Routine Maintenance Chart-LDS ITEM Pressure Test Patient circuits, hoses, Rebreathing bags
Machine surfaces CO2 absorber canister
Inhalation/Exhalation Valves Pop off valve
Scavenging system Gaskets(absorber and Valve)
Flutter discs
ACTION See Daily Machine Checklist Clean with dilute chlorhexidine type solution, rinse, Dry completely Inspect for damage Clean with soft cloth and mild soap/water Clean with mild soap/water and dry completely
Daily As needed
Inspect for wear/ damage
Daily
Change when 2/3 blue Visually inspect
As needed Daily
Clean with mild soap/water, dry completely Ensure smooth opening/closing
Daily
Can be cleaned with alcohol or replaced contact Clinical Support for assistance Check for proper operation, check hose Stack Visually inspect for wear, clean with mild Soap/water
INTERVAL Daily Daily
As needed Daily Weekly
Replace (Part# 32380B2-absorber, 32380B1- Annually valves) Visually inspect for damage or wear Weekly Clean with mild soap/ water, dry completely
**Please note it is recommended to pressure test and visually inspect the anesthesia machine on a daily basis. The intervals listed above are approximate. For further assistance please contact Clinical Support at 1-262-513-8500.