4 - DEACTIVATION & ACTIVATION FOR DOCTORS - Deactivation and activation of the ZSI 375 is performed by doctors during the healing period, to perform an endoscopy or to place a catheter into the urethra.
HOW TO DEACTIVATE THE ZSI 375? Deactivation means that you lock the mechanism of the sphincter. The cuff can no longer be inflated or deflated
FOR PATIENTS - Deactivation and activation are not for patient use. But if there are symptoms of an accidental deactivation, retention (partial or full), incontinence (partial or full), or the ZSI 375 does not seem to function properly, please perform a re-activation.
HOW TO RE-ACTIVATE THE ARTIFICIAL SPHINCTER ZSI 375?
YOUR SOLUTION FOR INCONTINENCE
ZSI 375 ARTIFICIAL URINARY SPHINCTER
Step 1
Activation Button
Pump Button
Hold the Pump Unit in the scrotum with one hand. Firmly press and release the Pump Button twice (the bulb) with the index finger of the other hand to empty the cuff.
Step 2
Pump Button (other side)
Hold the Pump Unit in the scrotum with one hand. Firmly press the activation button with the index finger of the other hand. The activation button is at the opposite side of the deactivation Button and the Pump Button (Bulb).The artificial sphincter is activated.
HOW TO CHECK THAT THE DEVICE IS DEACTIVATED
PATIENT MANUAL More information at: www.zephyr-si.com
- The patient is incontinent. - Perform a Cystoscope to check if urethra is open. - Perform a frontal Xray of device. The image should clearly show device placed upright. Spring should be compressed and sit below the top of the cylinder like so: Deactivation Button
Local Distributor
Top of Cylinder Top of Spring
Then press the deactivation button firmly with the index finger to deactivate the sphincter. The deactivation button is just above the Pump Button. The artificial sphincter is deactivated.
ZEPHYR Surgical Implants Route desJeunes 4 -1227 Acacias (Geneva) Switzerland TĂŠl : + 41 22 327 1026 / Fax : + 41 22 327 1027 Email : contact@zephyr-si.com / www.zephyr-si.com Version 2.1 November 2014