STEP-BY-STEP SURGICAL PROCEDURE
Drilling protocol For Z5m/Z5mlb and Z5c implants
TD285 | max. 600 RPM
3.6 mm 4.0 mm 5.0 mm
RD230 | max. 800 RPM
TD230 | max. 800 RPM
DP220
For Z5m(t) implants
4.0 mm
TD170 | max. 800 RPM
CD355-1 | max. 400 RPM
TD230 | max. 800 RPM
CD455-1 | max. 400 RPM
5.0 mm
RD230 | max. 800 RPM
TD285 | max. 600 RPM
TD325 | max. 500 RPM
TD285 | max. 600 RPM
TD325 | max. 500 RPM
Drilling depth in accordance with implant length: 8 ⁄ 10 ⁄ 12 mm
Z5m-40-12
14 mm 12 mm
T360L-2 | max. 15 RPM optional for D2–D3
10 mm 8 mm 6 mm
max. 1.3 mm TD325
T400-2 | max. 15 RPM optional for D2–D3
Insertion depth
Abutment 5 mm Biological width 2.5 mm Bone level
TD375 | max. 500 RPM
TD425 | max. 500 RPM
T500-2 | max. 15 RPM optional for D2–D3
Z5m /Z5m(t) / Z5c
Drill depths for bone class D3/D4 (in mm) ø 4 mm
ø 5 mm
Implant length (in mm)
8
10
12
TD170
8
10
12
TD230
8
10
12
8
10
12
For drilling protocol bone class D1/D2 please see our surgical manual.
Taking the implant out of the package All Z-SYSTEMS implants are delivered in a stable carton. You
sterilized internal blister and the implant, as well as the IFU
will find an non-sterile outer blister pack with the plasma-
and three detachable label strips for documentation.
1
2
3
Tear at the
Remove the outer blister,
Remove the sterile
lateral perforation
and open the sealed cover
internal blister pack
4
5
Hold the internal blister
6
Open the sealed cover
near the arrow with slight pressure
7
Carefully, gently press to stabilize the white inserts
8
Press on the angle
Remove the one-piece
or ratchet adapter
implant through a sideways movement. Removing the two-piece implant is shown on the following page.
Implant insertion Specific instruments are needed due to the different design
While for the Z5m ⁄ Z5mlb and Z5mlc implants the insertion
geometries for each type of implant. These instruments are
instruments are only plugged in, for Z5c implants we provi-
indicated accordingly in the surgical tray.
de a very firm connection between implant and instrument.
Z5m
Z5mlc
Z5mlb
Z5m(t)
Z5c
RA16 | HA12 | RA24
RA16lc
RA16lb | HA12lb | RA24lb
RA16 | HA12 | RA24
Z5cRA12 | HA12 | RA16 | RA12-1 | RA16-1
Z5c a) The white implant holder is rotated in a clockwise direction, and the implant is freely accessible. b) The clamping wedge can be installed. c) Constant vertical pressure is very important in order to ease insertion of the implant.
a b
Z5m ⁄ Z5mlb ⁄ Z5mlc
Z5c
Insertion speed (RPM)
15
15
Recommended torque
35–70
25–35
Min. torque for sufficient primary stability
20
20
Max. torque
70
50
c
only with a screwed ratchet adapter
Application tip: Turn the implant slightly to the left before insertion. The thread just reaches into the pre-cut-of alveolus, and then follows the threads when turning clockwise.
Healing Z5m ⁄ mlc ⁄ mlb
Z5c
The Z5m-, Z5mlc-, Z5mlb- and Z5c implants heal
The Z5c implant is provided with sterile healing caps
transgingivally.
for after the insertion.
There are various gingival formers available for these
There are various gingival formers available
implants for shaping after healing. These can
for tissue healing.
also be used to create provisional single tooth implants for a maximum of 180 days.
Healing caps
Z5c-HC40, Z5c-HC50
Gingival former
Gingival former 2.8 mm
3.3 mm
G36, G40, G50, G50R G36-0, G40-0, G50-0, G50R-0
Z5c-G40-28/Z5c-G50-28/ Z5c-G40-33/Z5c-G50-33