Case report of the fracture of a zygoma implant

Page 1

Fracture of zygoma implant after 8 years of function Luc@Vrielinck.be


Situation • Patient whishes: Fixed prosthesis in upper jaw and lower jaw • Diagnostic: – UJ: Dorsal bone atrophy – LJ: Sufficient bone volume – Bruxist (was not known before treatment)

• Treatment plan: – LJ: 6 implants, delayed loading, implant supported bridge – UJ: 4 implants, 2 long implants with pterygoid anchorage, delayed loading, implant supported bridge


Chronology • • • •

03/01/2002: 09/07/2002: 17/11/2008: 22/11/2010:

Implant placement UJ and LJ Second stage FU 6 years FU 8 years


Follow-up 17/11/2008 6 years follow-up after several teeth fractures and 1x bridge loosening


03/01/2002: Implant surgery UJ and LJ


27/08/2002: Try-in


27/08/2002: Try-in


17/11/2008: follow up 6 years


Follow-up 8 years 22/11/2010 Bridge lossening: fracture of 3 implants


22/11/2010: FU 8 years


22/11/2010: FU 8 years


22/11/2010: FU 8 years


22/11/2010: FU 8 years


22/11/2010: FU 8 years


22/11/2010: FU 8 years


22/11/2010: FU 8 years


22/11/2010: FU 8 years


SEM


Fatigue fracture of the zygoma implant


Fatigue fracture of the zygoma implant


Fatigue fracture of the zygoma implant


Causes of implantfractures • • • •

Bruxisme 2 piece bridge, no cross-arch stabilization Antagonistic implant supported bridge Direct implant fixation (no abutments used) = a mistake