Veterinary Division N7 W22025 Johnson Drive Waukesha, WI 53186-1856 USA Toll Free: +1 888-745-6562 Tel: +1 262-513-8500 Fax: +1 262-513-9069 www.surgivet.com

Routine Maintenance Chart-LDS ITEM Pressure Test Patient circuits, hoses, Rebreathing bags

Machine surfaces CO2 absorber canister

Inhalation/Exhalation Valves Pop off valve

Scavenging system Gaskets(absorber and Valve)

Flutter discs

ACTION See Daily Machine Checklist Clean with dilute chlorhexidine type solution, rinse, Dry completely Inspect for damage Clean with soft cloth and mild soap/water Clean with mild soap/water and dry completely

Daily As needed

Inspect for wear/ damage

Daily

Change when 2/3 blue Visually inspect

As needed Daily

Clean with mild soap/water, dry completely Ensure smooth opening/closing

Daily

Can be cleaned with alcohol or replaced contact Clinical Support for assistance Check for proper operation, check hose Stack Visually inspect for wear, clean with mild Soap/water

INTERVAL Daily Daily

As needed Daily Weekly

Replace (Part# 32380B2-absorber, 32380B1- Annually valves) Visually inspect for damage or wear Weekly Clean with mild soap/ water, dry completely

**Please note it is recommended to pressure test and visually inspect the anesthesia machine on a daily basis. The intervals listed above are approximate. For further assistance please contact Clinical Support at 1-262-513-8500.