dc-static-hysteresis-graph-plotter

Page 1

DC Static Hysteresis Graph

An ISO 9001:2208 | ISO 14001:2008 | ISO 13485 WHO:GMP Products | GLP Compliant Products


DC Static Hysteresis Graph Automatic measurement on basic magnetization curve and magnetic hysteresis loop under static state of soft magnetic material, accurate measurement on static magnetic characteristic parameters such as initial permeability ?i, maximum permeability ?m, saturation magnetic induction Bs, remanence Br, coercive Hc and hysteresis losses Pu. Windows measurement software applied simply. It conforms to China National Standards GB 3656-83, GB 3657-83, GB/T 13012-91, GJB 937-90, industry standard SJ/T 10281-91 and international standard IEC 60404-4, IEC 404-7. In accordance with measuring principle of ballistic method, combine computer control technology, A/D and D/A, replace conventional ballistic galvanometer with electronic integrator, realize analog ballistic method measurement under microcomputer control, can completely eliminate non-instant error caused by ballistic galvanometer in classical ballistic method, with high measurement accuracy, fast speed and good repeatability, can eliminate the influence of various artificial factors, and supply reliable basis for research on material magnetization process.

ACMAS TECHNOCRACY PVT. LTD. www.measuring-meters.com | www.acmasindia.com | www.scientificlaboratoryequipments.com


Testing sample varieties: soft magnetic ferrite, permalloy,

Operate under Windows 95/98, Windows 2000/XP

amorphous, nano crystal, iron dust core, electrician ferrite

operating systems, conform to Windows software

and stalloy.

specification, visual and simple operation.

Testing sample shapes: toroid, E and U closing

Full automatic control, intelligent identification, users'

samples, and strip (bar) and chip open samples. Closing sample direct winding measurement: sample, magnetizing coil (N1) and measuring coil (N2) form a

process. Sampling waveform and instrument status under real-

no-load transformer. Measure magnetic field intensity through measuring

time monitoring, and can be stopped at any time. The effective sectional area and length of magnetic

interventions are unnecessary in the whole testing

path of samples can be calculated according to the

magnetizing current, magnetic field lock precision as •

high as 0.1%. Adopt electronic integrator to measure magnetic

induction, zero drift of integrator can be self-corrected

overall dimension of samples. Automatic detection on sample test condition, facilitate users to test unknown samples. File system adopts database format, can directly print or output

through software. Solenoid or permeameter shall be equipped to measure open samples, the inner field can be

test result to Excel form. Powerful file management functions: data saving and

measured by the magnetic potentiometer inside. Testing method: analog ballistic method and magnetic

deleting, remove all the data etc. Data files contain complete sampled data, sample

field sweep. Adopt analog ballistic method to measure basic

parameters, instrument parameters and testing schemes, adopt text format, can be conveniently typed

magnetization curve, can accurately measure magnetic characteristic parameters on magnetization •

hysteresis loop or ?(H) permeability curve, and can

curve: ?i, ?m, Bs, value ? of any point. Adopt analog ballistic method and magnetic field

display coordinate information of every data point on •

curve. Support various models of printers, test report

accurately match printer sheet. Print preview function, it can conveniently regulate the

size and edge distance of test report. Test report can be directly printed, or generate JPG

image file. JPG image file can be sent directly through E-mail, or

saved into disk. Test reports contain complete curve diagrams, test

results, test conditions and sample parameters. Both Chinese and English format can be adopted to

sweep to measure saturation hysteresis loop, can accurately measure magnetic characteristic parameters on hysteresis loop: Bs, Br, Hc and Pu. Special explanation: analog ballistic method to measure hysteresis loop is an exclusive measuring method of our company, can greatly improve the •

accuracy of Hc and Br. Automatic demagnetization function before ?i testing,

10 Hz AC saturation demagnetization adopted. Select to test single parameter with analog ballistic method to save time, the time of every testing point

can be set from 0.1 second to 0.9 second. Adopt the B speed feedback, it can automatic regulate the density with the shape of sample magnetic characteristic curve, and curve can also be tested

according to the testing point predefined by users. Powerful software functions with minimal technical

into other software. Display B(H) magnetization curve, B(H) magnetic

facilitate adding user mark and enterprise name. •

Multiple unit systems optional to satisfy the application habits of different users.

requirements for testing staff.

ACMAS TECHNOCRACY PVT. LTD. www.measuring-meters.com | www.acmasindia.com | www.scientificlaboratoryequipments.com


DC Static Hysteresis Graph System Specifications Adopt ballistic method measurement, technical indices as follows:

Parameters measured

Bs (%)

Br (%)

Hc (%)

Pu (%)

µ m (%)

µ i (%)

Uncertainty (k = 2)

1

1

1.5

1

2

5

Repeatability (constant temperature)

± 0.5

± 0.5

± 0.5

± 0.5

±1

±3

When testing hysteresis loop with scanning method, H and B measuring errors can be estimated as:

B: 1% + ½ LSB H: 1% + ½ LSB

ACMAS TECHNOCRACY PVT. LTD. www.measuring-meters.com | www.acmasindia.com | www.scientificlaboratoryequipments.com


Instrument Specifications

PC6616 A/D D/A Card

MATS - 2010SD Static Hysteresisgraph • • •

Output Voltage Output Current Current Range

: 0 ~ ±35 V : 0.01 mA ~ 10 A : 1 mA, 2 mA, 5 mA, 10 mA, 20 mA, 50 mA, 100 mA, 200 mA, 500 mA, 1A, 2A, 5A, 10A total 13

• •

Stability Flux Range

automatic ranges : superior to 0.05% : 0.25, 0.5, 1, 2, 5, 10, 20 mWb,

• • • • • •

A / D Conversion: 16 bits, 16 channels Conversion Time: < 10 µs Conversion Accuracy: 0.02% (full range) D/A Conversion: 12 bits, 2 channels Conversion Accuracy: 0.1% (full range) Bus Structure: PCI busbar

total seven automatic ranges •

Sensitivity

: 0.05 µWb (ballistic method) 0.1 µWb (magnetic field sweep)

ACMAS TECHNOCRACY PVT. LTD. www.measuring-meters.com | www.acmasindia.com | www.scientificlaboratoryequipments.com


www.measuring-meters.com | www.acmasindia.com | www.scientificlaboratoryequipments.com

TECHNOCRACY PVT. LTD. SALES OFFICE (INDIA) 312-313, Vardhman Capital Mall, L.S.C. 10, GulabiBagh, Delhi-110052, INDIA Tel: 0091-11-23646703, 23643054 (M) +91-9313971681| Email: india@acmasindia.com

SALES OFFICE (HONG KONG) Unit D 28 11/f Wing Tat Comm, Bldg 97, Bonham Strand East, Sheung Wan, HONG KONG (PRC). Tel: 0086-13929598046 | 0086-18922303099 Email: hk@acmasindia.com

SHOWROOM 141, Rai Industrial Estate, Rai, Sonepat, Haryana-131029, INDIA Tel: 0091-0-9312219738 (M) +91-9717741167 Email: info@acmasindia.com

SALES OFFICE (RUSSIA) Inmed Trade Street, Ozerkovsky Embankment, Unit No 50 Straine-1, Off-502, Moscow, RUSSIA Tel: 0049-79592345 | Email: russia@acmasindia.com


Turn static files into dynamic content formats.

Create a flipbook
Issuu converts static files into: digital portfolios, online yearbooks, online catalogs, digital photo albums and more. Sign up and create your flipbook.