Operating table
SU-04 The product was made of materials that possess antimicrobial/antibacterial properties.
www.famed.com.pl
Operating table
SU-04
Head rest adjusted by the use of gas spring Head rest adjustment lever
Transport trolley movement possibility made of two sides of the operating table base
Antistatic removable mattresses made of polyurethane foam
Purpose
The SU-04 operating table has been designed to support patients during procedures in general surgery, urology, gynaecology, cardiosurgery, neurosurgery, proctology, ophthalmology, laryngology, laparoscopy, endoscopy, oncology, dentistry, plastic surgery and others. The entire offer of additional equipment is presented in the “Equipment for operating tables” catalogue.
Head rest detachment knob Longitudinal shift adjustment Back rest segment release Back rest segment adjustment lever Wire controller WS-35.0 (version I)
Base locking lever Adjustable foot
Dismantling lever of leg rest detachment
Leg rest segment angle inclination adjustment Wireless controller IR indicator Wire hand controller socket-contact Leg rest segment deflection adjustment and locking mechanism
Connection panel Transportation trolley movement limiter
Standard structure
Functions and drives
• The table is made of acid proof, stainless steel • The movable base is equipped with centrally locking wheels, controlled by foot lever • Table tops are fitted and removed by means of a transfer trolley • The antistatic, seamless mattresses are removable from the segments of the table top and resistant to disinfecting agents • The mat table surface is easy to clean and resistant to disinfecting agents • The table top is X-ray translucent enabling the use of a C arm
• Table top height adjustment, lateral tilts, Trendelenburg and reverse Trendelenburg tilts and the table top levelling are performed by means of a 24V battery powered electro-hydraulic drive system and a remote control • The back rest segment, head rest segment and one-piece leg rest segment adjustment is manually performed by locking gas springs • Divided, two parts leg rest segments are adjusted by means of gear transmission • Kidney elevator adjustment in case of a five or sixsection table top is manually performed by the mechanic gear transmission • Longitudinal shift and aside deflection movement of leg rests is manually performed
Table top automatic assemble and dismantling on the table column
Transport trolley incursion is possible from both sides of the table base (back and front). There is an additional possibility of placing the table top on the transport trolley with the head rest segment or leg rest segment toward driving direction. It ensures either patient enlighten and table top in optimal position for the treatment.
Transport trolley for table tops transport
Operating table SU-04 makes possible modern operating block arrangement. This is possible thanks to the automatic function operating table top dismantling out of operating table and automatic table top dismantling out of transport trolley • Inside of the operating table SU-04 column and inside transport trolley there is fastened inside a special sockets with additional lock mechanism safety measures against co-accidental disconnection • Proposed system makes possible patient transport for operating room and post operating room on the operating table top placed on transport trolley • Operating table electronic system in automatic way recognizes of the table top orientation and ensures compatibility selected functions for functions displayed on the hand controller
Trolley WB-01.2 is equipped with height adjustment and longitudinal shift adjustment
Standard functions and solutions
SU-04 table tops Total length: • 4-segments table top • 5-segments table top • 6-segments table top • Neurosurgery table top
Longitudinal table top shift
2025 mm 2025 mm 2160 mm 2025 mm
Total width:
Table top longitudinal shift and antysymetrical operating table column location allow C-arm use during patient treatment.
• 4-segments table top • 5-segments table top • 6-segments table top • Neurosurgery table top
550 mm 550 mm 550 mm 550 mm
Leg segment length: • 4-segments table top • 5-segments table top • 6-segments table top • Neurosurgery table top
Possibility of table top removal
X-rays permeability
Range of X-ray permeability with use of the RTG tray
615 mm 615 mm 750 mm 615 mm
Table top BL-20.0 technical data: • Total table top length (without of head rest segment) • Total table top width • Back rest segment adjustment
1735 mm 545 mm -25° to +80°
4-segments table top
5-segments table top
6-segments table top
Neurosurgery table top
Used solutions
Possibility of mattresses dismantling
Seating for orthopedic attachment
Seating for specialist head rest segment
Operating table structure allows for static charge outlet through antistatic wheels for conductive floor and / or equalization charge potential installation
Battery power
Examples of use
Operating table SU-04 execution options • Control panel placed on the table column • Additional alternative mechanic-hydraulic power system • Wireless hand controller • Wire foot controller
Operating table SU-04 equipped with OT base PS-04.11
Seating position (table top with laryngologyophthalmic head rest WS-21.5)
Gynaecological position with applying gynaecological attachment WS-28.5 and knee rests WS-05.5
Head and neck operation
Side fixed positioning (in kidneys operations). Table top with head rest segment WS-45.5
Orthopaedic attachment Attachment is made of acid proof, stainless steel. It’s fixed to table top and together with table top has the possibility of height, lateral and longitudinal adjustments. Attachment enables patient’s C arm monitoring during operation and ensures surgeon’s convenient access to the patient.
Execution options SU-04
SU-04 controlls
SU-04 alternative drive SU-04 table with alternate drive system (option) is equipped with two independently operating drive systems: electro-hydraulic and mechanical-hydraulic. The alternative mechanical-hydraulic system enables adjustment of height, lateral and longitudinal tilts of table top in case of electrohydraulic system failure. Particular functions are selected by the lever located in the table base and are performed thanks to the hydraulic foot pump.
Column control panel Panel makes possible electro-hydraulic system control of SU-04 operating table. There are the function buttons compatible to the buttons on the wire hand controller and battery charge indicator as well. Panel control is placed on the column side.
Wire foot controller WS-36 (option)
Wireless hand controller WS-37 (option)
Wire hand controller WS-35
Technical data Total table length • with 4 or 5-segments table top • with 6-segments table top
2025 mm 2160 mm
Total table top width
550 mm
Mattress width x thickness
500 x 50 mm
Table top height adjustment
790 to 1040 mm
Trendelenburg position
25°/16°
Reverse Trendelenburg position
16°/25°
Lateral tilts
20°
Back rest segment adjustment
-40° to +70°
Head rest adjustment
-20° to +50°
Kidney elevator height adjustment (5 and 6 segments table top)
160 mm
Inclination angle adjustment of undivided leg rests
-90° to +15°
Inclination angle adjustment of divided leg rests
-90° to +90°
Leg rest segment inclination angle adjustment
180°
Table top longitudinal shift
250 mm
Battery power
24 V DC
Power supply
230/127/110/100 V ~, 50/60 Hz
Power consumption
100 VA
Maximum permissible load
200 kg
Total table weight
265 kg
Famed Żywiec Sp. z o.o. ul. Fabryczna 1 34-300 Żywiec, Poland Head Office: tel.: +48 33 866 62 00 fax: +48 33 475 58 90 Export Department: tel.: +48 33 866 62 58 tel.: +48 33 866 62 65 tel.: +48 33 866 62 66 fax: +48 33 861 46 78 expo.famed@famed.com.pl www.famed.com.pl
Operating table functions executed with use of the hand controller 790 to 1040 mm
25°/16°
16°/25°
"0"
20°
Operating table functions executed manually -40° to +70°
160 mm *
250 mm * in 5 or 6-segments table top
180°
** i n 6-segments table top
ZY J ST SZA P ODUKC R
C
Famed Żywiec reserves the right to make modifications in the products in the course of technical progress. All illustrations and photos, used in this material, are for illustration and may not accurately depict the actual product.
RODUCTIO RP N NE
EWORK AM FR
The product meets the requirements of the European Directive MDD 93/42/EEC concerning safety of medical devices
A
-90° to +15°
-90° to +90° **
UNEP IE CLE A
-20° to +50°