Su 04 en

Page 1

Operating table

SU-04 The product was made of materials that possess antimicrobial/antibacterial properties.

www.famed.com.pl


Operating table

SU-04

Head rest adjusted by the use of gas spring Head rest adjustment lever

Transport trolley movement possibility made of two sides of the operating table base

Antistatic removable mattresses made of polyurethane foam

Purpose

The SU-04 operating table has been designed to support patients during procedures in general surgery, urology, gynaecology, cardiosurgery, neurosurgery, proctology, ophthalmology, laryngology, laparoscopy, endoscopy, oncology, dentistry, plastic surgery and others. The entire offer of additional equipment is presented in the “Equipment for operating tables” catalogue.

Head rest detachment knob Longitudinal shift adjustment Back rest segment release Back rest segment adjustment lever Wire controller WS-35.0 (version I)

Base locking lever Adjustable foot

Dismantling lever of leg rest detachment

Leg rest segment angle inclination adjustment Wireless controller IR indicator Wire hand controller socket-contact Leg rest segment deflection adjustment and locking mechanism

Connection panel Transportation trolley movement limiter

Standard structure

Functions and drives

• The table is made of acid proof, stainless steel • The movable base is equipped with centrally locking wheels, controlled by foot lever • Table tops are fitted and removed by means of a transfer trolley • The antistatic, seamless mattresses are removable from the segments of the table top and resistant to disinfecting agents • The mat table surface is easy to clean and resistant to disinfecting agents • The table top is X-ray translucent enabling the use of a C arm

• Table top height adjustment, lateral tilts, Trendelenburg and reverse Trendelenburg tilts and the table top levelling are performed by means of a 24V battery powered electro-hydraulic drive system and a remote control • The back rest segment, head rest segment and one-piece leg rest segment adjustment is manually performed by locking gas springs • Divided, two parts leg rest segments are adjusted by means of gear transmission • Kidney elevator adjustment in case of a five or sixsection table top is manually performed by the mechanic gear transmission • Longitudinal shift and aside deflection movement of leg rests is manually performed

Table top automatic assemble and dismantling on the table column

Transport trolley incursion is possible from both sides of the table base (back and front). There is an additional possibility of placing the table top on the transport trolley with the head rest segment or leg rest segment toward driving direction. It ensures either patient enlighten and table top in optimal position for the treatment.

Transport trolley for table tops transport

Operating table SU-04 makes possible modern operating block arrangement. This is possible thanks to the automatic function operating table top dismantling out of operating table and automatic table top dismantling out of transport trolley • Inside of the operating table SU-04 column and inside transport trolley there is fastened inside a special sockets with additional lock mechanism safety measures against co-accidental disconnection • Proposed system makes possible patient transport for operating room and post operating room on the operating table top placed on transport trolley • Operating table electronic system in automatic way recognizes of the table top orientation and ensures compatibility selected functions for functions displayed on the hand controller

Trolley WB-01.2 is equipped with height adjustment and longitudinal shift adjustment


Standard functions and solutions

SU-04 table tops Total length: • 4-segments table top • 5-segments table top • 6-segments table top • Neurosurgery table top

Longitudinal table top shift

2025 mm 2025 mm 2160 mm 2025 mm

Total width:

Table top longitudinal shift and antysymetrical operating table column location allow C-arm use during patient treatment.

• 4-segments table top • 5-segments table top • 6-segments table top • Neurosurgery table top

550 mm 550 mm 550 mm 550 mm

Leg segment length: • 4-segments table top • 5-segments table top • 6-segments table top • Neurosurgery table top

Possibility of table top removal

X-rays permeability

Range of X-ray permeability with use of the RTG tray

615 mm 615 mm 750 mm 615 mm

Table top BL-20.0 technical data: • Total table top length (without of head rest segment) • Total table top width • Back rest segment adjustment

1735 mm 545 mm -25° to +80°

4-segments table top

5-segments table top

6-segments table top

Neurosurgery table top

Used solutions

Possibility of mattresses dismantling

Seating for orthopedic attachment

Seating for specialist head rest segment

Operating table structure allows for static charge outlet through antistatic wheels for conductive floor and / or equalization charge potential installation

Battery power


Examples of use

Operating table SU-04 execution options • Control panel placed on the table column • Additional alternative mechanic-hydraulic power system • Wireless hand controller • Wire foot controller

Operating table SU-04 equipped with OT base PS-04.11

Seating position (table top with laryngologyophthalmic head rest WS-21.5)

Gynaecological position with applying gynaecological attachment WS-28.5 and knee rests WS-05.5

Head and neck operation

Side fixed positioning (in kidneys operations). Table top with head rest segment WS-45.5

Orthopaedic attachment Attachment is made of acid proof, stainless steel. It’s fixed to table top and together with table top has the possibility of height, lateral and longitudinal adjustments. Attachment enables patient’s C arm monitoring during operation and ensures surgeon’s convenient access to the patient.

Execution options SU-04

SU-04 controlls

SU-04 alternative drive SU-04 table with alternate drive system (option) is equipped with two independently operating drive systems: electro-hydraulic and mechanical-hydraulic. The alternative mechanical-hydraulic system enables adjustment of height, lateral and longitudinal tilts of table top in case of electrohydraulic system failure. Particular functions are selected by the lever located in the table base and are performed thanks to the hydraulic foot pump.

Column control panel Panel makes possible electro-hydraulic system control of SU-04 operating table. There are the function buttons compatible to the buttons on the wire hand controller and battery charge indicator as well. Panel control is placed on the column side.

Wire foot controller WS-36 (option)

Wireless hand controller WS-37 (option)

Wire hand controller WS-35


Technical data Total table length • with 4 or 5-segments table top • with 6-segments table top

2025 mm 2160 mm

Total table top width

550 mm

Mattress width x thickness

500 x 50 mm

Table top height adjustment

790 to 1040 mm

Trendelenburg position

25°/16°

Reverse Trendelenburg position

16°/25°

Lateral tilts

20°

Back rest segment adjustment

-40° to +70°

Head rest adjustment

-20° to +50°

Kidney elevator height adjustment (5 and 6 segments table top)

160 mm

Inclination angle adjustment of undivided leg rests

-90° to +15°

Inclination angle adjustment of divided leg rests

-90° to +90°

Leg rest segment inclination angle adjustment

180°

Table top longitudinal shift

250 mm

Battery power

24 V DC

Power supply

230/127/110/100 V ~, 50/60 Hz

Power consumption

100 VA

Maximum permissible load

200 kg

Total table weight

265 kg

Famed Żywiec Sp. z o.o. ul. Fabryczna 1 34-300 Żywiec, Poland Head Office: tel.: +48 33 866 62 00 fax: +48 33 475 58 90 Export Department: tel.: +48 33 866 62 58 tel.: +48 33 866 62 65 tel.: +48 33 866 62 66 fax: +48 33 861 46 78 expo.famed@famed.com.pl www.famed.com.pl

Operating table functions executed with use of the hand controller 790 to 1040 mm

25°/16°

16°/25°

"0"

20°

Operating table functions executed manually -40° to +70°

160 mm *

250 mm * in 5 or 6-segments table top

180°

** i n 6-segments table top

ZY J ST SZA P ODUKC R

C

Famed Żywiec reserves the right to make modifications in the products in the course of technical progress. All illustrations and photos, used in this material, are for illustration and may not accurately depict the actual product.

RODUCTIO RP N NE

EWORK AM FR

The product meets the requirements of the European Directive MDD 93/42/EEC concerning safety of medical devices

A

-90° to +15°

-90° to +90° **

UNEP IE CLE A

-20° to +50°


Turn static files into dynamic content formats.