5
4
UNCLASSIFIED AREA OR AREAS CLASSIFIED AS ZONE 0 OR CLASS I, DIV 1, GROUPS B, C, D
D
3
FLUKE 725Ex
2
FLUKE 725Ex
Source V, Frequency, RTD (Ohms)
CSA: IS Class I; Div 1; Groups B-D T = 171 ºC AEx ia IIB, T = 171 ºC
THE CONNECTIONS SHOWN FOR J1-J6 JACKS CAN BE USED SIMULTANEOUSLY WITH ANY CONNECTION SHOWN FOR J7-J8 JACKS
W ONLY CONNECTIONS SPECIFIED ON THIS
DRAWING ARE ALLOWED. OTHER CONNECTIONS MAY INVALIDATE INTRINSIC SAFETY
Approved Associated Apparatus
+ J1
J3
J2
J4 J6
J7
J1
J3
J8
J2
J4 J6
Approved Associated Apparatus or
Simulate mA
Mode
C
J1
J3
J7
+ Approved Associated Apparatus -
4W J1
J4 J6
2 Wire
2 W ONLY
3 Wire
2 W and 3 W
4 Wire
2 W,3 W and 4 W
J8
J2
J4 J6
J8
Simple Apparatus or Approved Intrinsically - Safe Apparatus
Source mA
2W 2W 3W
J3
J3-J4: Uo = 13.6 V, Io =25.2 mA, Po = 86 mW J5-J6: Uo = 13.7 V, Io = 26 mA, Po = 90 mW J7-J8: Uo = 13.7 V, Io = 76 mA, Po = 261 mW C
J1
J3
J2
J4 J6
J8
Loop
+
Input
FLUKE 725Ex
-
+ Simple Apparatus or Approved Intrinsically - Safe Apparatus
Measure V using J7-J8 jacks
Measure Thermocouple
J2
J4 J6
J1
J3
J8
J2
J4 J6
J7
J1
J3
J8
J2
J4 J6
J5
J5
FLUKE 725Ex
Measure V, Frequency
J7 J5
+ Simple Apparatus or Approved Intrinsically - Safe Apparatus
+ Simple Apparatus or Approved Intrinsically - Safe Apparatus
J8
Approved Intrinsically - Safe Apparatus
Fluke 700PEx pressure module only
FLUKE 725Ex FLUKE 725Ex
J3
J7
J2
J4 J6
J1
J3
THE PRESSURE MODULE CAN BE USED SIMULTANEOUSLY WITH ANY CONNECTION SHOWN ON THIS DIAGRAM FOR J1-J8 JACKS
J2
J4 J6
J5
J7 J5
J8
Approved Associated Apparatus or +
Approved Intrinsically - Safe Apparatus
J8
J1
+ Simple Apparatus or Approved Intrinsically - Safe Apparatus
J3
J1-J2 J3-J4 J5-J6 J7-J8
NOTES: 1. Connections to an Intrinsically Safe Apparatus /Associated Apparatus/Loop must be in acordance with notes 2 thru 5 2. Intrinsic Safety Entity Concept: The Intrinsic Safety Entity Concept allows the interconnection of two or more Approved Intrinsically Safe devices or Associated Apparatus with entity parameters not specifically examined in combination as a system when: Uo or Voc or Vt <= Vmax or Ui Io or Isc or It <= Imax or Ii Ca or Co >= Ci + Ccable La or Lo >= Li + Lcable Po <= Pi
4. The Fluke 725Ex Multi Function Calibrator may only be connected to terminals with compatible entity parameters except as indicated in Note 2 and must not be connected to internal circuitry of Associated Apparatus or Intrinsically Safe Apparatus. Failure to comply with this rule violates approval.
6. Refer to the 725Ex User Manual (provided on the CD) for additional safety information. 7.No revision to this drawing is permitted without approval from CSA and KEMA 12345
J7
J4 J6
Rev
B5
Description
By By
Original
Date Date Aug. 24, Nov. 18, 2004 2003
Fluke 725Ex 700PexCCD CCD
J8
Fluke Fluke PartPart Number: Number: XXXXXXXXXX 2117036 MARTEL 6800032
SCALE: None 5
4
3
B
3. Installation must be in accordance with Article 500 or Article 505 of the NEC (ANSI/NFPA70) and ANSI/ISA RP12.6 or equivalent.
J5 J2
Lo(mH) D(IIA)
BC(IIB )
D(IIA) 18.1 18.6 18.1 18.1
5. Applying more than 30 V or 100 mA to the input terminals invalidates the Calibrator's Ex Approval and may damage the unit. Immediately remove the unit from the Ex-hazardous area and contact a Fluke Service Center.
Pressure Mode
A
J1
Approved Associated Apparatus or +
Source Thermocouple
Co (µF) BC(IIB) 5.0 5.2 5.0 5.0
J7
+ Simple Apparatus or Approved Intrinsically - Safe Apparatus
FLUKE 725Ex
J7
0.01 µF 0 µF 0.01 µF 0 µF
J1-J2: Uo = 13.7 V, Io =96.5 mA, Po = 331 mW
B
J1
J1-J2: Ci = J3-J4: Ci = J5-J6: Ci = J7-J8: Ci =
Output Entity Parameters:
J5
+
FLUKE 725Ex
-
+ Simple Apparatus or Approved Intrinsically - Safe Apparatus
Measure mA feedback while providing power to the loop
Connections used
J5
J5 J2
J7
J3
FLUKE 725Ex
Measure RTD (Ohms) 2,3,4 Wire Mode
D
ENTITY PARAMETERS : All Inputs: J1-J2, J3-J4, J5-J6, J7-J8: Ui = 30 V, Ii = 100 mA Pi = 750 mW, Li= 0 mH
Loop
+
Input
FLUKE 725Ex
T1 = -10 ºC... +55 ºC
Input Entity Parameters:
J8
Approved Intrinsically - Safe Apparatus
Measure mA
-
J7 J5
J5
+
FLUKE 725Ex
1
Fluke 725Ex Multi function Calibrator
Measure mA feedback
2
Sheet 1 of 1 1