PURELAB Flex Range

Page 1

Pure and ultrapure laboratory water from a single system


Your laboratory water specialists ELGA has been a trusted name in water purification for over 50 years, pioneering innovative technologies and award winning product design for our customers. Part of the world’s leading water services company Veolia Water Solutions and Technologies, ELGA have the most comprehensive range of water purification systems available today for laboratory research, healthcare and diagnostics. Water is the most common reagent used in today’s laboratories for research and testing applications and we understand how important accurate consistent test results are. The PURELAB flex range is designed around your needs, delivering accuracy, flexibility and ease of use with an innovative and ergonomic design.

Find a PURELAB flex to match your application

PURELAB flex 2

Daily volume

>10 Liters

Water quality Feedwater source Typical applications

PURELAB flex 3&4

<10 Liters

Type I ultrapure water Pre-purified water Electrophoresis GFAAS (Graphite Furnace Atomic Absorption Spectrophotometry) HPLC IC (Ion Chromatography) ICPAES (Inductively Coupled Plasma Atomic Emission Spectrometry) ICPMS (Inductively Coupled Plasma Mass Spectrometry) Mammalian and bacterial cell culture Plant tissue culture Qualitative analyses

>10 Liters Type II/Type II+ pure water

Pre-purified or tap water

Electrochemistry

Molecular biology

PURELAB flex 1

Pre-purified water AAS (Atomic Absorption Spectrophotometry) Buffer and media preparation Electrophysiology FAAS (Flame Atomic Absorption Spectrophotometry) Feed to ultrapure water systems Glassware washing/rinsing General chemistry Histology Microbiological analysis RIA (Radioimmunoassay) / ELISA (Enzyme Linked Immunoabsorbant Assay) Sample dilution and reagent preparation Spectrophotometry

PURELAB flex 3&4

PURELAB flex 1

<10 Liters

>10 Liters

Type III RO water Pre-purified or tap water

Pre-purified water

Autoclave feed Feed to ultrapure water systems Hydroponics Plant growth cabinets Stability chambers Steam generators Sterilizer feed


Water purification made easy Intuitive to use, ergonomic handset design Clear water purity display for absolute confidence as you dispense Flexible dispensing in four different ways:

Realtime TOC monitoring for critical analytical applications

Variable flow – drop by drop or up to 2 liters per minute

Autovolume dispense from 50ml to 60 liters and repeat dispensing

• Hands free with optional foot pedal • Locked dispense for glassware filling

Prioritized information shown at all times (system status, TOC, alarm)

POU filters for multiple applications •

Remove endotoxins, DNase, RNase and bacteria

• Conforms to international water

standards e.g. CLSI, CLRW, ISO 3696: Grade 1,2,3, ASTM D1193-06, Pharmacopeia USP, EP and JP

Optional accessories

• Wall mounting bracket • Foot pedal • Thermal printer • Point of Use filters • Leak sensor


Designed for today’s laboratories

Customized settings to suit your application

185nm/254nm UV lamp for low TOC water production needed for organic sensitive application

Easy-access for routine maintenance

Multiple dispense positioning, wall, bench, height adjustable arm, hand-held dispensing

Designed for simple easy installation

Purification pack quick and easy to replace

Data capture via USB for system performance validation

Quick and easy sanitization

Minimize microbial growth to give you absolute confidence in the water purity. The simple procedure involves one single sanitization pack so no harsh chemicals need to be ordered, handled or mixed. No additional consumables need to be replaced during this short process.

Global network, local support

Our highly trained service engineers will apply their expertise to the installation, validation and maintenance of your water purification system. We aim to provide a first-time-fix service.


Inside the PURELAB flex PURELAB flex 2 Process Flow Diagram

Water Purity Sensor 185/254 nm UV Lamp Feedwater Inlet

Purification Pack

Water Purity and Temperature Sensor

0.2µm Point of Use Filter or Biofilter (optional)

Water Purity and Temperature Sensor

(Pre-purified Water)

Dispense Handset

Recirculation Pump

PURELAB flex 3 Process Flow Diagram

Reverse Osmosis

CVF Composite Vent Filter Reservoir

Water Purity Sensor 185/254 nm UV Lamp

Purification Pack

Water Purity and Temperature Sensor

Water Purity and Temperature Sensor Feedwater Inlet (potable tap water)

Manual Outlet

Dispense Handset 0.2µm Point of Use Filter or Biofilter (optional)

Recirculation Pump

Want to know more?

Find out more or book your demonstration at www.purelabflex.com

Commitment to quality

PURELAB flex systems are supplied with a Certificate of Conformity, ensuring they have been manufactured and tested at ELGA LabWater Global Operations, whose Quality Management System is approved by Lloyds Register Quality Assurance (LRQA) and complies with BS EN ISO 9001:2008.

ELGA’s manufacturing facility has been approved by LRQA and complies with the environmental management system ISO 14001:2004. The PURELAB flex has been designed to conform to the requirements of the Waste Electrical & Electronic Equipment (WEEE) Directive and the Restriction of Hazardous Substance (RoHS) Directive.

Assured safety

The PURELAB flex is compliant with CE directives associated with safety and electromagnetic compatibility and is approved and tested by an accredited external company. The PURELAB flex is also compliant with the requirements of IEC/EN61010-1 and UL61010-1 and is marked with an ETL label showing compliance.


Treated Water Specifications MODEL

PURELAB flex 1

Daily volume Delivery flow rate – maximum Inorganics (resistivity @25°C) Organics (TOC) – typical Direct from internal reservoir

Bacteria – typical (when fitted with POU Filter) Bacteria – typical (when fitted with Biofilter) Endotoxin (when fitted with Biofilter)

Dependant on feedwater

2

PURELAB flex 2

PURELAB flex 3

PURELAB flex 4

>10 liters Up to 2 l/min As per feedwater Dependant on feedwater N/A

>10 liters Up to 2 l/min 18.2 MΩ-cm Dependant on feedwater N/A

>10 liters Up to 2 l/min 18.2 MΩ-cm <5ppb 1

<10 liters Up to 2 l/min 18.2 MΩ-cm <5ppb 1

<10 liters Up to 2 l/min 18.2 MΩ-cm <5ppb 1

N/A

Type III / RO water 2

Type III / RO water 2

N/A N/A

N/A N/A

<20 pg/ml <0.002 ng/ml

<20 pg/ml <0.002 ng/ml

<20 pg/ml <0.002 ng/ml

<1 CFU/10ml <1 CFU/10ml N/A

DNase (when fitted with Biofilter) RNase (when fitted with Biofilter) 1

PURELAB flex 1

With Purification Pack

<1 CFU/10ml <1 CFU/10ml <0.001 EU/ml

<1 CFU/10ml <1 CFU/10ml <0.001 EU/ml

<1 CFU/10ml <1 CFU/10ml <0.001 EU/ml

<1 CFU/10ml <1 CFU/10ml <0.001 EU/ml

(Equal to or >) Type III RO Water

Dimensions and weights Dimensions Operational weight Installation

Width 236mm, Depth 374mm, Height minimum 900mm, Height maximum 1020mm 10kg (22lb) Bench/wall

10.5kg (23.1lb) Bench/wall

11kg (24.2lb) Bench/wall

Width 236mm, Depth 470mm, Height minimum 900mm, Height maximum 1020mm 23kg (57.3lb) Bench/wall

23kg (57.3lb) Bench/wall

Potable Tap Water

Originally from potable supply, then pre treated. Preferably reverse osmosis (RO) or filtered service deionization (SDI) or distilled.

Feedwater requirement Source

Conductivity Contaminant Hardness Free Chlorine Chloramine Total Chlorine Silica Carbon Dioxide Maximum Fouling index Iron / Manganese Organics (TOC) Particulates Temperature Flowrate (requirement at 15°C) Drain requirements (gravity fall with air gap) 3

Originally from potable supply, then pre treated. Preferably reverse osmosis (RO) or filtered service deionization (SDI) or distilled.

<1 µS/cm 3

<1 µS/cm 3

<1 µS/cm 3

<2000 µS/cm 4

N/A <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <2 ppm <0.1 ppm

N/A <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <2 ppm <0.1 ppm

N/A <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <2 ppm <0.1 ppm

<350 ppm as CaCO3 <0.5 ppm Cl2 <0.2 ppm Cl2 <0.5 ppm Cl2 <30 ppm SiO2 <30 ppm

<1 N/A N/A

<1 N/A N/A

<1 N/A N/A

(recommended <20 ppm)

<10 <0.05 ppm <2 ppm

A 0.2 micron membrane pre filter is recommended for all non-RO feeds to extend point-of-use filter life 4 - 40°C (Recommended 10 - 25°C) <2 l/min <2 l/min <2 l/min Up to 75 l/hr None required None required None required >90L/hr 4

If feedwater is in the range of 1 μS/cm – 30 μS/cm you will need a pre-conditioning cartridge

<30 µS/cm

TBC <0.05 ppm Cl2 <0.05 ppm Cl2 <0.05 ppm Cl2 <2 ppm SiO2 <30 ppm

(recommended <20 ppm)

<1 N/A <50 ppb recommended N/A

Up to 60 l/hr >70L/hr

Purification pack life may vary with feedwaters >1400 μS/cm

Feedwater pressure Maximum Minimum 5

22psi (1.5bar) Flooded suction

22psi (1.5bar) Flooded suction

22psi (1.5bar) Flooded suction

90psi (6bar) 5 30psi (2bar) 5

90psi (6bar) 1psi (0.0 7bar)

100VA

100VA

If <60psi (4 bar) a separate boost pump is recommended

Electrical requirements Mains Input System control voltage (not including pumps and UV) Power consumption during peak demand Noise level

100-240V ac, 50-60Hz 24V dc 120VA

120VA

120VA <40dBa

ELGA LabWater Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505 Email: info@elgalabwater.com Website: www.elgalabwater.com ELGA is the global laboratory water brand name of Veolia Water Solutions and Technologies. The information contained in this document is the property of VWS (UK) Ltd, trading as ELGA LabWater, and is supplied without liability for errors or omissions. © VWS (UK) Ltd. 2010 – All rights reserved. ELGA® and PURELAB® are registered trademarks of VWS (UK) Ltd.