PLANMECA DENTAL UNITS
Maintenance and infection control for scheduled down-time periods Before down-time period Dental unit waterlines Emptying waterlines If the dental unit will not be used for many weeks, the dental unit waterlines must be emptied to prevent biofilm formation.
1 Close the incoming water tap.
AIR
WATER
DR
1
AI
SU C I O N T
2 Flush the cup fill, bowl rinse and all instrument water-
lines, including the assistant’s syringe, until all water has run out. 80 6
2
remove the bottle and clean it.
3
After down-time period Dental unit waterlines 1 Run plenty of water through the dental unit. The easiest way to do this is to perform long instrument flushing several times. Remember to select all instruments, including the assistant’s syringe. Note! If the waterlines are empty, the water will splash in the beginning when you start flushing the instruments.
2 Clean the waterlines over night with Planmeca Planosil /
Planmeca PlanPure using the Waterline Cleaning System (WCS) or the Clean Water System (CWS) and flush the waterlines the following morning. Start the cleaning process the day before you want to take the dental unit into use. Repeat the waterline cleaning two more times for two subsequent days.
1
2
Suction lines 1 Aspire at least 0.5 litres of water through each suction tube.
1
0
160
10
3 If the dental unit is equipped with a clean-water bottle,
N
2
Surfaces 1 Wipe all surfaces.
Do not spray!
For detailed instructions, please see the user’s manual. Copyright Planmeca. Rights for changes reserved.
30025366B/04.2020/en
www.planmeca.com
Planmeca imaging units
Maintenance for scheduled down-time periods Planmeca ProX™ intraoral X-ray unit 1 Before down-time unplug the power cord.
1
2 After the down-time period verify the image quality
control check according to the requirements of local authorities, using for example Quart phantom or similar.
Planmeca ProSensor® HD intraoral sensor 1 Before down-time unplug the power cord.
1
2 After the down-time period verify the image quality
control check according to the requirements of local authorities, using for example Quart phantom or similar.
Planmeca ProScanner® 2.0 imaging plate scanner 1 Before down-time unplug the power cord.
1
2 After the down-time, if the imaging plate has been
stored for longer than one week, erase the image plate prior use.
3 Verify the image quality control check according to the requirements of local authorities, using for example Quart phantom or similar.
Planmeca ProMax® and Planmeca Viso™ extraoral X-ray units 1 Before down-time unplug the power cord.
1
2 After the down-time, if the X-ray unit has not been used for a week or more, perform tube head seasoning.
3 Verify the image quality control check according to the
Please note that country-specific directives, standards and specifications for medical products must always be considered, as well as specifications for dental practices and clinics. For detailed instructions, please see the user’s manual. Description and instructions for Planmeca QA tests can be found in the technical manuals. Copyright Planmeca. Rights for changes reserved.
30025459A/04.2020/en
requirements of local authorities, using for example Quart phantom or similar.
Planmeca PlanMill® 40 S, 30 S and 40
Milling unit maintenance for scheduled down-time periods Before down-time period Clean spindle caps and collets 1 Touch the Collet cleaning icon.
2 For PlanMill 40 select Yes and follow the cleaning
wizard’s instructions. For PlanMill 40 S and 30 S select Clean caps and collets and follow the cleaning wizard’s instructions.
3 Close the wizard after successful cleaning process.
Empty and clean fluid tank 1 Touch the Clean fluid tank icon.
2 Empty and clean the fluid tank according to the milling unit’s instructions and fill the tank with hot water only when asked to fill the tank.
3 With PlanMill 40 S and 30 S circulation system priming
starts automatically after filling the tank. With PlanMill 40 go to Support Console High Level touch Mechanical Control box Start Pump. Circulate the hot water 2-3 minutes and Stop Pump.
4 After successful circulation system priming process
open the drawer and empty and clean the tank. Rinse the filter and leave both the tank and the filter out for drying.
5 Clean the milling chamber and lid thoroughly. 6 Shut down the milling unit and switch off the main
power from the back of the unit. Leave the lid open. Shut down the PlanMill 40 Job Server by pressing the power button for 5 seconds.
After down-time period Taking the milling unit into use 1 Turn on the milling unit. 2 If Clean fluid tank icon is red, proceed to the cleaning process. Otherwise refill the fluid tank with water and cooling liquid.
3 Start milling. For detailed instructions, please see the user’s manual. Copyright Planmeca. Rights for changes reserved.