Micro Vessel Occlusion Clamps

Page 1


MICRO VESSEL OCCLUSION CLAMPS

● Gold standard micro clamps for small vessel occlusion

● Light & compact design

● Precise pressure calibration

● Five clamp sizes for vessels 0.4mm to 5mm Ø

● Option of matt or black surface

● Arterial or venous clamp options

● Smooth slide bar for approximator clamps

● Unique gripping surface

ACLAND® MICRO VESSEL CLAMPS 'The

“V” “A”

● Light and compact design

● Extremely durable

● A unique gripping surface

● Matt or black finish

● Smooth sliding bar on approximator clamps

Ordering details:

“A”

or “V” Pattern?

The “V” pattern is the all-purpose clamp, used on all veins, and almost all arteries. The “A” pattern, available in all sizes, is designed for exceptionally thick-walled vessels and slippery arteries. The slight incurve of the tips of the “A” pattern jaw provides extra security against slippage.

SINGLE CLAMPS - supplied in pairs

00396 V 00397 A 00398 V 00399 A 00400 V 00401 A 00325

00462 V 00461 A

V

A

V

A

Matt Black Clamp Vessel Clamping code code size size force

ACLAND® MICRO VESSEL CLAMPS

DOUBLE CLAMPS WITH FRAME - frames are used to tie sutures onto during anastomosis

Ordering details:

Matt Black Clamp Vessel Clamping code code size size force 00474 V 00473 A

- 3 00408 V 00409 A

V 00411 A 00412 V 00413 A

V 00475 A

V 00477 A

DOUBLE APPROXIMATING CLAMPS - used for end to end anastomosis on small vessels

00414 V 00415 A 00416 V 00417 A

Matt Black Clamp Vessel Clamping code code size size force 00480 V 00479 A 00482 V 00481 A

V 00419 A

V 00483 A

- 22

33

MICRO VESSEL CLAMPS

The Biover ® microvascular clamp is a high precision instrument and has been developed under the guidance of international surgeons.

● Highest quality (Polycarbonate & Stainless Steel)

● Single or double clamps sterile, ready for use

● Four sizes available for veins & three sizes for arterys

● Latex Free and Radiolucent

VEIN CLAMPS

Single Vein Clamps

Double Approximating Clamps

● Exact pressure every use

● Supplied in boxes of 10

● Each box easily identified by a different colour

ARTERIAL CLAMPS

Single Artery Clamps Double Approximating Clamps

TKFV -1 TKFV - 2

Clamping Force 15g/mm²

Clamping Force 15g/mm²

For vessels between 0.2mm to 1mm Ø

TKSV -1 TKSV - 2

Clamping Force

20g/mm²

Clamping Force 20g/mm²

For vessels less than 1mm Ø

TKMV -1 TKMV - 2

Clamping Force

30g/mm²

Clamping Force 30g/mm²

For vessels between 1mm to 2mm Ø

TKLV -1 TKLV - 2

Clamping Force

40g/mm²

Clamping Force 40g/mm²

For vessels over 2mm Ø

TKS -1 TKS- 2

Clamping Force 40g/mm²

Clamping Force 40g/mm²

For vessels less than 1mm Ø

TKM -1 TKM- 2

Clamping Force 60g/mm²

Clamping Force 60g/mm²

For vessels between 1mm to 2mm Ø

TKL- 1

Clamping Force 120g/mm²

TKL- 2

Clamping Force 120g/mm²

For vessels between 2mm to 4mm Ø

10 clamps per box

ACLAND® MICRO VESSEL CLAMPS

Acland Clamp Applicator Forceps

● 14cm/7" length

CAF - 4 Forceps for smaller sizes of the Acland Clamps: B -1 to B- 3

CAF - 5 Forceps for larger sizes of the Acland Clamps: HD- S & RD- S

Acland Clamp Container

MV4400- 00

Compartments to store each clamp size

Clamp Box

● With perforated sliding lid, holds a selection of clamp sizes safely

CB -1

PRESSURE TESTER GAUGE

ACLAND CLAMP PRESSURE TESTER GAUGE

Ensuring your Acland Micro Vessel Clamps are applying the correct pressure onto delicate vessels is an essential consideration for successful anastomosis.

Each size of vessel clamp has its own validated pressure parameters carefully calibrated to the vessel size the clamp is designed to be used on.

This pressure can often vary depending on the age of the clamp and the amount of use.

Vessels can suffer damage if the clamp is applying too much pressure.

Equally if not sufficient pressure is applied then the vessel is not completely occluded resulting in a difficult anastomosis of the vessel with the vessel leaking blood obscuring the operating field.

An annual audit of Acland Micro clamps is recommended to test the pressures thus preventing incidents with non-conforming clamps considered good practice for Microsurgery sets.

Clamp is applied between two sensors on the Clamp Gauge where the applied pressure/force can then be accurately calculated.