 cd•ft -2), the cones no longer function and the responsivity is that of the rods. This is known as dark adapted, or scotopic vision or parafoveal vision. After being light adapted, the eye usually requires a considerable time to become dark adapted when the light level is lowered. The time needed depends on the initial luminance of the starting condition but is usually achieved in 30 to 45 minutes.

Charlie Chong/ Fion Zhang


Photopic and Scotopic Vision

http://www.solarlightaustralia.com.au/2013/05/30/photopic-scotopic-and-mesotopic-lumens/

Charlie Chong/ Fion Zhang


Photopic and Scotopic Vision

Charlie Chong/ Fion Zhang


Between the levels at which the eye exhibits photopic and scotopic responses the spectral and other responses of the eye are continuously variable; this is known as the mesopic state, in which properties of both cone and rod receptors contribute. Many visual tests are made under photopic conditions but most measurements of fluorescent and phosphorescent materials are made under scotopic and mesopic conditions. Because of the changes in the eye's spectral response at these levels it is necessary to take luminance into account when evaluating the results of such measurements.

Charlie Chong/ Fion Zhang


Photopic and Scotopic Vision

http://lumenistics.com/consider-photopic-scotopic-mesopic-vision-before-specifying-lumen-requirements/

Charlie Chong/ Fion Zhang


2.5.3

Measurable Quantities

As indicated in Table 2, many characteristics of light, light sources, lighting materials and lighting installations may be measured, including 1. 2. 3. 4. 5. 6. 7. 8. 9.

illuminance, luminance, luminous intensity, luminous flux, contrast, color (appearance and rendering), spectral distribution, electrical characteristics and radiant energy.

Charlie Chong/ Fion Zhang


TABLE 2. Measurable characteristics of light, light sources and lighting materials

Charlie Chong/ Fion Zhang


2.6

Principles of Photometry

2.6.0

General

Photometric measurements frequently involve a consideration of the cosine law and the inverse square law (strictly applicable only for point sources). 2.6.1

Inverse Square Law

The inverse square law (see Fig. 6a) states that the illumination E (in lux) at a point on a surface varies directly with the luminous intensity I of the source and inversely as the square of the distance d between the source and the point. If the surface at the point is normal to the direction of the incident light, the law may be expressed as: E= I/d2

(Eq. 5)

This equation is accurate within 0.5 percent when d is at least five times the maximum dimension of the source, as viewed from the point on the surface.

Charlie Chong/ Fion Zhang


Inverse Square Law

http://pondscienceinstitute.on-rev.com/svpwiki/tiki-index.php?page=Square%20Law

Charlie Chong/ Fion Zhang


2.6.2

Lambert Cosine Law

The Lambert cosine law (see Fig. 6b) states that the illuminance of any surface varies as the cosine of the angle of incidence. The angle of incidence 0 is the angle between the normal to the surface and the direction of the incident light. The inverse square law and the cosine law can be combined to yield the following relationship (in lux): E = I/d2 Cos ϴ

Charlie Chong/ Fion Zhang

(Eq. 6)


Lambert Cosine Law

http://webx.ubi.pt/~hgil/FotoMetria/HandBook/ch06.html

Charlie Chong/ Fion Zhang


2.6.3

Photometric Reference Standards

Reference standards for candlepower, luminous flux and color are established by national standard laboratories. A primary reference standard is reproducible from specifications and is typically found only in a national laboratory. Secondary reference standards are usually derived directly from primary standards and must be prepared using precise electrical and photometric equipment. Preservation of the reference standard's rating is very important. Accordingly, a standard is used as seldom as possible and is handled and stored with care. Photometric reference lamps are used when accuracy warrants the highest attainable precision. Because of the cost of reference standards, so-called working standards are prepared for frequent use A laboratory can prepare its own working standards for use in calibrating photometers. The working standard is not used to conduct a test, except where a direct comparison is necessary.

Charlie Chong/ Fion Zhang


2.6.4

Photometric Applications

Photometric measurements make use of the basic laws of photometry, applied to readings from visual photometric comparison or photoelectric instruments. Various procedures are discussed below Direct photometry is the simultaneous comparison of a standard lamp and an unknown light source. Substitution photometry is the sequential evaluation of the desired photometric characteristics of a standard lamp and an unknown light source in terms of an arbitrary reference. To avoid the use of standard lamps, relative photometry is often used. Relative photometry is the evaluation of a desired photometric characteristic based on an assumed lumen output of the test lamp. Alternately, relative photometry refers to the measurement of one uncalibrated light source to another uncalibrated light source. It is sometimes necessary to measure the output of sources that are nonsteady or cyclic and, in such cases, extreme care should be taken.

Charlie Chong/ Fion Zhang


2.6.5

Means of Achieving Attenuation

During photometric measurement, it often becomes necessary to reduce the luminous intensity of a source in a known ratio to bring it within the range of the measuring instrument. A rotating sector disk with one or more angular apertures is one means of doing this. If such a disk is placed between a source and a surface and is rotated at such speed that the eye perceives no flicker, the effective luminance of the surface is reduced in the ratio of the time of exposure to the total time (Talbot's law). The reduction is by the factor ϴ/360 degrees. The sector disk has advantages over many filters: (1) it is not affected by a change of characteristics over time and (2) it reduces luminous flux without changing its spectral composition. Sector disks should not be used with light sources having cyclical variation in output.

Charlie Chong/ Fion Zhang


Various types of neutral filters of known transmittance are also used for attenuation. Wire mesh or perforated metal filters are perfectly neutral but have a limited range. Partially silvered mirrors have high reflectance but the reflected light must be controlled to avoid errors in the photometer. When a mirror filter is perpendicular to the light source photometer axis, serious errors may be caused by multiple reflections between the filter and receiver surface. This can be avoided by mounting the filter at a small angle (not over 3 degrees) from perpendicular at a sufficient distance from the receiver surface to throw reflections away from the photometric axis. In this canted position, care must be taken not to reflect light from adjacent surfaces on to the receiver. Also, it is difficult to secure completely uniform transmission over all parts of the surface. So-called neutral glass filters are seldom neutral and transmission characteristics should be checked before use. In general, they have a characteristic high transmittance in the red and low transmittance in the blue, so that spectral correction filters may be required. However, this type of filter varies in transmittance with ambient temperature, as do many other optical filters.

Charlie Chong/ Fion Zhang


Neutral gelatin filters are satisfactory, although not entirely neutral. Some have a small seasoning effect (losing neutrality over a period of time). They must be protected by mounting between two pieces of glass and must be watched carefully for loss of contact between the glass and gelatin. Filters should not be stacked unless cemented, because of errors that may be created by interference between surfaces. With modem metering techniques, electronic alterations can be accomplished to keep the output of a receiver and amplifier combination in range of linearity and readability.

Charlie Chong/ Fion Zhang


2.7

Photometers

A photometer is a device for measuring radiant energy in the visible spectrum. Various types of physical instruments consist of an element sensitive to radiant energy and appropriate measuring equipment and are used to measure ultraviolet and infrared energy. When used with a filter to correct their response to the standard observer, such devices can measure visible light. In general, photometers may be divided into two types: 1. laboratory photometers are usually fixed in position and yield results of highest accuracy, 2. portable photometers are of lower accuracy for making measurements in the field.

Charlie Chong/ Fion Zhang


Both types of meters may be grouped according to function, such as the photometers used to measure luminous intensity (candlepower), luminous flux, illuminance, luminance, light distribution, light reflectance and transmittance, color, spectral distribution and visibility Illuminance Photometers. Visual photometric methods have largely been supplanted by physical methods. Because of their simplicity, vision based photometry methods are still used in educational laboratories for demonstrating photometric principles and for less routine types of photometry. Photoelectric photometers' may be divided into two classes: 1. those employing solid state devices such as photovoltaic and photoconductive cells and 2. those employing photoemissive tubes.

Charlie Chong/ Fion Zhang


Photometry

http://safety.fhwa.dot.gov/roadway_dept/night_visib/lighting_handbook/

Charlie Chong/ Fion Zhang


2.8

Photovoltaic Cell Meters

2.8.0

General

A photovoltaic cell is one that directly converts radiant energy into electrical energy. It provides a small current that is about proportional to the incident illumination and also produces a small electromotive force capable of forcing this current throtigh a low resistance circuit. Photovoltaic cells provide much larger currents than photoemissive cells and can directly operate a sensitive instrument such as a microammeter or galvanometer. However, as the resistance of their circuit increases, photovoltaic cells depart from linearity of response at higher levels of incident illumination. Therefore, for precise results, the external circuitry and metering should apply nearly zero impedance across the photocell.

Charlie Chong/ Fion Zhang


Some portable illuminance meters consist of a photovoltaic cell or cells, connected to a meter calibrated directly in lux or footcandles. However, with solid state electronic devices, operational amplifiers have been used to amplify the output of photovoltaic cells. The condition that produces the most favorable linearity between cell output and incident light is automatically achieved by reducing the potential difference across the cell to zero. The amplifier power requirements are small and easily supplied by small batteries. In addition, digital readouts may be conveniently used to eliminate the ambiguities inherent in deflection instruments.

Charlie Chong/ Fion Zhang


2.8.1

Spectral Response

The spectral response of photovoltaic cells is quite different from that of the human eye and color correcting filters are usually needed.."-" As an example, Fig. 7 illustrates the degree to which a typical commercially corrected selenium photovoltaic cell commonly used in illuminance meters approximates the standard spectral luminous efficiency curve. Cells vary considerably in this respect and for precise laboratory photometry each cell should be individually color corrected. The importance of color correction can be illustrated by comparing the human eye match under illumination generated by a monochromatic source. For example, if a predominant blue light source is used, the majority of the visible energy is concentrated near 465 nm. It can be seen in Fig. 7 that the relative eye response and the filtered receptor response are about 10 and 15 percent. This represents a 50 percent differential and indicates that the photoreceptor could read as much as 50 percent high under the blue light source. Care should be taken to correct for this difference.

Charlie Chong/ Fion Zhang


FIGURE 7. Average spectral sensitivity characteristics of selenium photovoltaic cells, compared with relative eye response (luminous efficiency curve)

Charlie Chong/ Fion Zhang


2.8.2

Transient Effects

When exposed to constant illumination, the output of photovoltaic cells requires a short finite rise time to reach a stable output. The output may decrease slightly over time because of fatigue. 4U-42 Rise times for silicon cells often are considerably shorter than for selenium cells.

Charlie Chong/ Fion Zhang


2.8.3

Effect of Incidence Angle

At high incidence angles, part of the light reaching a photovoltaic cell is reflected by the cell surface and the cover glass and some may be obstructed by the rim of the case. The resulting error increases with angle of incidence. When an appreciable portion of the flux comes at wide angles, an uncorrected meter may read illuminance as much as 25 percent below the true value. The cells used in most illuminance meters are provided with diffusing covers or some other means of correcting the light sensitive surface to approximate the true cosine response. The component of illuminance contributed by single sources at wide angles of incidence may be determined by positioning the cell perpendicular to the direction of the light and multiplying the reading by the cosine of the incidence angle.

Charlie Chong/ Fion Zhang


The possibility of cosine error must be taken into consideration for some laboratory applications of photovoltaic cells. One satisfactory solution to the problem consists of placing a nonfluorescent opal diffusing acrylic plastic disk with a matte surface over the cell. At high angles of incidence, the disk reflects the light specularly so that the readings are too low. This can be compensated by allowing light to reach the cell through the edges of the plastic. The readings at very high angles are then too high but can be corrected using a screening ring. In general it is important that the opal diffusing plastic disk with a matte surface should be nonfluorescent or erroneous values of illuminance may be obtained in the presence of blueviolet and ultraviolet radiations; such a situation is common in fluorescent penetrant and magnetic particle testing applications in which measurements of the ambient visible light, in he presence of the blacklight are required by certain industrial and military specifications. Certain photometers are actually provided with fluorescent diffusers and should be avoided in such situations.

Charlie Chong/ Fion Zhang


2.8.4

Effect of Temperature.

Wide temperature variations affect the performance of photovoltaic cells, particularly when the external resistance of the circuit is high. Prolonged exposure to temperatures above 50 째C (120 째F) permanently damages selenium cells. Silicon cells are considerably less affected by temperature. Measurements at high temperatures and at high illuminance levels should therefore be made rapidly to avoid overheating the cell. Hermetically sealed cells provide greater protection from the effects of temperature and humidity. When using photovoltaic cells at other than their calibrated temperature, conversion factors may be used or means may be provided to maintain cell temperatures near 25 째C (77 째F).

Charlie Chong/ Fion Zhang


2.8.5

Effect of Cyclical Variation of Light

When electric discharge sources are operated on alternating current power supplies, precautions should be taken with regard to the effect of frequency on photocell response. In some cases, these light sources may be modulated at several kilohertz. Consideration should then be given to whether the response of the cell is exactly equivalent to the Talbot's law response of the eye for cyclic varying light. Because of the internal capacitance of the cell, it cannot always be assumed that its dynamic response exactly corresponds to the mean value of the illuminance. It has been found that a low or zero resistance circuit is the most satisfactory for determining the average intensity of modulated or steady state light sources with which photovoltaic cell instruments are generally calibrated. Although a microammeter or galvanometer appears to register a steady photocell current, it may not be receiving such a current. The meter actually may be receiving a pulsating current which it integrates because its natural period of oscillation is long compared to the pulses. Meters are available that can average over a period of time, eliminating the effect of cyclic variation.

Charlie Chong/ Fion Zhang


2.8.6

Photometer Zeroing

It is important to check photometer zeroing before taking measurements. If an analog meter is used, this requires manual positioning of the indicator to zero. For any type of equipment using an amplifier, it may be necessary to zero both the amplifier and the dark current (current flowing through the device while it is in absolute darkness). When possible, it should be verified that the meter remains correctly zeroed when the photometer scale is changed. Alternately, any deviation from zero under dark current conditions may be measured and subtracted from the light readings.

Charlie Chong/ Fion Zhang


2.8.7

Electrical interference

With electronic meters, care should be taken to eliminate interference induced in the leads between the cell and the instrumentation. This can be achieved by filter networks, shielding, grounding or combinations of the above.

Charlie Chong/ Fion Zhang


2.9

Meters Using Photomultiplier Tubes

2.9.0

General

Photoelectric tubes produce current when radiant energy is received on a photoemissive surface and then amplified by a phenomenon known as secondary emission. These tubes require a high voltage to operate (2 to 5 kV) and an amplifier to provide a measurable signal. The resulting current may be measured by a deflection meter, oscillograph or a digital output device. Dark current (current flowing through the device while it is in absolute darkness) must be compensated for in the circuitry or subtracted from the lighted tube output. Meters using this device are often extremely sensitive. Photomultiplier tubes can he damaged by shock and the calibration of the meter can be altered by strong magnetic fields. In addition, the device is temperature sensitive and should be operated at or below room temperature. As with other photoelectric devices, the photomultiplier spectral response curve does not match the human eye and color correcting filters are required.

Charlie Chong/ Fion Zhang


Because of the large number of photomultiplier types available, manufacturers commonly supply the proper optical filter for their design. When a photomultiplier tube is used in conjunction with an optical lens system, the resulting luminance meter can be of high sensitivity and broad range. Keywords: Secondary emmision

Charlie Chong/ Fion Zhang


2.9.1

Luminance Photometers

The basic principles for the measurement of illuminance apply equally well for the measurement of luminance. Luminance meters are essentially a photoreceptor in front of a focusing mechanism. By suitable optics, the luminance of a certain size spot, when cast onto the receptor, generates an electrical signal that is dependent on the object luminance. This signal can be measured and, assuming that the necessary calibration has been performed, a reading is produced that directly measures luminance. Usually an eyepiece is provided so that the user is able to see the general field of view through the instrument. By changing the lens system in front of the photoreceptor, different fields of view and therefore different sizes of measurement area may be achieved. This can vary from areas subtending a few minutes of arc up to several degrees. Photoreceptors may be selenium but are usually silicon cells or photomultipliers. The meter reading may be analog or digital and either built into the meter or remote. Amplifier controls for zeroing and scale selection are usually provided. Other options include optical filters for color work or scale selection by means of neutral density filters.

Charlie Chong/ Fion Zhang


Luminance Photometers

Charlie Chong/ Fion Zhang


2.9.1

Brightness Spot Meter

The brightness spot meter is a photoelectric device for measuring the luminance of small areas, typically 0.25, 0.5 or 1 degree field of view. A beam splitter allows a portion of the light from the objective lens to reach a reticule viewed by the eyepiece. The remainder of the light is reflected in front of the photomultiplier tube. The output of the tube after amplification is read on a microammeter with a scale calibrated in candelas per square meter or footlamberts. One of the filters provided for such instruments corrects the response of the photomultiplier to the standard spectral luminous efficiency curve. Full scale deflection is produced by 10-1 to 10-7 cd•m-2

Charlie Chong/ Fion Zhang


2.9.3

High Sensitivity Photometer

One version of the photomultiplier photometer has interchangeable field apertures covering fields from arc minutes to 3 degrees in diameter. Full scale sensitivity ranges are from 10-4 to 108 cd•m-2. In this meter, readings of the measured light are free from the effects of polarization because there are no internal reflections of the beam. The spectral response of each photometer is individually measured. The filters to match it best to the standard spectral luminous efficiency curve are then inserted. Filters are also included to permit evaluation of polarization and color factors.

Charlie Chong/ Fion Zhang


High Sensitivity Photometer

Charlie Chong/ Fion Zhang


High Sensitivity Photometer

Charlie Chong/ Fion Zhang


2.10

Equivalent Sphere Illumination

2.10.1 Photometers Equivalent sphere illumination (ESI) may be used as a tool as part of the evaluation of lighting systems. The equivalent sphere illumination of a visual task at a specific location in a room illuminated with a specific lighting system is defined as that level of perfectly diffuse (spherical) illuminance that makes the visual task as visible in the sphere as it is in the real lighting environment. Measurements may be made visually and/or physically. In the visual method the measurement is made by comparison between a task viewed in the measured (actual) environment and the task viewed in a luminous sphere by using a visibility meter. The physical method, however, is based on certain algorithms and requires only measurements in situ of background illuminance Lb and task illuminance Lt. All physical equivalent sphere illumination devices measure Lb and Lt in one way or another. Measuring devices are discussed below in chronological order of development.

Charlie Chong/ Fion Zhang


2.10.2 Visual Task Photometer The visual task photometer is a basic, reference instrument against which others are often compared. Equivalent sphere illumination is not measured directly: Lb and Lt are measured and equivalent sphere illumination is subsequently calculated. The task to be evaluated is mounted on a target shifter and a telephotometer is aimed at it from the desired viewing angle. The task and telephotometer (usually mounted on a cart) are then positioned so that: â– â–

the task is in the location where the measurement is to he made and the telephotometer is facing the proper direction of view.

The standard body shadow (attached to the telephotometer) shades the task in a manner similar to an actual observer. The Lb value is measured, then the shifter is activated to bring the target into view of the telephotometer. The Lt value is then measured.

Charlie Chong/ Fion Zhang


2.10.3 Visual Equivalent Sphere Illumination Meter The visual equivalent sphere illumination meter' consists of an optical system, variable luminous veil, target carrier, luminous sphere, illuminance meter (inside the sphere) and a body shadow. A task is placed on the target carrier and viewed through the optical system. The contrast of the task is then reduced to threshold by adjusting a variable luminous veil. Field luminance is automatically kept constant so as not to alter the adaptation luminance of the observer's eye. The task is then carried inside the sphere and the optical system is adjusted until the target is again at threshold (task visibility is the same inside the sphere as it was outside). The illuminance in the sphere is measured directly to determine equivalent sphere illumination.

Charlie Chong/ Fion Zhang


Visual Equivalent Sphere Illumination Meter

Charlie Chong/ Fion Zhang


Visual Equivalent Sphere Illumination Meter

Charlie Chong/ Fion Zhang


Visual Equivalent Sphere Illumination Meter

Charlie Chong/ Fion Zhang


Visual Equivalent Sphere Illumination Meter

Charlie Chong/ Fion Zhang


Visual Equivalent Sphere Illumination Meter

Charlie Chong/ Fion Zhang


2.10.4 Physical Equivalent Sphere Illumination Meters Two devices are available that do not rely on the actual presence of a task for their precision. Instead, they use numerical data that represents the task's reflectance characteristics: bidirectional reflectance distribution functions. One meter uses cylinders that represent an optical analog of the visual task photometer." There are two cylinders used per measurement: one representing a task's Lb is called the background cylinder and one representing Lb - Lt is called the difference cylinder. These two parameters can be used to calculate equivalent sphere illumination in place of Lb and Lt alone. Each cylinder has its own body shadow. A cosine corrected illuminance probe is placed where the measurement is desired. The background cylinder is placed atop the probe and oriented in the appropriate viewing direction.

Charlie Chong/ Fion Zhang


Background illuminance is then recorded. The background cylinder is replaced by the difference cylinder, oriented in the same direction and the difference illuminance is recorded. Equivalent sphere illumination is then calculated from the background and difference illuminance readings. A second meter using bidirectional reflectance distribution functions is a scanning luminance meter." This instrument contains a narrow field luminance probe attached to a motorized scanning apparatus and a minicomputer to control scanning, store the distribution function data and perform calculations. To use this device for measuring equivalent sphere illumination, the probe is positioned at the desired location and the minicomputer is instructed to begin scanning. Luminances are multiplied by their appropriate bidirectional reflectance distribution functions to determine Lb and Lt. The minicomputer then calculates equivalent sphere illumination directly. The scanning luminance meter has the capabilities of rotating the task in any viewing direction and of determining the equivalent sphere illumination on different tasks with only one set of scanning measurements.

Charlie Chong/ Fion Zhang


2.10.5 Visual Task Photometers The bidirectional reflectance distribution functions used with physical meters are obtained by illuminating a task from a particular direction and by viewing the task from some other unique direction. A visual task photometer is used to perform these measurements. The visual task photometer is the same as that used for equivalent sphere illumination measurements except that it includes a collimated light source that can be positioned anywhere on a hemisphere over the task. The task is illuminated from each azimuth and declination angle (usually in 5 degree increments) and the reflectance is measured at each angle. The collection of bidirectional reflectance data for the task and its background form the distribution function.

Charlie Chong/ Fion Zhang


2.11

Reflectometers

Reflectometers are photometers used to measure reflectance of materials or surfaces in specialized ways. The reflectometer measures diffuse, specular and total reflectance. Those instruments designed to determine specular reflectance are known as glossmeters. One popular reflectometer uses a collimated beam and a photovoltaic cell. The beam source and cell are mounted in a fixed relationship in the same housing. The housing has an aperture through which the beam travels. This head or sensor is set on a standard reflectance reference with the aperture against the standard. The collimated beam strikes the standard at a 45 degree angle. The photovoltaic cell is constructed so that it measures the light reflected at 0 degrees from the standard. The instrument is then adjusted to read the value stated on the standard. The sensor is placed on the test surface and the reading is recorded. Two cautions are recommended for use of reflectometers. The reference standard should be in the range of the value expected for the surface to be measured. Also, if the area to be considered is large, several measurements should he taken and averaged to obtain a representative value.

Charlie Chong/ Fion Zhang


Another type of reflectometer (see Fig. 8) measures both total reflectance and diffuse transmittance."' The instrument consists of two spheres, two light sources and two photovoltaic cells. The upper sphere is used alone for the measurement of reflectance. The test object is placed over an opening at the bottom of the sphere and a collimated beam of light is directed on it at about 30 degrees from normal. The total reflected light, integrated by the sphere, is measured by two cells mounted in the sphere wall. The tube carrying the light source and the collimating lenses is then rotated so that the light is incident on the sphere wall and a second reading is taken. The test object is in place during both measurements, so that the effect on both readings of the small area of the sphere surface it occupies is the same. The ratio of the first reading to the second reading is the reflectance of the object for the conditions of the test. Test objects made of translucent materials should be backed by a nonreflecting diffuse material. Transmittance for diffuse incident light is measured by using the light source in the lower sphere and taking readings with and without the test object in the opening between the two spheres. The introduction of the test object changes the characteristics of the upper sphere. Correction must be made to compensate for the introduced error. Charlie Chong/ Fion Zhang


Various instruments are available for measuring such properties as specular reflectance and the gloss characteristics of materials. For example, an instrument similar to that described above for the measurement of diffuse reflectance may be used, except that the cell is fixed at 45 degrees on the side of the test object opposite to the light source, thus measuring the specularly reflected beam. The angle subtended by the photocell to the test object affects the reading and appropriate compensations are recommended.

Charlie Chong/ Fion Zhang


FIGURE 8. A light cell reflectometer In an arrangement for transmittance measurement

Charlie Chong/ Fion Zhang


2.12

Radiometers

Radiometers are sometimes called radiometric photometers and are used to measure radiant power over a wide range of wavelengths, including the ultraviolet, visible or infrared spectral regions. Radiometers may use detectors that are nonselective in wavelength response or that give adequate response in the desired wavelength band. Nonselective detectors (response varies little with wavelength) include thermocouples, thermopiles, bolometers and pyroelectric detectors. One class of wavelength selective detectors is photoelectric and includes photoconductors, photoemissive tubes, photovoltaic cells and solid state sensors such as photodiodes, phototransistors and other junction devices. The overall response of such detectors can he modified by using appropriate filters to approximate some desired function. For example, these detectors can be color corrected by means of a filter to duplicate the standard luminous efficiency curve in the visible range or to level a detector's response to radiant power over some hand of wavelengths. The corrections must compensate for any selectivity in the spectral response of the optical system. Care must be exercised to eliminate a detector's response to radiation lying outside the range of interest.

Charlie Chong/ Fion Zhang


When a monochromator is used to disperse the incoming radiation, the radiant power can be determined in a very small band of wavelengths. Such an instrument is called a spectroradiometer and is used to determine the spectral power distribution (the radiant power per unit wavelength as a function of wavelength) of the radiation in question. The spectral power distribution is fundamental; from it radiometric, photometric and colorimetric properties of the radiation can be determined. The use of digital processing has greatly facilitated both the measurement and the use of the spectral power distribution. The range of spectral response generally depends on the nature of the detector. Photomultiplier tubes extend wavelength sensitivity from 125 to 1,100 nm.

Charlie Chong/ Fion Zhang


Various types of silicon photodiodes cover the range from 200 to 1,200 nm. In the infrared range are intrinsic germanium (0.9 to 1.5 μm), lead sulfide (1.0 to 4.0 μm), indium arsenide (1.0 to 3.6 μm), indium antimonide (2.0 to 5.4 μm), mercury cadmium telluride (1.0 to 13 μm) and germanium doped with various substances such as zinc (2.0 to 40 μm). The response of nonselective detectors ranges from near ultraviolet to 30μm (300,000 A) and beyond. The electrical output of detectors (voltage, current or charge) is very small and special precautions are often required to achieve acceptable signal levels, signal-to-noise ratios and response times (for rapidly varying signals). Photon counting and charge integration techniques are used for extremely low radiation level.

Charlie Chong/ Fion Zhang


In all radiometric work, it is important to avoid stray radiation and care must be taken to ensure its exclusion. This is difficult because stray radiation is not visible and a surface seen as black may actually be an excellent reflector of radiant energy outside the visible spectrum. Often, unwanted radiation can be absorbed by an appropriate filter. Sometimes such a high flux must be removed to avoid the absorption filter's heating to the point of breaking or its transmittance for other desired wavelengths is altering. Because radiated flux of some wavelengths is dispersed or absorbed by a layer of air between the radiator and the detector, consideration must be given to the placement of the source and the detector and to the medium surrounding them.

Charlie Chong/ Fion Zhang


2.13

Spectrophotometers

Photometry is the measurement of power in the visible spectrum, weighted according to the visual response curve of the eye. When the power is measured as a function of wavelength, the measurement is referred to as spectrophotometry. Its applications extend from precise quantitative chemical analysis to the exact determination of the physical properties of matter. Spectrophotometry is important for the determination of spectral transmittance and spectral reflectance. It is also applied to the measurement of the spectral emittance of lamps, in which case it is known as spectroradiometry. This form of measurement commonly covers the visible portion of the spectrum, the ultraviolet and near infrared wavelengths. Instruments used for performing such measurements are called spectrophotometers and spectroradiometers.

Charlie Chong/ Fion Zhang


These devices consist basically of a monochromator (separates or disperses the wavelengths of the spectrum using prisms or gratings) and a receptor (measures the power contained within a certain wavelength range of the dispersed light). If the spectrum is examined visually rather than by a photoreceptor, the instrument is known to as a spectroscope. In the visible spectrum, the only fundamental means of examining a color for analysis, standardization and specification is by spectrophotometry. In addition, this is the only means of color standardization that is independent of material color standards (always of questionable permanence) and independent of the abnormalities of color vision existing among so-called normal observers. Commercial development of spectrophotometers has extended the wavelength range from about 200 to 2,500 nm, made them automatically record and added tristimulus integration. Self scanned silicon photodiode arrays provide nearly instantaneous determination of spectral power distributions. Charlie Chong/ Fion Zhang


Spectrophotometers

Charlie Chong/ Fion Zhang


Spectrophotometers

Charlie Chong/ Fion Zhang


Spectrophotometers

Charlie Chong/ Fion Zhang


Spectrophotometers

Charlie Chong/ Fion Zhang


Spectrophotometers

Charlie Chong/ Fion Zhang


Spectrophotometers

http://www.nature.com/nature/journal/v512/n7512/full/nature13382.html

Charlie Chong/ Fion Zhang


2.14

Types of Photometers

2.14.1 Optical Bench Photometers Optical bench photometers are used for the calibration of instruments for illumination measurement. They provide a means for mounting light sources and photocells in proper alignment and a means for easily determining the distances between them. If the source is of known luminous intensity (candlepower), the inverse square law is used to compute illuminance, provided that the source-to-detector distance is at least five times the maximum source dimension.

Charlie Chong/ Fion Zhang


2.14.2 Distribution Photometers Luminous intensity (candlepower) measurements are made on a distribution photometer which may be one of the following types: (1) goniometer and single cell, (2) fixed multiple cell, (3) moving cell and (4) moving mirror. All types of photometers have advantages and disadvantages. The significance attached to each advantage or disadvantage depends on factors such as available space and facilities, polarization requirements and economic considerations.

Charlie Chong/ Fion Zhang


2.14.3 Goniometer and Single Cell The light source is mounted on a goniometer, which allows the source to rotate about horizontal and vertical axes. The candlepower is measured by a single fixed cell. There are several kinds of goniometers, each related to the type of source being photometered and the facilities in which it is located. With the use of computers, the coordinate system of one goniometer system can be easily changed to another coordinate system and the compatibility of data reporting becomes practical. Figure 9 shows two types of goniometer systems.

Charlie Chong/ Fion Zhang


FIGURE 9. Goniometer variations: (a J the projector turns about a fixed horizontal axis and about a second axis which, in the position of rest, is vertical and, on rotation, follows the movement of the horizontal axis; and (b) the light source turns about a fixed vertical axis and also about a horizontal axis following the movement of the vertical axis; the grid lines shown represent the loci traced by the photocell as the goniometer axes are rotated

Charlie Chong/ Fion Zhang


2.14.4 Fixed Multiple Cell Photometer In a multiple cell photometer, many individual photocells are positioned at various angles around the light source under test. Readings are taken on each photocell to determine the light intensity or candlepower distribution (see Fig. 10).

Charlie Chong/ Fion Zhang


FIGURE 10. Schematic side elevation of a fixed multiple cell photometer

Charlie Chong/ Fion Zhang


2.14.5 Moving Cell Photometer The moving cell photometer (Fig. 11) has a photocell that rides on a rotating boom or an arc shaped track. The light source is centered in the arc traced by the cell. Readings are collected with the cell positioned at the desired angular settings. Sometimes a mirror is placed on a boom to extend the test distance.

Charlie Chong/ Fion Zhang


FIGURE 11. Schematic side elevation of a moving cell photometer

Charlie Chong/ Fion Zhang


2.14.6 Moving Mirror Photometer In the moving mirror photometer, a mirror rotates around the light source, reflecting the candlepower to a single photocell. Readings are taken at each angle as the mirror moves to that location.

Charlie Chong/ Fion Zhang


2.14.7 Integrating Sphere Photometer The total luminous flux from a source can be measured by a form of integrator sphere. Other geometric forms are sometimes used. The theory of the integrating sphere assumes an empty sphere whose inner surface is perfectly diffusing and of uniform nonselective reflectance. Every point on the inner surface reflects to every other point and the illuminance at any point is made up of two components: the flux coming directly from the source and that reflected from other parts of the sphere wall. With these assumptions, it follows that, for any part of the wall, the illuminance and the luminance from reflected light only is proportional to the total flux from the source, regardless of its distribution. The luminance of a small area of the wall or the luminance of the outer surface of a diffusely transmitting window in the wall, carefully screened from direct light from the source but receiving light from other portions of the sphere, is therefore a relative measurement of the flux output of the source.

Charlie Chong/ Fion Zhang


The presence of a finite source, its supports, electrical connections, the necessary shield and the aperture or window, are all departures from the assumptions of the integrating sphere theory. The various elements entering into the considerations of a sphere, as an integrator, make it undesirable to use a sphere for absolute measurement of flux unless various correction factors are applied. This does not detract from its use when a substitution method is employed

Charlie Chong/ Fion Zhang


Integrating Sphere Photometer

Charlie Chong/ Fion Zhang


Integrating Sphere Photometer

Charlie Chong/ Fion Zhang


Charlie Chong/ Fion Zhang


Turn static files into dynamic content formats.

Create a flipbook
Issuu converts static files into: digital portfolios, online yearbooks, online catalogs, digital photo albums and more. Sign up and create your flipbook.