/2d35dd7694a32f7d7c73df6d83211ffc

Page 1

Operating Manual Classic Physician Beam Scale with height rod M20313

ADE GmbH & Co. Hammer Steindamm 27 – 29 D – 22089 Hamburg Germany Fax E-mail:

+49(40)432776-0 +49(40)432776-10 info@ade-hamburg.de


Your ADE Mechanical sliding weight Column scale is a class III approved scale which has been manufactured for medical use in hospitals, by doctors or medical specialists. The scale is a precision instrument. The weighing result is used as an essential basis for medical application decided by doctors or medical personnel in hospitals, nursing homes, rehabilitation- and dialysis centres. Make sure disabled patients are only weighed by trained personnel and are never left unattended. In order to achieve best possible results and for safety reasons kindly read and follow the Operating Manual in all respects. Installation Check carton contents: -

Column with head part

-

Base/platform

-

Height rod

-

Pair of wheels (incl. 2 cross-slotted crews)

-

Mounting material: ( 6 x cross-slotted screw, 2 x hexagonal screw, 1 wrench )

-

Operating Manual

Assembling The assembling of the scale is easily done as all parts and components are pre-assembled during manufacturing: -

Put the base on the floor.

-

Make sure the column brackets are not bent, otherwise straighten them.

-

Slip the column into the brackets.

-

Keep the column upright to secure the smooth entering into the column brackets.

-

Connect the column and the column brackets by using two of the cross-slotted screws from the front side and four of them from the backside. Make sure the screws are pulled tight carefully.

-

Put the scale on the floor, column in horizontal position.

-

Lift up the lever system and push it.

-

Hook the transmission rod in the extension hole on top of the lever pushing it towards the column for easier insertion.

Attention: The transmission rod must move freely in the head. -

Fix the pair of wheels by using the attached 2 cross-slotted screws.

Assembling of the scale is completed. Attach the height rod: -

After having unpacked the height rod insert the 2 hex screws into the holes in front of the column. Tighten them carefully by using the wrench. Now place the two brackets of the height rod on the pre installed hex screws and pull them tight by using the wrench. Make sure not to over tighten them.

Take the scale to its working area and put it on an even ground in upright position. 1

050100-Rev000-M20313


Zero adjustment: Before first use of the scale and to make sure that the scale indicates the correct weight it has to be set to ZERO manually. -

First step on and off the scale for several times to make sure that all parts come back into the right position after transport.

-

For zero setting procedure the platform must remain empty.

-

Set both sliding weights on the beam (head part) into zero position.

-

The beam has to move freely equilibrating very close to even.

-

Turn the equilibrating screw slowly to the right respectively to the left to bring the beam into the exact even position

-

The scale is now ready to be used.

Attention: A periodically zero setting according to the previous instructions should be done. This applies especially if the scale is used in a different place, on a different surface. Use of the scale where approved scales are imperative The scale corresponds to standards and requirements of the European Directive 90/384/EEC. Quality check ADE products correspond to the applicable EC Directives and European standards to manufacture and distribute world-wide weighing and measuring equipment as per the latest technical standard offering a long lifetime and continuous reliability. This symbol stands for fulfillment of the European Directive 93/42 EEC for medical devices and declares the conformity as per this Directive.

Due to the certification as per ISO 9001/2000 the high standards for Research & Development, Manufacturing, Sales and Service requirements are fulfilled. A copy of our ISO Certificate is available on request.

Approved scales class III respecting the standards of the Directive 90/384 EEC for non automatic scales are carrying this symbol. Scales carrying this symbol in combination with class III sign are calibrated in accordance with the high precision standards laid down in EC Directive 90/384 EEC. Products carrying this symbol meet the requirements of: • Directive for non automatic scales for medical use 90/384 EEC • Applicable EC Directive governing the electromagnetic compatibility (89/336 EEC). • Directive 73/23 EEC for low voltage regulations

050100-Rev000-M20313

2


Weighing -

Step on the scale.

-

Slide the larger counterweight to the estimated weight.

-

Now use the small sliding weight to reach the correct result.

-

Wait until the beam comes close to a standstill in even position.

-

Even if the head is still moving you can read the correct result +- 100 g.

-

Total weight is the addition of both results (smaller and lager counterweight).

Measuring –

Before measuring make sure the measuring tongue is placed in a horizontal position and pull the height rod up.

Place the person on the scale and place the measuring tongue on top of the head.

If the person is shorter than 101.5 cm push the latch to the right and lower the measuring tongue until it reaches the person’s head.

The reading is done either on the inner or outer height rod.

If the measuring tongue itself points to the outer height rod the results may be taken directly.

If the measuring tongue points to the inner height rod the person’s height should be read from the top of the outer height rod. The word READ is helping you to find the right reading point.

Before to leave the scale hold the measuring tongue up and wait until the person left the platform.

Take the measuring tongue in a vertical position and put the height rod down if no measuring is required. Cleaning and maintenance

Security -

Your ADE Column scale is a precision instrument. Handle with care. Do not drop, jump on or let heavy objects fall onto it.

-

When using the scale always precisely respect the rules and advices of this Operating Manual.

-

Make sure no fluids are penetrating the scale.

Cleaning For cleaning just use a damp cloth or common disinfection fluids. Do not use any aggressive cleaning fluids.

3

050100-Rev000-M20313


Warranty – Liability ADE will repair or replace this product within a period of 2 years from date of delivery applying to defects occurred due to poor material or workmanship (presentation of invoice is necessary). All moveable parts – batteries, cables, mains units, rechargeable batteries etc. – are excluded. Warranty does not include natural wear, damages caused by incorrect or negligent handling, overuse or unsuitable or improper storage. The cost of transport will be borne by the customer should the equipment be located anywhere other then the customer’s premises. In the event of transport damage, claims under warranty can be accepted subject to using the complete original packaging and the equipment secured the way it was packed and secured originally. All the packaging should therefore be retained. Any warranty claim will be rejected if the equipment is opened by persons not authorized expressively by ADE to do so. Customers abroad are kindly requested to contact their local dealer for any warranty claim, customer service and spare parts. Technical data •

Max. Capacity:

200 kg

Graduation:

100 g

Platform size:

375 x 275 mm

Dimensions:

280 x 1470 x 560 mm

Height rod:

Measuring range:

750 – 2000 mm

Graduation:

5 mm

Total Weight: approx.

15 kg

Supplier:

050100-Rev000-M20313

ADE GmbH & Co. Hammer Steindamm 27 – 29 D – 22089 Hamburg Germany

4


D

S ... entspricht dem in der Bescheinigung über die Bauartzulassung beschriebenen Baumuster. Die Waage erfüllt die geltenden Anforderungen folgender Richtlinien: 90/384/EWG über nichtselbsttätige Waagen, 93/42/EWG über Medizinprodukte und die Europäische Norm DIN EN 45501.

DK

... motsvarar beskrivningen enligt mönstertillstándet. Vágen uppfyller gällande krav i direktiven angáende icke automatiska vágar: (90/384/EWG), 93/42/EWG över medicinska produkter samt den europeiska normen DIN EN 45501.

N

... svarer til det typemonster, som er beskrevel i attesten over typegodkendelsen. Vaegten opfylder de gaeldende krav fra folgende direktiver: 90/384/EOF om ikke-automatiske vaegte, 93/42/EOF om medicinprodukter og den eurpaeiske standard DIN EN 45501.

GB

... er i samsvarmed monsteret som er beskrevet i sertifikatet over konstruksjonstype-godkjennelsen. Vekten oppfyllerde gjelende kravene i folgende direktiver: 90/384/EEC om ikke-automatiske vekter, 93/42/EEC om medisinske proukter og den europeiske standarden DIN EN 45501.

… corresponds to the type described in the certificate in respect of type approval permit. The scales conform to the applicable requirements of the following directives: 90/384/EEC on non-automatic scales, 93/42/EEC on medical products and European standard DIN EN 45501.

FIN

… conforme au modèle type décrit dans le certificat d’homologation. Cette balance est conforme aux directives et normes suivantes : directive 90/384/CEE concernant l’harmonisation des Etats membres relative aux instruments de pesage à fonctionnement non automatique ; directive 93/42/CEE sur les produits et appareils médicaux ; norme européenne DIN EN 45501.

NL

F

I

... on tyyppihyväksyntätodistuksessa kuvatun raken nemallin mukainen. Vaaka täyttää seuraavien direktiivien voimassa olevat vaatimukset: 90/384/ETY ei-automaattisista vaaoista, 93/42/ETY lääketieteellisistä tuotteista ja Euro-normi DIN EN 45501.

... komt oveeen met de in de verklaring van over de modelkeuring beschreven constructie. De weegschaal voldoet aan de geldende eisen van de volgende richtlijnen: 90/364/EEG inzake niet-automatische weegschalen 93/42/EEG inzake medische hulpmiddelen en de Europese norm DIN EN 45501.

P … è conforme al campione descritto nel certificato di omologazione costruttiva. La bilancia risponde ale vigenti esigenze poste dalle seguenti direttive: 90/384/CEE sulle bilanci non automatiche, 93/42/CEE sui prodotti medicali e la norma europea DIN EN 45501.

RUS/UA/BY СертиФикат качества ISO 9001

E … corresponde a la muestra descrita en el certificado sobre el permiso de construcción. La báscula cumple las exigencias de las siguientes directrices : 90/384/CEE sobre básculas no automáticas 93/42/CEE sobre productos médicos y la norma europea DIN EN 45501.

ADE GmbH & Co. Hammer Steindamm 27 – 29, 22089 Hamburg

Germany

5

… corresponde ao padräo de construcäo descrito no certificado de homologacäo de tipo de construcäo. A balanca satisfaz os requistios aplic#aveis da directiva sobre balancas näo automáticas 90/384/CEE e 93/42/CEE sobre artigos médicos do Conselho, bem come da norma Europeia DIN EN 45501.

Konformitätserklärung Overensstemmelsesttest Declaration of conformity Certificat de conformité Declaration de conformidad Dichiarazione di conformità Försäkran om överensstämmelse Konformitetserklaering vaatimuksenmukaisuusvakuutus Verklaring van overeenkornst Declaracao de conformidade Δηλωση Σuμβατστητας Die nichtselbsttätige Personenwaage The non-automatic personal scales Le pèse-personnes non automatique La bilancia pesapersone nn automatica La báscula no automàtica pesapersona Den ikke-automatiske personvaegr Den icke-automatiske personvekten Ei-automaattinen henkilövaaka De niet-automatische personenweegschaal A balanca näo automática para pessoas Η μη αντοματη ζνγαρια ατομων

ADE M20313

GR ... ανταηοκριεται στο καταοκενατικο προτθπο πον περιγραφεται στην εγκριτκη βεβαιωση τον κατασκεναοτικον τροπον. Η ζνγαρια εκπληρωει τις προνποΟεσεις των ακολονΘων Οδπγιων: 90/384/ΕΟΚ περι μη αντοματων ζνγαριων, 93/42/ΕΟΚ περι ιατρικων προιοντων, και την Ενρωπαικη ΠροδιαγραΦη DIN EN 45501.

Hamburg, Januar 2004 Andreas Schmitt Geschäftsführer

℡ +49(40)432 776 - 0 +49(40)432 776 - 10 info@ade-hamburg.de

050100-Rev000-M20313


Turn static files into dynamic content formats.

Create a flipbook
Issuu converts static files into: digital portfolios, online yearbooks, online catalogs, digital photo albums and more. Sign up and create your flipbook.