Sinus-T Surgical Instruments Sinus-T
Osteotome Technique by TR Sinus System “TR Sinus System” contains trephine drills and round drills, will provide easy & safe sinus membrane lifting.
Order Code: SNT 01
Surgical Instruments
To overcome the short of bone length on maxillary molar area in case of sinus membrane lifting, osteotome technique is useful to obtain sufficient bone for placing implant.
Ø3.0 / Ø3.5 / Ø4.0
Ø3.0 / Ø3.5 / Ø4.0
Ø2.0 / Ø2.5 / Ø3.0
8
7
6 5
Ø5.0
4 3
15
15 O-Ring
29
29
ST 4006
ST 4007
ST 4008
ST 4009 ST 4010
Drill Stopper
Ratchet Adapter
Depth Gauge
RA 8011
Trephine Drill
Round Drill
STD 3020L STD 3525L STD 4030L
SRD 3008L SRD 3508L SRD 4008L
Ø2.5
72
SDG 2515
ST 4011
Sinus Protocol Precise X-Ray reading is very important to approach closely under the sinus floor with trephine drill.
0.5~1mm
7mm
15~20rpm
1 Drilling
2 Push
3 Drilling
4 Insertion
1,000~1,500rpm
1 Drilling (Trephine Drill)
Choose correct length of stopper and put it on the trephine drill and then drill 0.5~1mm under the sinus floor. (Drill Speed : 1,000 ~ 1,500 RPM) Putting stopper on the trephine drill reduce the risk of perforation of sinus floor accidentally during the surgery.
2 Push (Osteotome)
After malleting at correct position of osteotome, do microfracture safely on sinus floor. Mallet 2 or 3 times to obtain a microfracture on sinus floor. If the several malletings do not work for the microfracture, do drilling again with trephine drill toward sinus floor. (Drill Speed : 15 ~ 20 RPM)
3 Drilling (Round Drill)
Enter the osteotomy site with round drill, thereby lift the sinus membrane slowly. (Drill Speed : 15 ~ 20 RPM) Round drill minimizes the risk of perforation of sinus membrane with rounded shape.
4 Insertion (Fixture) After checking the lift of the sinus membrane, place the fixture.
73
Surgical Instruments
15~20rpm
Dr.SOS Surgical Instruments Dr.SOS
Immediate Action For Emergency The Kit for the easy removal of Broken Abutment Screw Regeneration of damaged internal threads in the fixture
External Guide
Order Code: DRS 00
Internal Guide
Remove Bar
0.3 Slot Driver Reverse Cutting Drill
Surgical Instruments
Screw Tapping
0.3 Slot Driver When Torx Abutment Screw Head was damaged, you can use 0.3 Slot Driver instead of 1.7 Torx Driver.
Machine MD 0322
Ratchet HD 0315A
1.7 Torx Abutment Screw
74
External Guide (External Connection) It is connected to fixture and guides through Remove Bar to enter and support broken abutment screw.
Hex
2.4 Hex Branemark Connection
2.7 Hex Branemark Connection
3.4 Hex Branemark Connection
EXT 24S
EXT 27S
EXT 34S
EXT 24L
EXT 27L
EXT34L
Short
Long
External Hex
Surgical Instruments
Internal Guide (Internal Taper Connection) It is connected to fixture and guides through Remove Bar to enter and support broken abutment screw. Mores Taper
Morse Taper 6째
Morse Taper 8째
Morse Taper 11째
INT 6S Torx
INT 8S Torx
INT 11S Hex
Morse Taper 5.7째
Torx
*HEX Type
INT 6L
INT 8L
INT 11L
INT 5.7S
Non-Torx
*Option: ITI Guide
75
*Narrow Fix
Reverse Cutting Drill
Remove Bar
Connect the Reverse Cutting Drill to low speed handpiece, with 1,000~2,000 RPM in reverse rotation. Then, repeat drilling up and down to grind and change the top of the broken screw.
Under 20 RPM in reverse rotation
*CAUTION): Be careful not to break it with overheating.
RBEX 24S
RBEX 24L
IRBS
IRBL
RCD 36
Ø1.25 External 2.4 Hex
External 2.7/3.4 Hex Internal 6°, 8°, 11°
Surgical Instruments
Screw Tapping When internal threads in the fixture were damaged, use Screw Tapping to regenerate fixture internal threads.
SGT 1620A
SGT1820A
SGT2020A
SGT2520A
Screw External Fixture
Ø1.6
Ø1.8
Narrow
Non Submerged Internal Fixture Submerged Internal Fixture
Ø2.0
Ø2.5
Regular T-Wide
Wide
Internal Fixture (PSI, IFI) SM Narrow
SM Regular SM Wide Extra Wide
Handle
Torque Wrench
RBH 10
DTW 0060
76
Immediate Action For Emergency, Dr. SOS The Kit for the easy removal of Broken Abutment Screw Regeneration of dameged internal threads in the fixture
Ready After cleaning thoroughly broken screw area, repeat inserting oral-use oil to let the broken screw out.
Guide Fixing Locate a proper guide from the "Dr.SOS Kit" onto a fixture inside mouth and support it securely to make it vertical. *You can use ratchet to support the guide when it's hard to support the guide by hand.
A) Remove Bar Connect Remove Bar to the Handle, reverse it by hand to remove the Broken Abutment Screw, and be careful not to drop off the product into the mouth. *Confirm the level of line on the Remove Bar.
B) Remove Bar + Handpiece After connecting the Remove Bar into low speed handpiece, it is also possible to remove it with reverse rotation of low RPM.
Surgical Instruments
*less than 25 RPM. *Remove Bar and drill extension connection enables a longer use.
Explore Use explore to completely remove it after reaching certain height.
Screw Tapping When fixture internal threads were damaged, use Screw Tapping to regenerate fixture internal threads and connect Abutment.
Tip)
TIP) When pulling, hand will feel it touches. If no touch on hands, lean guide and bur left and right until getting feel. 77
Bone Expander Set Surgical Instruments Bone Expander
Order Code: BEP(M) 00
Osteotomes have been utilized in a variety of techniques designed to prepare implant sites in maxilla, to elevate the floor of the maxillary sinus and to expand the atrophic edentulous ridge. Despite its effectiveness, the surgical mallet is not well tolerated by patients. Using a drill to prepare implant sites in maxilla and mandibular also have a risk of maxillary perforation and penetration into the nerve 2 3 0 1 mandibularis. 7 9 As safe alternative and complementary instruments to the 4 5 6 8 10 osteotomes and drills, 11 Bone Expander can be used for preparation of implant sites, atraumatic ridge expansion and condensing of bone.
Description
Code
0
1.2mm Initial Drill
MSD 1218M
1
1.0/1.6mm Bone Expander
2
Apical/Neck(Ă˜)
Length
1.2
16
BEP 1716
1.0 / 1.6
17
1.3/2.3mm Bone Expander
BEP 1723
1.3 / 2.3
17
3
1.7/3.1mm Bone Expander
BEP 1731
1.7 / 3.1
17
4
2.4/3.7mm Bone Expander
BEP 1737
2.4 / 3.7
17
5
3.0/4.4mm Bone Expander
BEP 1744
3.0 / 4.4
17
6
3.5/5.1mm Bone Expander
BEP 1751
3.5 / 5.1
17
7
Driver(short)
MHDC 2520
Color Code
For Contra Angle
8
Driver(long)
MHDC 2525
9
Driver(short)
MHDR 2513
For Ratchet Wrench
10 11
Driver(long)
Torque Wrench
Open Wrench
MHDR 2518
DTW 0060
OW 002
78
(mm)
0
1
2
Pic 1
Pic 2
Pic 3
3
IFI35xx
Pic 4
4
IFI40xx SFN38xx
5
Pic 5
IFI48xx IFI65xx SFR45xx
Pic 6
6
IFI55xx IFI65xx SFW53xx
Pic 7
*Option: MSD 2016(Pic 3->)
Protocol DIO Implant Bone Expander hardens Bone density to erase difficulties caused from Osseointegration of soft or Poor Bone. (This procedure shows the method of Implantation of SFW 5312) Pic 1
In order to give depth and direction to desired surgical area, 1.2mm drill is used. (800 rpmn 1200rpm is the capable engine speed)
Pic 2
To expand the implantation site, 1.6mm Bone Expander (Color Code: Yellow) is used (Check Laser Marking) *Connect the Bone Expander with Adapter. Then finish with Ratchet Wrench.
Pic 3
Use 2.3 Bone Expander (Color Code: Yellow) to expand the desired surgical area. (Check Laser Marking)
Pic 4
Use 3.1mm Bone Expander (Color Code: Yellow) to expand the desired surgical area. (Check Laser Marking)
Pic 5
Use 3.7mm Bone Expander (Color Code: Red) to expand the desired surgical area. (Check Laser Marking)
Pic 6
Use 4.4mm Bone Expander (Color Code: Green) to expand the desired surgical area. (Check Laser Marking)
Pic 7
Continue by using 5.1mm Bone Expander to expand the desired surgical area. (Check Laser Marking) Hardening the Bone has a risk of Ischemia which makes the Doctor to check whether sufficient bleeding is occurring. Give Scrape to the desired surgical area to increase bleeding - Bleeding help Osseo Integration. Implant SFW 5312 to the expanded hole by Bone Expander
79
Sinus Lift Kit Surgical Instruments Sinus Lift Kit Custom designed kit comprised of 6 tools used in sinus procedures
Code
Surgical Instruments
Sinus Elevator & Membrane Lift Tips
80
Order Code: SLK 01
Description
DSNK 01
Membrane Detach Elevator
DSNK 02
Membrane Elevator
DSNK 03
Membrane Elevator
DSNK 04
Bone Graft Packer
DSNK 05
Membrane Elevator
Osteotome Kit Surgical Instruments Osteotome Kit
Diameter(L)
0
Order Code: OST 01
Image
Code
OSE 2228
Ø2.8
Application
Sinus Elevation
Ø2.2
1
OSL 2028
Ø2.8
SM/ IFI/ FTN Initial Drill
Ø2.0
2
OSL 2130
Ø3.3
OSL 2133
Ø3.9
OSL 3039
SFR 45xx/ IFI 40xxM Compaction
Ø4.6
OSL 3546
SFW 53xx/ IFI 48xxM/ FTN 50xxB Compaction
Ø2.1
3 Ø2.1
4
FTN 33xxB
Ø3.0
Compaction
SFN 38xx/ IFI 35xxPM Compaction
Ø3.0
5 Ø3.5
81
Bone Planer Set Bone Planers are used to flatten the uneven bone during osteotomy reparation
Size & Code Narrow
Regular
Wide
Ă˜5.65
Ă˜5.9
30 10
5.5
Ă˜4.95 BP 5010
BP 5710
BP 5910
Protocol
1 Guide Drilling Drill through the cortical bone at the selected site for the implant placement.
2 Initial Drilling Select an appropriate drill stopper according to the desired depth of implant placement.
*Handpiece speed is recommended to be in a range between 400~600 rpm.
82
3 Bone Planing Insert its guide part into a drilled osteotomy site and remove the uneven bone.