● Gold standard micro clamps for small vessel occlusion
● Light & compact design
● Precise pressure calibration
● Five clamp sizes for vessels 0.4mm to 5mm Ø
● Option of matt or black surface
● Arterial or venous clamp options
● Smooth slide bar for approximator clamps
● Unique gripping surface
ACLAND® MICRO VESSEL CLAMPS 'The
“V” “A”
● Light and compact design
● Extremely durable
● A unique gripping surface
● Matt or black finish
● Smooth sliding bar on approximator clamps
Ordering details:
“A”
or “V” Pattern?
The “V” pattern is the all-purpose clamp, used on all veins, and almost all arteries. The “A” pattern, available in all sizes, is designed for exceptionally thick-walled vessels and slippery arteries. The slight incurve of the tips of the “A” pattern jaw provides extra security against slippage.
SINGLE CLAMPS - supplied in pairs
00396 V 00397 A 00398 V 00399 A 00400 V 00401 A 00325
00462 V 00461 A
V
A
V
A
Matt Black Clamp Vessel Clamping code code size size force
ACLAND® MICRO VESSEL CLAMPS
DOUBLE CLAMPS WITH FRAME - frames are used to tie sutures onto during anastomosis
Ordering details:
Matt Black Clamp Vessel Clamping code code size size force 00474 V 00473 A
- 3 00408 V 00409 A
V 00411 A 00412 V 00413 A
V 00475 A
V 00477 A
DOUBLE APPROXIMATING CLAMPS - used for end to end anastomosis on small vessels
00414 V 00415 A 00416 V 00417 A
Matt Black Clamp Vessel Clamping code code size size force 00480 V 00479 A 00482 V 00481 A
V 00419 A
V 00483 A
- 22
33
MICRO VESSEL CLAMPS
The Biover ® microvascular clamp is a high precision instrument and has been developed under the guidance of international surgeons.
● Four sizes available for veins & three sizes for arterys
● Latex Free and Radiolucent
VEIN CLAMPS
Single Vein Clamps
Double Approximating Clamps
● Exact pressure every use
● Supplied in boxes of 10
● Each box easily identified by a different colour
ARTERIAL CLAMPS
Single Artery Clamps Double Approximating Clamps
TKFV -1 TKFV - 2
Clamping Force 15g/mm²
Clamping Force 15g/mm²
For vessels between 0.2mm to 1mm Ø
TKSV -1 TKSV - 2
Clamping Force
20g/mm²
Clamping Force 20g/mm²
For vessels less than 1mm Ø
TKMV -1 TKMV - 2
Clamping Force
30g/mm²
Clamping Force 30g/mm²
For vessels between 1mm to 2mm Ø
TKLV -1 TKLV - 2
Clamping Force
40g/mm²
Clamping Force 40g/mm²
For vessels over 2mm Ø
TKS -1 TKS- 2
Clamping Force 40g/mm²
Clamping Force 40g/mm²
For vessels less than 1mm Ø
TKM -1 TKM- 2
Clamping Force 60g/mm²
Clamping Force 60g/mm²
For vessels between 1mm to 2mm Ø
TKL- 1
Clamping Force 120g/mm²
TKL- 2
Clamping Force 120g/mm²
For vessels between 2mm to 4mm Ø
10 clamps per box
ACLAND® MICRO VESSEL CLAMPS
Acland Clamp Applicator Forceps
● 14cm/7" length
CAF - 4 Forceps for smaller sizes of the Acland Clamps: B -1 to B- 3
CAF - 5 Forceps for larger sizes of the Acland Clamps: HD- S & RD- S
Acland Clamp Container
MV4400- 00
Compartments to store each clamp size
Clamp Box
● With perforated sliding lid, holds a selection of clamp sizes safely
CB -1
PRESSURE TESTER GAUGE
ACLAND CLAMP PRESSURE TESTER GAUGE
Ensuring your Acland Micro Vessel Clamps are applying the correct pressure onto delicate vessels is an essential consideration for successful anastomosis.
Each size of vessel clamp has its own validated pressure parameters carefully calibrated to the vessel size the clamp is designed to be used on.
This pressure can often vary depending on the age of the clamp and the amount of use.
Vessels can suffer damage if the clamp is applying too much pressure.
Equally if not sufficient pressure is applied then the vessel is not completely occluded resulting in a difficult anastomosis of the vessel with the vessel leaking blood obscuring the operating field.
An annual audit of Acland Micro clamps is recommended to test the pressures thus preventing incidents with non-conforming clamps considered good practice for Microsurgery sets.
Clamp is applied between two sensors on the Clamp Gauge where the applied pressure/force can then be accurately calculated.