1 minute read

Valve with Integrated Pressure Regulator for medical

Oxygen

Technical Features

NEOS incorporates a low torque non rotating spindle shut off valve with an integrated ten position flow selector

• Valve with integrated pressure regulator for Medical Oxygen, MRI conditional certified up to 3 Tesla

Non-return valve with sintered bronze filter integrated in the filling port

Piston regulator

Sintered bronze smart filter in the cylinder connection

• Tested and approved in accordance with the International Standards EN ISO 10524-3, CGA E-18

Material Components

Chrome plated brass body Valve Main Sealing in Nylon Regulator Sealing in Nylon

Elastomer in EPDM

• The valve does not contain any ferrous material or steel

Ordering Information

Viproxy 1 Touch Series

Valve with Integrated Pressure Regulator for medical Oxygen Adapters and Shrouds

Fill Adapters

Short

Shrouds

Filling Adapters

Protective Shrouds

(*) Additional components and assembly tooling may be required.

Factory Repair Center in Somerset, NJ

Cavagna offers complete service and repair for VIPROXY and IVIPR valves with integrated regulators.

Factory trained and certified technicians

Complete diagnostic capabilities

All repairs made using OEM component parts

Warranty and non-warranty repairs offered to include: o Replacing protective guard o Gauge replacement (analog or digital) o Battery replacement o Hose barb replacement o Outlet cap replacement o Inlet O-ring replacement o Inner hand wheel replacement o Hand wheel kit o Control group replacement (digital VIPOXY only) o Circuit board replacement (digital VIPROXY only) o All factory repairs performed in an ISO class clean room o Installation of new valves in customer owned cylinders

• 212 square foot cleanroom

• Two 24-hour HEPA filters

Meets ISO 14644-1 requirements for ISO class 8 cleanroom

Contact our New Jersey office to arrange the return of items in need of repair. Parts and Labor pricing available upon request.