THE SOLUTION FOR MALE INCONTINENCE
ZSI 375 ARTIFICIAL URINARY SPHINCTER
ALL SIZES WITH ONE ADJUSTABLE CUFF More information at: www.zephyr-si.com ZEPHYR Surgical implants is ISO 13485 : 2004 Artificial Urinary Sphincter ZSI 375 is EC certified. LOCAL DISTRIBUTOR
ZEPHYR Surgical Implants Route des Jeunes 4 1227 Acacias (Geneva) SWITZERLAND Email: contact@zephyr-si.com / www.zephyr-si.com
Flyer, English version V2 September 2014
Swiss Made
ADJUSTABLE CUFF FROM 3.75 TO 5 cm 7 1
5
6
3 1
9
3
2 4
8
2
1 ZSI 375
1
ACTIVATED WHEN THE BUTTON ( ) PORTRUDES AT THE SIDE OF THE PUMP BUTTON
4
PUMP BUTTON TO EMPTY THE CUFF (PRESS-RELEASE)
7
WINGS TO SEW THE ZSI 375 TO THE INTERNAL SCROTAL TISSUE
2
COMPENSATION POUCH SEPTUM
5
DEACTIVATED WHEN THE BUTTON ( ) PORTRUDES AT OPPOSITE SIDE OF THE PUMP BUTTON
8
3,75 TO 5 CM ADJUSTABLE CUFF, MOULDED IN A CURVE
3
PRESSURE REGULATING TANK
6
PUMP
9
HYDRAULIC CIRCUIT SEPTUM
2 5 ML SYRINGE 3 HUBER NEEDLES
PERFORM A SCROTAL X RAY TO CHECK SPRING POSITION
The spring is compressed. It sits under the upper part of the cylinder. The ZSI 375 is deactivated.
The spring is decompressed. It sits above the top of the cylinder. The ZSI 375 is activated.
POSSIBILITY OF INCREASING ISSUED PRESSURE
The spring is fully decompressed. There is a lack of solution in the hydraulic circuit.
FUNCTIONING
The cuff is inflated, the urethra is compressed. The patient is continent.
Step 2
When the patient wants to urinate, he squeezes and releases the pump button located in the scrotum.
Step 3
The unit draws the fluid out of the cuff releasing the pressure around the urethra. The patient voids, the bladder is empty.
Step 4
Within 2-3 minutes the cuff automatically re-inflates. The patient is continent again.
ZSI375-A4-MARS2012
Step 1