Neonatal cranial ultrasound safety aspects 2013 mcqs

Page 1

ECMUS – The Safety Committee of EFSUMB : Tutorial

MCQs Neonatal cranial 1.

Imaging of the neonatal brain can be done with neonatal cranial ultrasound (CUS) or magnetic resonance imaging (MRI) a) CUS is better than MRI for preterm babies b) MRI is better than CUS for preterm babies c) CUS is better than MRI for term babies d) Both methods have advantages and disadvantages, and there is no agreement on which modality is better for imaging the neonatal brain

2.

Neonatal cranial ultrasound is used a) To assess brain maturation and structural and acquired anomalies b) To screen for hemorrhage in term babies c) To verify brain abnormalities demonstrated by magnetic resonance imaging d) Only when there are clinical signs of brain abnormalities (eg seizures in the newborn)

3.

The standard planes obtained though the anterior fontanelle are a) 5 sagittal planes b) 6 coronal planes c) 6 coronal and 5 sagittal planes d) 5 coronal and 6 sagittal planes 4. During neonatal cranial ultrasound TI values should be kept lower than the maximum recommended by BMUS and ECMUS because a) The impedance of the fontanels are difficult to measure and calculation of TI may suffer from large variation in impedance b) The model for calculating TI ignores transducer self heating and there is experimental evidence that this can occur during neonatal cranial ultrasound c) Calculated TI values may be wrong (up to a factor of 3-4) when compared with in vivo temperatures d) The brain is more sensitive to heating than other organs ***

Š 2014 ECMUS Tutorial: Technical Quality Assurance– safety aspects (Edt: CK & LD)|www.efsumb.org/ecmus 1/2


ECMUS – The Safety Committee of EFSUMB : Tutorial Correct answers: 1d: Both methods have advantages and disadvantages, and there is no agreement on which modality is better for imaging the neonatal brain 2a: To assess brain maturation and structural and acquired anomalies 3c: 6 coronal and 5 sagittal planes 4b The model for calculating TI ignores transducer self heating and there is experimental evidence that this can occur during neonatal cranial ultrasound

© 2014 ECMUS Tutorial: Technical Quality Assurance– safety aspects (Edt: CK & LD)|www.efsumb.org/ecmus 2/2


Turn static files into dynamic content formats.

Create a flipbook
Issuu converts static files into: digital portfolios, online yearbooks, online catalogs, digital photo albums and more. Sign up and create your flipbook.