VIS Spectrophotometer

Page 1

R

VIS SPECTROPHOTOMETER MODEL NO.- ACM-UV-1601


R

VIS SPECTROPHOTOMETER

FEATURES Ÿ Wide wavelength range, satisfying requirements of various fields Ÿ The split-beam ratio monitoring system provides accurate measurements and enhances baseline stability. Ÿ Five options for spectral bandwidth selection, 5nm, 4nm, 2nm, 1nm and 0.5nm, made according to customer's need

and satisfying the requirements of pharmacopoeia Ÿ Fully automated design, realizing the simplest measurement. Ÿ Optimized optics and large scale ingegrated circuits design, light source and receiver from world famous

manufacturer all add up to high performance and reliability Ÿ Rich measurement methods: wavelength scan, time scan, multi-wavelength determination, multi-order derivative

determination, double-wavelength method and triple-wavelength method etc., meet different measurement requirements Ÿ Automatic 10mm 8-cell holder, changeable to automatic 5mm-50mm 4-posion cell holder for more choices.


R

Ÿ Data output can be obtained via a printer port and a RS-232 interface (RS485 and USB port optional) Ÿ Parameters and data can be saved for user's convenience Ÿ PC controlled measurement can be achieved for more accurate and flexible requirement.

TECHNICAL SPECIFICATIONS

Wavelength Range

190-1100nm

Spectral Bandwidth

2nm(5nm, 4nm, 1nm, 0.5nm optional)

Wavelength Accuracy

±0.3nm

Wavelength Reproducibility

0.15nm

Photometric System

Split-beam ratio monitoring; Auto scan; Dual detectors

Photometric Accuracy

±0.3%T(0-100%T), ±0.002A(0-0.5A), ±0.004A(0.5A-1A)

Photometric Reproducibility

0.2%T

Working Mode

T, A, C, E

Photometric Range

-0.3-3A

Stary Light

=0.1%T(NaI, 220nm, NaNO2 340nm)

Baseline Flatness

±0.002A

Stability

0.001A/30min (at 500nm, after warming up)

Noise

±0.001A (at 500nm, after warming up)

Display

6 inches high light blue LCD

Detector

Silicon photodiode

Power

AC:220V/50Hz, 110V/60Hz, 180W

Dimensions

630×470 ×210mm

Weight

26kg


An ISO 9001 : 2208 | ISO 14001 : 2008 | ISO 13485 WHO: GMP Products | GLP Compliant Products

www.acmasindia.com | www.cleanroom-equipments.com | www.measuring-meters.com

TECHNOCRACY PVT. LTD. SALES OFFICE (INDIA) Office no. 506 & 507, 5th Floor, Plot no.- C-9, Pearls Best Height-II, Netaji Subhash Place, Pitam Pura, Delhi- 110034 Land Line No.: +91 - 11- 47619688, Fax: +91-11-47619788 E-mail.: info@acmasindia.com

SALES OFFICE (HONG KONG) Unit D 28 11/F Wing Tat Comm,Bidg 97, Bonham Strand East, Sheung Wan, Hong Kong (PRC) Tel.: 0086-13929598046 0086- 18922303099 E-mail.: hk@acmasindia.com

SHOWROOM J-284 Sector 1 Bawana Industrial Estate, Bawana, Delhi, INDIA-110039 Tel.: 0091-0-9312219738 (M) +91- 9717741167 E-mail.: info@acmasindia.com

SALES OFFICE (RUSSIA) Inmed Trade Street Ozerkovsky Embankment, Unit No 50, Straine- 1, Off- 502, Moscow, Russia E-mail.: info@acmasindia.com Tel.: 0049- 79592345 | Email: russia@acmasindia.com


Turn static files into dynamic content formats.

Create a flipbook
Issuu converts static files into: digital portfolios, online yearbooks, online catalogs, digital photo albums and more. Sign up and create your flipbook.