masterclave_comparative

Page 1

MASTERCLAVE MEDIA PREPARATORS SPECIFICATIONS Masterclave 9

Masterclave 528

Masterclave 60

Adaptability

Great flexibility

Large media volume users

Volume

1 to 8 liters

5 to 28 liters

10 to 60 liters

Stirring Speed

Constant at 45 rpm

May be adjusted between 40 and 80 rpm

Constant at 40 rpm

Timer

Via microprocessor, set from 1 to 60 minutes, + /‐ Via microprocessor, set from 1 to 99 minutes, + /‐ 1 second for any cycle stage 1 second for any cycle stage

P Power supply l

230V, +/‐10%, 16A (50,60Hz) 230V / 10% 16A (50 60H )

Simple cycle

Sterilization : 95 to 125°C, Dispensing : 30 to 80°C

Special cycle Temperature precision Units dimensions (WxHxD) Weight Safety devices

Stage 1 : 25 to 125°C, Stage 2 : 25 to 125°C, Stage 3 : 25 to 125°C, Dispensing: from 25°C to stage 3 +/‐ 1°C in stabilized phase 50 x 43.5 x 59 cm 60 Kg

From 1 to 99 minutes

400V+/ 10% 3 phases with neutral, 16A (50,60Hz) or 230V, 400V+/‐ 10% 3 phases with neutral, 32A (50,60Hz) or 230V, 400V+/‐ 10% 3 phases with neutral, 16A (50,60Hz) or 230V, 400V+/ 10% 3 phases with neutral, 32A (50,60Hz) or 230V, +/‐10% 3 phases without neutral, 20A (50,60Hz) +/‐10% 3 phases without neutral, 32A (50,60Hz) Sterilization : 95 to 125°C, Dispensing : 25 to 80°C

Sterilization : 95 to 125°C, Dispensing : 25 to 79°C

Stage 1 : 25 to 125°C, Stage 2 : 25 to 125°C, Stage 3 : 25 to 125°C, Dispensing: 25 to 80°C +/‐ 1°C in stable phase 76 x 110 x 58 cm 91 Kg 1 safety valve on lid ( set at 1,7 bars) 1 safety valve on double skin (set at 3 bars)

+/‐ 0.5°C in stabilised phases 100 x 125 x 70 cm 170 Kg 2 solenoid valves for water flow, 1 decompression valve, 1 valve for steam condenser