PH METERS CARE AND MAINTENANCE OF pH ELECTRODES
INSPIRING TOMORROW’S SCIENTISTS THROUGH PRACTICAL SCIENCE
HANNA pH Electrode Cleaning and Storage
CLEANING & STORAGE SOLUTIONS
SKU
Type
Qty per Pack
PH25000 PH25010
pH Cleaning Solution pH Storage Solution
500 ml 500 ml
HI-7061L is a general-purpose cleaning solution that can be used for a variety of applications. Electrodes become dirty from use and will produce inaccurate results even if they read correctly in a pH buffer. HANNA’s cleaning solutions eliminate impurities and residues which are left on electrode surfaces when immersed in samples during measurement or when stored incorrectly. HANNA suggests cleaning the bulb and junction of your electrode regularly in the ready-to-use cleaning solutions to ensure that the probe is always clean and to prevent any clogging of the junction. All cleaning solutions are packaged in an airtight bottle with a tamper-proof seal to ensure quality and freshness. All solutions show a batch number and expiry date.
Useful tips: After cleaning, rinse the electrode in tap water and then place it in a HANNA Storage Solution before calibrating or taking measurements. Always follow up with a HANNA Buffer/calibrating solution for the best results.
Buffer Solution Capsules These buffer solution capsules for use with pH electrodes are convenient and easy to use. Simply split the capsule in two and add to 100 ml of deionised/ distilled water and mix. No minimum shelf life, so will keep for years!
CLEANING & STORAGE SOLUTIONS
SKU
Type
Qty
PH120120
pH4 (orange capsules)
pk 10
PH120125
pH7 (green capsules)
pk 10
PH120130
pH9 (white capsules)
PH120135
pH10 (blue capsules)
pk 10 pk 10
HANNA pH COMBINATION ELECTRODE - GLASS BODIED Plastic bottles, air-tight with grommet to stop pH electrodes from drying out. SKU
Type
BO80400
Storage bottle for 12 mm diameter probes
HANNA buffer solutions are prepared, ready-to-use, according to precise formulas and standardised in accordance with the National Institute of Standards and Technology (NIST) to ensure quality and accuracy. All solutions show a batch number, an expiry date, and the correlation table between pH and temperature. A certificate of analysis is available on request. All bottled solutions are sealed with an airtight, tamper-proof seal to prevent any Carbon Dioxide from entering and causing a change in the pH buffer value.
Top Tips: • pH electrodes must NEVER be allowed to dry out • pH probes should be stored in a bottle (BO80400) with the following in order of preference Long term: 1. pH storage solution (PH25010) Short term: 2. Saturated KCl solution
CARE AND MAINTENANCE OF pH ELECTRODES
3. pH 7 or pH 4 buffer 4. Deionised water (last resort to prevent drying out) • Electrodes should never be stored in deionised water, unless as a last resort to prevent drying out, testing samples or solvents • Electrode tips should always be rinsed after use • If electrodes need cleaning physically, use a soft tissue and mild detergent or propan-1-ol • Regularly inspect electrode glass for cracks or chips; if any are present, replace the electrode • It is not unusual for electrodes to have salt deposits present – these will disappear when rinsed • If the bulb becomes dry during storage, soak in storage solution, pH 7, or pH 4 buffer for at least an hour prior to use
Timstar
Phoenix House, Stafford Drive, Battlefield Enterprise Park, Shrewsbury, Shropshire SY1 3FE, United Kingdom timstar.co.uk customer.services@wf-education.com +44 (0)1743 812 200 Connect with us