UNCLASSIFIED AREA OR AREAS CLASSIFIED AS ZONE 0 OR CLASS I, DIV 1, GROUPS B, C, D Fluke 725Ex Multi

Page 1

5

4

UNCLASSIFIED AREA OR AREAS CLASSIFIED AS ZONE 0 OR CLASS I, DIV 1, GROUPS B, C, D

D

3

FLUKE 725Ex

2

FLUKE 725Ex

Source V, Frequency, RTD (Ohms)

CSA: IS Class I; Div 1; Groups B-D T = 171 ºC AEx ia IIB, T = 171 ºC

THE CONNECTIONS SHOWN FOR J1-J6 JACKS CAN BE USED SIMULTANEOUSLY WITH ANY CONNECTION SHOWN FOR J7-J8 JACKS

W ONLY CONNECTIONS SPECIFIED ON THIS

DRAWING ARE ALLOWED. OTHER CONNECTIONS MAY INVALIDATE INTRINSIC SAFETY

Approved Associated Apparatus

+ J1

J3

J2

J4 J6

J7

J1

J3

J8

J2

J4 J6

Approved Associated Apparatus or

Simulate mA

Mode

C

J1

J3

J7

+ Approved Associated Apparatus -

4W J1

J4 J6

2 Wire

2 W ONLY

3 Wire

2 W and 3 W

4 Wire

2 W,3 W and 4 W

J8

J2

J4 J6

J8

Simple Apparatus or Approved Intrinsically - Safe Apparatus

Source mA

2W 2W 3W

J3

J3-J4: Uo = 13.6 V, Io =25.2 mA, Po = 86 mW J5-J6: Uo = 13.7 V, Io = 26 mA, Po = 90 mW J7-J8: Uo = 13.7 V, Io = 76 mA, Po = 261 mW C

J1

J3

J2

J4 J6

J8

Loop

+

Input

FLUKE 725Ex

-

+ Simple Apparatus or Approved Intrinsically - Safe Apparatus

Measure V using J7-J8 jacks

Measure Thermocouple

J2

J4 J6

J1

J3

J8

J2

J4 J6

J7

J1

J3

J8

J2

J4 J6

J5

J5

FLUKE 725Ex

Measure V, Frequency

J7 J5

+ Simple Apparatus or Approved Intrinsically - Safe Apparatus

+ Simple Apparatus or Approved Intrinsically - Safe Apparatus

J8

Approved Intrinsically - Safe Apparatus

Fluke 700PEx pressure module only

FLUKE 725Ex FLUKE 725Ex

J3

J7

J2

J4 J6

J1

J3

THE PRESSURE MODULE CAN BE USED SIMULTANEOUSLY WITH ANY CONNECTION SHOWN ON THIS DIAGRAM FOR J1-J8 JACKS

J2

J4 J6

J5

J7 J5

J8

Approved Associated Apparatus or +

Approved Intrinsically - Safe Apparatus

J8

J1

+ Simple Apparatus or Approved Intrinsically - Safe Apparatus

J3

J1-J2 J3-J4 J5-J6 J7-J8

NOTES: 1. Connections to an Intrinsically Safe Apparatus /Associated Apparatus/Loop must be in acordance with notes 2 thru 5 2. Intrinsic Safety Entity Concept: The Intrinsic Safety Entity Concept allows the interconnection of two or more Approved Intrinsically Safe devices or Associated Apparatus with entity parameters not specifically examined in combination as a system when: Uo or Voc or Vt <= Vmax or Ui Io or Isc or It <= Imax or Ii Ca or Co >= Ci + Ccable La or Lo >= Li + Lcable Po <= Pi

4. The Fluke 725Ex Multi Function Calibrator may only be connected to terminals with compatible entity parameters except as indicated in Note 2 and must not be connected to internal circuitry of Associated Apparatus or Intrinsically Safe Apparatus. Failure to comply with this rule violates approval.

6. Refer to the 725Ex User Manual (provided on the CD) for additional safety information. 7.No revision to this drawing is permitted without approval from CSA and KEMA 12345

J7

J4 J6

Rev

B5

Description

By By

Original

Date Date Aug. 24, Nov. 18, 2004 2003

Fluke 725Ex 700PexCCD CCD

J8

Fluke Fluke PartPart Number: Number: XXXXXXXXXX 2117036 MARTEL 6800032

SCALE: None 5

4

3

B

3. Installation must be in accordance with Article 500 or Article 505 of the NEC (ANSI/NFPA70) and ANSI/ISA RP12.6 or equivalent.

J5 J2

Lo(mH) D(IIA)

BC(IIB )

D(IIA) 18.1 18.6 18.1 18.1

5. Applying more than 30 V or 100 mA to the input terminals invalidates the Calibrator's Ex Approval and may damage the unit. Immediately remove the unit from the Ex-hazardous area and contact a Fluke Service Center.

Pressure Mode

A

J1

Approved Associated Apparatus or +

Source Thermocouple

Co (µF) BC(IIB) 5.0 5.2 5.0 5.0

J7

+ Simple Apparatus or Approved Intrinsically - Safe Apparatus

FLUKE 725Ex

J7

0.01 µF 0 µF 0.01 µF 0 µF

J1-J2: Uo = 13.7 V, Io =96.5 mA, Po = 331 mW

B

J1

J1-J2: Ci = J3-J4: Ci = J5-J6: Ci = J7-J8: Ci =

Output Entity Parameters:

J5

+

FLUKE 725Ex

-

+ Simple Apparatus or Approved Intrinsically - Safe Apparatus

Measure mA feedback while providing power to the loop

Connections used

J5

J5 J2

J7

J3

FLUKE 725Ex

Measure RTD (Ohms) 2,3,4 Wire Mode

D

ENTITY PARAMETERS : All Inputs: J1-J2, J3-J4, J5-J6, J7-J8: Ui = 30 V, Ii = 100 mA Pi = 750 mW, Li= 0 mH

Loop

+

Input

FLUKE 725Ex

T1 = -10 ºC... +55 ºC

Input Entity Parameters:

J8

Approved Intrinsically - Safe Apparatus

Measure mA

-

J7 J5

J5

+

FLUKE 725Ex

1

Fluke 725Ex Multi function Calibrator

Measure mA feedback

2

Sheet 1 of 1 1

A