Application Notes VT PLUS HF Ventilator Compensation Mode Settings Correction Mode BTPS BTPS BTPS ATP ATP ATP ATP ATP
Manufacturer
Model
Viasys/Bird Viasys/Bird Viasys/Bird BMD BMD BMD Breas Breas
Avion Gold T-Bird Legacy IC2A MVP10 Crossvent 4 PV-403, 501 IMP-2
Draeger
Evita 2 and 4
BTPS (or NTPD)
Draeger
Babylog 8000
BTPS (or NTPD)
Hamilton Hamilton Infrasonics
Galileo 2 Veolar 3 Adult Star E100i & E100m
ATP ATP STPD 21째
Newport
Breeze E150
ATP
Puritan Bennett Puritan Bennett Puritan Bennett Puritan Bennett Pulmonetics Maquet (formerly Siemens) Maquet (formerly Siemens) Maquet (formerly Siemens)
7200 760 840 MA-I LTV Series
BTPS BTPS BTPS BTPS BTPS
Servo i
STPD 0째
300C
STPD 21째
900C
ATP
Newport
Comment
Not available from manufacturer, assumed Not available from manufacturer, assumed Uses BTPS in normal mode but both in Service mode, NTPD for high pressure barometer (use STPD21 in this case) Uses BTPS in normal mode but both in Service mode, NTPD for high pressure barometer (use STPD21 in this case)
ATP Use BTPS when connecting after the water trap and heat up
Setting of BTPS (on the VT PLUS HF) will give higher readings on flow (and tidal volumes) than ATP, STP21 and STP0 (in order).
Fluke Biomedical Corporation 5200 Convair Drive, Carson City, NV 89706, USA Tel: 775.883.3400, Toll Free: 800.648.7952, Email: sales@flukebiomedical.com www.flukebiomedical.com
Application Notes Definitions: STPD: Standard Temperature and Pressure, Dry (0ยบC or 21ยบC) BTPS: Body Temperature and Pressure, Saturated (think of warm, moist exhaled air) ATP: Ambient Temperature and Pressure NTPD: Normal Temperature and Pressure, Dry (see STPD at 21ยบC)
Fluke Biomedical Corporation 5200 Convair Drive, Carson City, NV 89706, USA Tel: 775.883.3400, Toll Free: 800.648.7952, Email: sales@flukebiomedical.com www.flukebiomedical.com