The Zenyth 340 is a high-performance microplate reader with optional temperature control providing an indispensable tool research and routine applications
Biochrom® Anthos Zenyth 340 Microplate Reader
Key Features
PC-controlled microplate reader
Five interference filters supplied as standard that cover most common absorbance assays
Reading can be performed at timed intervals-ideal for kinetic assays
Linear and area scan capability
Additional wavelength filters from 340-750nm can be purchased separately to allow more specialized assays to be performed
Handles multiple plate formats from 6-384 wells microplates
Optional temperature control for temperature sensitive assays
Applications
Software ADAP BASIC SOFTWARE Supplied as standard with the Biochrom Anthos Zenyth 340, ADAP basic can be used to control all mechanical functions of the microplate reader including the plate transporter, filter selection and calibration as well as endpoint, kinetic and multiwavelength measurements. ADAP Basic allows the user to obtain raw data from the instrument for single, dual wavelength, endpoint and kinetic measurements. ADAP basic controls user access at different authorizations levels allowing users to work within FDA 21 CFR part 11 compliance and data can be stored or transferred to a spreadsheet for further analysis.
ADAP PLUS SOFTWARE ADAP Plus has the same functionality as ADAP Basic, as well as integrated analytical capabilities:
Absorbance measurements: ELISA and protein quantitation, cell proliferation and cell toxicity assays
Enzyme kinetic assays requiring automatic readings at timed intervals
Linear scan: for verification of agglutination and coagulation studies
Area scan: for precipitation and insolubility studies. Cell counting
Quantitative evaluation using standard curve including forecasting and curve-fitting Qualitative analysis including cut-off formulas and sample group identification Transformation formulas Plate layout, including assignment of blanks, standards, and controls Replicate elimination and test validation formulas Detailed graph view of multi-wavelength
ADAP EXPERT SOFTWARE ADAP Expert has the same functionality as ADAP Plus as well as the following
Data reduction methods of kinetic assays Tracking of up to 32,000 sample IDs Multi-test functions applying multiple methods to one plate
TECHNICAL DETAILS Biochrom Anthos Zenyth 340 Microplate Reader Light source Tungsten halogen lamp Photodetector Silicon photodiode Wavelength range 340 - 750nm Wavelength accuracy ± 2.0nm Wavelength ± 0.5nm reproducibility Bandwidth 8nm Resolution 0.0001 OD Indication range 0.000-4.000 OD (end point) Accuracy < ± 1.5% at 1 OD <±0.5% and ±0.005 OD from 0.1 OD to 1.5 OD Linearity <±0.75% from 1.5 OD to 3.0 OD (400 - 750nm) <±0.7% and ±0.005 OD from 0.1 OD to 2 OD (340 - 399nm) Reproducibility <±0.3% at 1 OD, <±0.5% at 2 OD Reading speed 96 well 18 seconds Temperature control 4°C above ambient up to 45°C in 1°C steps) - Zenyth 340rt only Shaking 3 speed Microplates 6- 384 well plates 340, 405, 450,492, 620nm (standard). Filter wheel for up to 8 filters. Additional filters Filters optional accessories. Power requirement 1100 - 240V (autosensing), 50/60Hz Serial interface 9-pin (RS232) for remote control and data transfer Self control Auto-calibration, auto lamp-adjustment, status report Dimensions (wxhxd) 46.8cm x 25.1cm x 40.8cm (18.4 x 9.9 x 16.0 inches) Weight 15Kg (33lb) ORDERING INFORMATION: GF 25 100 01
Biochrom Anthos Zenyth 340r includes: Dustcover, Spare Fuses, Power Cable, Serial Cable, ADAP Basic Standard Software and User Manual on CD
GF 25 300 01
Biochrom Anthos Zenyth 340rt includes: Dustcover, Spare Fuses, Power Cable, Serial Cable, ADAP Basic Standard Software and User Manual on CD
B0 32082
ADAP Plus Software
B0 32083
ADAP Expert Software
217142066 217145674
info@ert.pt www.ERT.pt Rua Soeiro Pereira Gomes, Lote 5, Cave Drt. P2845-387 Amora Portugal
Distributors Worldwide
Biochrom is a Harvard Bioscience Company
Biochrom Ltd
Tel: +44 1223 423723 Fax: +44 1223 420164 Email: enquiries@biochrom.co.uk www.biochrom.co.uk
Biochrom US
Tel: 877- BIO-CHROM (877-246-2476) (Toll free) Tel: 508-893-8999 Fax: 508-429-5732 Email: sales@biochrom-us.com www.biochrom-US.com 80-4001-07 Issue 0810