Scatter Radiation Protection by Smart Armor

Page 1

Scatter Radiation Protection

Shields & Drapes for Interventional Procedures

KIARMOR RADIATION PROTECTION THE ONLY BI-LAYER LEAD-FREE CORE MATERIAL WHICH PASSES DIN (6875-1), ASTM (F3094-14) AND IEC (61331-1-2014) STANDARDS FROM 50 TO 150 KVP.


Scatter Radiation “The scattered radiation from the patient comprises the main source of radiation dose to staff.”

Direct Beam Radiation

VS

High energy radiation penetrates and passes through the patient.

Scatter Radiation Low energy radiation interacts with body tissue then scatters in different directions.

High Energy Absorbed

Low Energy Scattered

Scatter Radiation Absorbed Dose Effects on Staff

Photoelectric Effect

Direct Beam High Energy Radiation Direct Beam Low Energy Radiation

Compton Effect

male Patient

2

portrait

Passes through patient for imaging

Low Energy Scatter Radiation Absorbed by Staff

The energy absorbed from ionizing radiation can cause cataracts, tumors, skin dermatitus, hair-loss and germ cell mutations. High Risk (WT = 0.12)

Moderate Risk (WT = 0.05)

Low risk (WT = 0.01)


Scatter Armor with KIARMOR Radiation Protection against High Energy Direct Beam & Low Energy Scatter Radiation Lead-Free Bi-Layer Homogenous Material

BI-LAYER LEAD-FREE

Sb

Antimony

51 Low atomic weight absorbs low energy

Bi

Bismuth

83 High atomic weight absorbs high energy

Two individual layers within one homogenous material: #1 Lower Atomic Weight Metal (Antimony) Placed close to the radiation source. #2 Higher Atomic Weight Metal (Bismuth) Placed closest to the user.

Certified to DIN, IEC & ASTM Standards KIARMOR is the only lead-free material that passes all 3 universally recognized testing procedures: ASTM F309414, IEC 61331-1:2014, DIN 6857-1

DIN 6857-1

IEC 61331-1:2014

1040 Avenida Acaso Camarillo, CA 93012, USA

ASTM F3094-14

Ph: 805.987.5255 Fx: 805.482.8424

info@infabcorp.com www.infabcorp.com

3


Scatter Armor Shields & Drapes Scatter Armor Multipurpose Shield Transradial Access: Left and Right Radial, Shunt, Biliary Drainage, Spinal, Fibroid Embolization

Scatter Armor Multipurpose Shield with Fenestration Needle Biopsy or Puncture Procedures.

Scatter Armor Angiography Shield

Subclavian Access: Bi-Ventricular Pacemaker Implants, ICS procedures, and Bi-Ventricular Pacing.

Transfemoral Access: Angiography, Coronary Catheterization

Scatter Protection

One-time Use

Scatter Radiation Testing

Shields provide scatter radiation protection for radial and many other types of procedures. KIARMOR Bi-Layer is the most protective core material against scatter radiation.

Scatter Armor Shields are lead-free and disposable.

According to ASTM Test Method F3094-14.

Scatter Armor Shields utilize the highest absorbency covering material in the industry, critical in heavy fluid situations. Absorbent Rate Weight Absorbent Capacity

Sterile

Sample Information

50 kV

70 kV

90 kV

110 kV

Scatter Armor Shield

98.7%

96.1%W

92.0%

87.8%

Packaged individually in sterile packets.

Absorbent

1.3 sec 2.36 osy / 80 gsm 686% 16.2osy / 549 gsm

GSM - Grams per square meter, OSY - Ounce per square yard

4

Scatter Armor EP Shield

Flexible Scatter Armor Shields easily mold to the patients body for ease of use and maximum protection

Positioning Scatter Armor Shields are placed with adhesive, but can be easily re-positioned.

Direct Beam Attenuation Sample Information

60 kV

80 kV

100 kV

120 kV

Scatter Armor Shield

0.967

0.911

0.846

0.788


Scatter Armor Biliary Split Shield

Scatter Armor Peripheral Shield

Scatter Armor Dialysis Drape

Biopsy Procedures, Transjugular Intraheptic Portosystemic Shunt (TIPS)

Peripheral procedures where the access is through the femoral artery.

Dialysis Access and Declotting Procedures.

Ordering Scatter Armor Shields Scatter Armor Multi-Purpose Shield, Absorbent, No Fenestration Box Scatter Armor Multi-Purpose Shield, Absorbent, No Fenestration Case Scatter Armor Multi-Purpose Shield with Fenestration, Absorbent, Rectangular Fenestration Box Scatter Armor Multi-Purpose Shield with Fenestration, Absorbent, Rectangular Fenestration Case Scatter Armor EP Shield, Absorbent, Scoop Fenestration Box Scatter Armor EP Shield, Absorbent, Scoop Fenestration Case Scatter Armor Angiography Shield, Absorbent, Circular Fenestration Box Scatter Armor Angiography Shield, Absorbent, Circular Fenestration Case Scatter Armor Biliary Split Shield, Absorbent, Center Slit Fenestration Box Scatter Armor Biliary Split Shield, Absorbent, Center Slit Fenestration Case Scatter Armor Dialysis Drapes Box Scatter Armor Dialysis Drapes Case Scatter Armor Peripheral Shield, Absorbent, No Fenestration Box Scatter Armor Peripheral Shield, Absorbent, No Fenestration Case

Part Number SA-MPS SA-MPS-CS SA-MPF SA-MPF-CS SA-EPS SA-EPS-CS SA-AGS SA-AGS-CS SA-BSS SA-BSS-CS SA-DDD SA-DDD-CS SA-PER SA-PER-CS

Unit Box of 15 Case of 45 Box of 15 Case of 45 Box of 15 Case of 45 Box of 15 Case of 45 Box of 15 Case of 45 Box of 10 Case of 30 Box of 10 Case of 30

1040 Avenida Acaso Camarillo, CA 93012, USA

Ph: 805.987.5255 Fx: 805.482.8424

info@infabcorp.com www.infabcorp.com

5


TRANSRADIAL ACCESS, LEFT AND RIGHT RADIAL, SHUNT, BILIARY DRAINAGE, KYPHOPLASTY, PVE, ANGIOGRAPHY, TAVI/TAVR, FIBROID EMBOLIZATION

SA-MPS SA-MPS-CS

Scatter Armor

Multipurpose Shield Model Number

Description

SA-MPS

Scatter Armor Multipurpose Shield, Absorbent, No Fenestration Box of 15

SA-MPS-CS

Scatter Armor Multipurpose Shield, Absorbent, No Fenestration Case of 45

Dimensions

12” x 17”

Attenuation

Greater than 92% at 90 kVp

Protection

Greater than 0.25mm LE

Core Material

KIARMOR Bi-Layer Lead-Free

17”

12”

6


SCATTER ARMOR MULTIPURPOSE SHIELD

Multipurpose Shield Placement Place Scatter Armor Multipurpose Shield on top of surgical drape

3”

BACK Adhesive on back with ability to reposition.

7”

1040 Avenida Acaso Camarillo, CA 93012, USA

Ph: 805.987.5255 Fx: 805.482.8424

info@infabcorp.com www.infabcorp.com

7


NEEDLE BIOPSY OR PUNCTURE PROCEDURES

SA-MPF SA-MPF-CS

Scatter Armor

Multipurpose Shield with Fenestration 3”

2”

Model Number

Description

SA-MPF

Scatter Armor Multipurpose Shield, Absorbent, Rectangular Fenestration Box of 15

SA-MPF-CS

Scatter Armor Multipurpose Shield, Absorbent, Rectangular Fenestration Case of 45

Dimensions

12” x 17” Fenestration: 3.5″ x 2″

Attenuation

Greater than 92% at 90 kVp

Protection

Greater than 0.25mm LE

Core Material

KIARMOR Bi-Layer Lead-Free

17”

12”

8


SCATTER ARMOR MULTIPURPOSE SHIELD WITH FENESTRATION

Multipurpose Shield with Fenestration Placement Place Scatter Armor Multipurpose Shield with Fenestration on top of surgical drape

7”

BACK 1½” Adhesive on back with ability to reposition.

1040 Avenida Acaso Camarillo, CA 93012, USA

Ph: 805.987.5255 Fx: 805.482.8424

info@infabcorp.com www.infabcorp.com

9


SUBCLAVIAN ACCESS

SA-EPS SA-EPS-CS

Bi-Ventricular Pacemaker Implants, ICS procedures and Bi-Ventricular Pacing.

Scatter Armor

EP Shield

Model Number

Description

SA-EPS

Scatter Armor EP Shield, Absorbent, Scoop Fenestration Box of 15

SA-EPS-CS

Scatter Armor EP Shield, Absorbent, Scoop Fenestration Case of 45

Dimensions

12” x 17” Fenesration: 3.5″ x 1.5″

Attenuation

Greater than 92% at 90 kVp

Protection

Greater than 0.25mm LE

Core Material

KIARMOR Bi-Layer Lead-Free

17”

12”

10


SCATTER ARMOR EP SHIELD

EP Shield Placement Place Scatter Armor EP Shield on top of surgical drape

2” BACK Adhesive on back with ability to reposition.

8”

1040 Avenida Acaso Camarillo, CA 93012, USA

Ph: 805.987.5255 Fx: 805.482.8424

info@infabcorp.com www.infabcorp.com

11


ANGIOGRAPHY, CORONARY CATHETERIZATION, UFE, PELVIC SURGERY AND TACE.

SA-AGS SA-AGS-CS

Scatter Armor

Angiography Shield

Model Number

Description

SA-AGS

Scatter Armor Angiography Shield, Absorbent, Circular Fenestration Box of 15

SA-AGS-CS

Scatter Armor Angiography Shield, Absorbent, Circular Fenestration Case of 45

Dimensions

12” x 17” Circle Fenestration: 1.5”

Attenuation

Greater than 92% at 90 kVp

Protection

Greater than 0.25mm LE

Core Material

KIARMOR Bi-Layer Lead-Free

17”

3”

12”

12


SCATTER ARMOR ANGIOGRAPHY SHIELD

Angiography Shield Placement Place Scatter Armor Angiography Split Shield on top of surgical drape

BACK Adhesive on back with ability to reposition.

11”

1040 Avenida Acaso Camarillo, CA 93012, USA

Ph: 805.987.5255 Fx: 805.482.8424

info@infabcorp.com www.infabcorp.com

13


TRANSJUGULAR INTRAHEPTIC PORTOSYSTEMIC SHUNT (TIPS)

SA-BSS SA-BSS-CS

Scatter Armor

Biliary Split Shield

Model Number

Description

SA-BSS

Scatter Armor Biliary Split Shield, Absorbent, Center Slit Fenestration Box of 15

SA-BSS-CS

Scatter Armor Biliary Split Shield, Absorbent, Center Slit Fenestration Case of 45

Dimensions

12” x 17”

Attenuation

Greater than 92% at 90 kVp

Protection

Greater than 0.25mm LE

Core Material

KIARMOR Bi-Layer Lead-Free

17”

12”

14


SCATTER ARMOR BILIARY SPLIT SHIELD

Biliary Split Shield Placement Place Scatter Armor Biliary Split Shield on top of surgical drape

6½” BACK Adhesive on back with ability to reposition. 6½”

1040 Avenida Acaso Camarillo, CA 93012, USA

Ph: 805.987.5255 Fx: 805.482.8424

info@infabcorp.com www.infabcorp.com

15


DIALYSIS ACCESS AND DECLOTTING, FISTULAGRAM

SA-DDD SA-DDD-CS

Scatter Armor

Dialysis Drape

38.5”

36”

16

Model Number

Description

SA-DDD

Scatter Armor Dialysis Drape Box of 10

SA-DDD-CS

Scatter Armor Dialysis Drape Case of 30

Dimensions

38.5″ x 36″ Shield Size 4.5″ x 9″ Fenestration Size

Attenuation

Greater than 92% at 90 kVp

Protection

Greater than 0.25mm LE

Core Material

KIARMOR Bi-Layer Lead-Free


SCATTER ARMOR DIALYSIS DRAPE

Dialysis Drape Placement Place Scatter Armor Dialysis Drape on top of surgical drape

BACK 1” Adhesive border around fenestration with ability to reposition.

1040 Avenida Acaso Camarillo, CA 93012, USA

Ph: 805.987.5255 Fx: 805.482.8424

info@infabcorp.com www.infabcorp.com

17


PERIPHERAL PROCEDURES where access is through the femoral artery.

SA-PER SA-PER-CS

Scatter Armor

Peripheral Shield

34”

12” 18

Model Number

Description

SA-PER

Scatter Armor Peripheral Shield, Absorbent, No Fenestration Box of 10

SA-PER-CS

Scatter Armor Peripheral Shield, Absorbent, No Fenestration Case of 30

Dimensions

12” x 34”

Attenuation

Greater than 92% at 90 kVp

Protection

Greater than 0.25mm LE

Core Material

KIARMOR Bi-Layer Lead-Free


SCATTER ARMOR PERIPHERAL SHIELD

Scatter Armor Peripheral Shield Placement Place Scatter Armor Peripheral Shield on top of surgical drape

11” BACK Adhesive with ability to reposition.

1040 Avenida Acaso Camarillo, CA 93012, USA

Ph: 805.987.5255 Fx: 805.482.8424

info@infabcorp.com www.infabcorp.com

19


“The scattered radiation from the patient comprises the main source of radiation dose to staff.”

Bi-Layer Lead-Free Radiation Protection The only bi-layer lead-free core material which passes DIN (6875-1), ASTM (F3094-14) and IEC (61331-1-2014) standards from 50 to 150 kVp. TRA N SPARE N CY IN

R A D I AT I O N PROTECTION INFAB, LLC 1040 Avenida Acaso Camarillo, CA 93012, USA

Ph: 805.987.5255 Fx: 805.482.8424

info@infabcorp.com www.infabcorp.com

scatter-armor-catalog-infab-220209.pdf


Turn static files into dynamic content formats.

Create a flipbook
Issuu converts static files into: digital portfolios, online yearbooks, online catalogs, digital photo albums and more. Sign up and create your flipbook.