Multi Function Process Calibrator

Page 1

R

MULTI FUNCTION PROCESS CALIBRATOR MODEL NO.- ACM-MFPC


R

MULTI FUNCTION PROCESS CALIBRATOR MODEL NO.- ACM-MFPC

Weiber's Multi-Function Process Calibrator is unique and ideal equipment for research and general applications in the field of measurement and analysis of electricity (electric current) in terms of - voltage current (DC); electrical resistance, frequency, thermocouple, thermo resistance, Switch meter measure, and continuity test.

Major Technical Features Ÿ

Large LCD display of screen

Ÿ

Portable and hand-held equipment

Ÿ

Easy to operate

Ÿ

White LED for backlight

Application Ÿ

Kitchen appliances

Ÿ

Washing and drying machines

Ÿ

Electrical systems

Ÿ

Electronics used in automobile

TECHNOCRACY PVT. LTD.


R

MULTI FUNCTION PROCESS CALIBRATOR MODEL NO.- ACM-MFPC

Technical Features Range

50 mV

Resolving Power

1µV

Remark

Input Impedance : about 100 M Ω

Measuring Range

5.000 mV - 55.000 mV

Accuracy

0.02 % + 10

Range

500mV

Resolving Power

10µV

Remark

Input Impedance:about 100 M Ω

Measuring Range

10.00mV - 550.00mV

Accuracy

0.02%+2

Range

5V

Resolving Power

0.1mV

Remark

Input Impedance : about 1 M O

Measuring Range

0.1000V - 5.5000V

Accuracy

0.02%+5

Range

50 V

Resolving Power

10mV

Remark

Input Impedance : about 1 M O

Measuring Range

0.00V - 55.00V

Accuracy

0.03 % + 5

TECHNOCRACY PVT. LTD.


R

MULTI FUNCTION PROCESS CALIBRATOR MODEL NO.- ACM-MFPC

DC Current Ÿ

Range

50 mA

Ÿ

Resolving Power

1µA

Ÿ

Remark

Input Impedance 10Ω

Ÿ

Measuring Range

5.000mA - 55.000mA

Ÿ

Accuracy

0.02%+2

OHM Range

500 Ω About 1 mA exciting

Resolving Power

0.01Ω

Remark

Plough voltage : about 2.5V not including the accuracy of down-lead resistance

Measuring Range

0.00Ω- 550.00Ω

Accuracy

0.02% + 0.1Ω

Range

5 KΩ About 0.1 mA exciting

Resolving Power

0.1Ω

Remark

Plough voltage about 2.5V not including the accuracy of down-lead resistance

Measuring Range

0.0000 KΩ; - 5.5000KΩ 0.02%+0.5Ω

TECHNOCRACY PVT. LTD.


R

MULTI FUNCTION PROCESS CALIBRATOR MODEL NO.- ACM-MFPC

Frequency Range

100Hz

Resolving Power

0.1Hz

Remark

input impedance: 100 k立; or more Square wave at least 1Vp-

Measuring Range Accuracy Range Resolving Power Remark Measuring Range

p 50% duty cycle 2Hz - 110.0Hz +/-2 1KHz 1Hz Input Impedance : 100 k立 or more Square wave at least 1Vp-p 50% duty cycle 0.100KHz1.100KHz

Accuracy

賊2

Range

10KHz

Resolving Power

0. 1KHz

Remark

Input Impedance : 100 k立; or more Square wave at least

Measuring Range

1Vp-p 50% duty cycle 10KHz11.0KHz

TECHNOCRACY PVT. LTD.


R

MULTI FUNCTION PROCESS CALIBRATOR MODEL NO.- ACM-MFPC

Thermocouple Range : Measuring Range Resolving Power Accuracy Remark Range : Measuring Range : Resolving Power : Accuracy : Remark : Range : Measuring Range : Resolving Power : Accuracy : Remark : Range : Measuring Range : Resolving Power : Accuracy : Remark : Range : Measuring Range : Resolving Power : Accuracy : TECHNOCRACY PVT. LTD.

R -40C -1760 C(o) 1 C(o) -4500 : 1.8 &n- 501760 : 1.5 Adopt ITS-90 thermometric scale not including the accuracy of inner temperature compensatory error S - 20ºC ----- 1760ºC 1 ºC 4500 : 1.8 - 501760 : 1.5 dopt ITS-90 thermometric scale not including the accuracy of inner temperature compensatory error K - 200ºC ---- 950ºC 0.1ºC - 2000 : 1.2 ----- 01370 : 0.8 Adopt ITS-90 thermometric scale not including the accuracy of inner temperature compensatory error E -200ºC ------ 500 C 0.1C -2000 : 0.9 01000 : 1.5 Adopt ITS-90 thermometric scale not including the accuracy of inner temperature compensatory error J - 200ºC ----- 700ºC 0.1ºC -2000 : 1.0 ---- 01200 : 0.7


R

MULTI FUNCTION PROCESS CALIBRATOR MODEL NO.- ACM-MFPC

Remark : Range : Measuring Range : Resolving Power : Accuracy : Remark :

Range : Measuring Range : Resolving Power : Accuracy : Remark : Range : Measuring Range : Resolving Power : Accuracy : Remark : Range : Measuring Range : Resolving Power : Accuracy : Remark : Range :

Adopt ITS-90 thermometric scale not including temperature compensatory error T -200C - 400C 1 ยบC -2000 : 1.0 ----- 0400 : 0.7 Adopt ITS-90 thermometric scale not including temperature compensatory error N -200ยบC ----- 1000C 0.1C -2000 : 1.5 ---- 01300 : 0.9 Adopt ITS-90 thermometric scale not including temperature compensatory error B 400C ---- 1800ยบC 1Cยบ 400800 : 2.2 8001000 : 1.8 10001800 : 1.4 Adopt ITS-90 thermometric scale not including temperature compensatory error PT 100 - 200C ---- 700C 0.1C - 2000 : 0.5 0400 : 0.7 400850 : 0.8 Adopt ITS-90 thermometric scale not including temperature compensatory error PT 1000

the accuracy of inner

the accuracy of inner

the accuracy of inner

the accuracy of inner

the accuracy of inner

TECHNOCRACY PVT. LTD.


R

MULTI FUNCTION PROCESS CALIBRATOR MODEL NO.- ACM-MFPC

TECHNOCRACY PVT. LTD.


R

MULTI FUNCTION PROCESS CALIBRATOR MODEL NO.- ACM-MFPC

Measuring Range : Resolving Power : Accuracy : Remark :

- 200C ----- 630C 0.1C - 200100 : 0.22 100300 : 0.3 300630 : 0.4 Adopt ITS-90 thermometric scale not including the accuracy of inner temperature compensatory error

Switch Meter Measure Range

1 mA (approx.) exciting

Measuring Range

CLOSE / OPEN

Continuity Test Range

500 立 About 1 mA exciting

Resolving Power

0.01 立

Measuring Range

50 立 vocalism

TECHNOCRACY PVT. LTD.


An ISO 9001 : 2208 | ISO 14001 : 2008 | ISO 13485 WHO: GMP Products | GLP Compliant Products

www.acmasindia.com | www.cleanroom-equipments.com | www.measuring-meters.com

TECHNOCRACY PVT. LTD. SALES OFFICE (INDIA) Office no. 506 & 507, 5th Floor, Plot no.- C-9, Pearls Best Height-II, Netaji Subhash Place, Pitam Pura, Delhi- 110034 Land Line No.: +91 - 11- 47619688, Fax: +91-11-47619788 E-mail.: info@acmasindia.com

SALES OFFICE (HONG KONG) Unit D 28 11/F Wing Tat Comm,Bidg 97, Bonham Strand East, Sheung Wan, Hong Kong (PRC) Tel.: 0086-13929598046 0086- 18922303099 E-mail.: hk@acmasindia.com

SHOWROOM J-284 Sector 1 Bawana Industrial Estate, Bawana, Delhi, INDIA-110039 Tel.: 0091-0-9312219738 (M) +91- 9717741167 E-mail.: info@acmasindia.com

SALES OFFICE (RUSSIA) Inmed Trade Street Ozerkovsky Embankment, Unit No 50, Straine- 1, Off- 502, Moscow, Russia E-mail.: info@acmasindia.com Tel.: 0049- 79592345 | Email: russia@acmasindia.com


Turn static files into dynamic content formats.

Create a flipbook
Issuu converts static files into: digital portfolios, online yearbooks, online catalogs, digital photo albums and more. Sign up and create your flipbook.