EMC 500

EMC 500 Gas Quality Measuring Device For Determining Calorific Value and Standard Density without Burning Samples Application

For determining superior and inferior calorific value, standard density, Wobbe index and CO2 concentration without burning samples For determining energy content thanks to integrated corrector function (optional) Not influenced by ambient air Accuracy better than Âą 0.5% Suitable for natural, non-aggressive and manufactured gases; nitrogen, carbon dioxide and propane; and other gases upon enquiry

Technical description EMC_500_00_DE

Works without carrier or aux. gases Measures calorific values of altered natural gases up to 20% added air

EMC_500_00_GB EMC_500_00_PL_2002_01 EMC_500_00_RU

Determines the methane index (optional) Mobile version available

Cut-away views EMC_500_s

PTB-approved Operating pressure

pmin 0.5 bar, pmax 3.0 bar

Actuating delay

< 60 s

Measuring gas consumption

15 l/h

Electrical connections

Supply voltage 230 VAC, 24 VDC

Operation & maintenance, spare parts EMC_500_20_DE EMC_500_20_GB EMC_500_20_PL EMC_500_20_RU

Output signals 0/4–20 mA DSfG interface according to DVGW G 485 To top RMG Regel + Messtechnik GmbH | Osterholzstrasse 45 | D-34123 Kassel, Germany | Phone +49 (0)561 5007-0 | rmg@honeywell.com | Imprint / Legal Notice | Privacy Statement