Gravimetric dust measurement
GMD 06
PRODUCT INFORMATION 3
Gravimetric dust measuring device
Filter controller Dust measuring devices
Measuring principle The GMD 06 is an automatic sampling device for gravimetric dust measurements. These measurements are used as reference measuring method for the calibration of dust measuring devices (in Germany according to DIN VDI 2066). Furthermore it is possible to measure flow. The GMD 06 is able to determine independently all exhaust gas parameters being necessary for the dust measurement (e.g. humidity of the measuring gas, velocity in the stack as well as temperature and pressure). Under consideration of these parameters the GMD 06 extracts automatically a partial gas flow in isocinetic way from the exhaust gas channel. This partial gas flow is sucked via a filter, in which the dust contained is precipitated. A f t e r t h e measurement the filter is treated and weighed. When the filter mass weighed out has been entered manually, the GMD 06 calculates the dust content of the sample in operation and norm state.
Technical data control unit
The manual one For the gravimetric determination of the dust concentration Application: Measurement 1
of the absolute exhaust gas humidity Measurement of the 1 exhaust gas velocity (and D p, temperature) Measurement of the 1 exhaust gas content semi-automatic 1 calculation of the dust concentration after the determined dust mass has been entered
!Advantages:
Compact 1
system consisting of probes and control unit, easy use Variable application 1 possibilities due to probe adjustment to customer conditions On-site diagnosis of the 1 measuring values due to high resolution display 1 Excellent priceperformance ratio
Technical data portable case (control unit integrated) plus additional cases for accessories
probes
1 x humidity measurement, 1 x D p, temperature and flow
display
graphic display
humidity
0 ... 40 Vol%
dynamic pressure
0 ... 1.500 Pa
static pressure
-30 ... 10 kPa
flow (sampling) media temperature ambient temperature dew point difference flow velocity
10 ... 50 l/min max. 400 °C -20 ... +50 °C min. +5 K independent from gas flow
reaction time
< 8 sec
data memory
125 samples
power supply
230 VAC / 50 Hz, 200 W
TUV approval
none
Dr. Födisch Umweltmesstechnik AG
Function:
Dr. Födisch Umweltmesstechnik AG Zwenkauer Straße 159 D - 04420 Markranstädt Tel. Fax. Mail. Web.
+49-34205-755-0 +49-34205-755-40 vertrieb@foedisch.de http://www.foedisch.de
© Dr. Födisch Umweltmesstechnik AG - February 2008 subject to change
rd ext end editio ed n
Filter controller
PFM 92
PFM 97 ED
PFM 02 V
Filter controller
Dust measuring device
Dust measuring device
Measuring principle Function:
If 2 solid bodies get into contact by friction or touching, an electrical charge transfer will be caused. The arising charge difference is the base for triboelectric filter controllers and dust measuring devices which use the charge exchange between measuring probe and nearby streaming or direct impacting dust particles. Auswerteeinheit Staubsensor The triboelectric signal depends on the mechanical and electrical properties of the dust. Apart from the dust concentration it seems that the gas velocity has the biggest influence on the triboelectric charge transfer.
General advantages: In-situ online measurement of the dust content 1 High measuring accuracy 1 simple assembly without special tools 1 low demand for maintenance (long operating times) 1 no influence of probe pollution on the measuring 1 result no purging air required 1 low operating costs 1
Technical data control unit dimensions and degree of protection (control unit) probe immersion depth and degree of protection (probe) measuring principle
measuring range calibration display media temperature ambient temperature dew point difference flow velocity
The original one For the continuous monitoring of filter plants Application: monitoring of filtering 1 precipitators (bag filters, cartridge filters) monitoring of centrifugal 1 separators (cyclones) tendency measurement 1 of dust concentrations in raw gas
The extractive one Advantages: measuring ranges adjustable display of measuring 1 values on LCD display at the control unit Effective filter diagnosis 1 (localisation of defective filter ele-ments; determination of damaged membrane va l v e s , n o t t i g h t l y shutting valves ...) 4 1
For the continuous dust concentration measurement in wet and sticky gases Application: wet scrubber in waste 1 incinerations land reclamation plants 1 monitoring of exhaust 1 gases in woodprocessing industry (wood dryer, MDFdryer) application in insulant 1 material industry application behind wet 1 FGD
The delicate one Advantages: compact design 1 automatic dilution of the 1 gas flow sucked in self-monitoring of the 1 measuring gas flow automatic back-purging 1 for self-cleaning low demand for mainte1 nance
For the monitoring of filter plants with increased comfort Application: as dust 1
measuring system in combination with the signal of exhaust gas velocity (e.g. FMD 02) monitoring of dust 1 emissions in all industries with dry exhaust gas, especially behind filtering precipitators
Advantages: device (no separate control unit necessary) operating and status 1 displays via LCD display of measuring 1 values on LCD fast identification of 1 filter defects by on-line line diagram display of dust content 1 in mg/m³ after calibration compact 1
Technical data
Technical data
Technical data
weather proof aluminium case
steel-sheet case, mounted on profile frame (incl. blower)
control unit in the probe head (no separate control unit)
210 x 240 x 280 mm (H x W x D), IP 65
600 x 1700 x 500 mm (W x H x D), IP 65
-
stainless steel probe rod with pre-amplifier in the probe head
extractive sampling with GRP weather protection hood
stainless steel probe rod at the probe head
300 mm (standard), IP 65
500 x 750 x 1000 mm (W x H x D), IP 65
300 mm (Standard), IP 65
dust: measurement with 1 triboelectric sensor
dust and internal operating variables
dust: measurement with 1 triboelectric sensor
0,3 ... 1.000 mg/m³ (special measuring ranges on request)
0 ... 15 mg/m³ (max. 500 mg/m³)
0 ... 10 mg/m³ (max. 1.000 mg/m³)
by gravimetric reference measurements
by gravimetric reference measurements
by gravimetric reference measurements
LCD display (0 ... 100 %)
LCD display
LCD display (meas. value in mg/m³ after gravim. calibration)
max. 280 °C (higher temperatures on request)
max. 280 °C
max. 280 °C
-20 ... +50 °C
-20 ... +50 °C
-20 ... +50 °C
min. +5 K
independent from gas humidity
min. +5 K
from ca. 3 m/s
independent from gas flow
from ca. 3 m/s
1 x 4 ... 20 mA (dust)
5 x 4 ... 20 mA (incl. 1 x dust)
2 x 4 ... 20 mA (galvanically isolated)
digital signals
3 potential-free contacts (failure, limit value 1 and 2)
5 potential-free contacts
3 potential-free outputs (failure/maint., limit value 1 and 2)
power supply
110 VAC, 230 VAC or 24 VDC
400 VAC, 50 Hz, 3~
110 VAC, 230 VAC or 24 VDC
TUV approval
TI Air
TI Air, 13th, 17th and 27th BImSchV
TI Air, 13th, 17th and 27th BImSchV
analogue signals
DE 100 22 391 C2 | PCT/DE 01/01674
FMD 02 to combine with PFM 02 V for dust concentration measurement
Dust measuring devices
PFM 97 and 97 W
PFM 92 C
PFM 92 K
Dust measuring device
Filter controller
Filter controller
Influence of flow The triboelectric signal depends on the mechanical and electrical properties of the dust. Apart from the dust concentration it seems that the gas velocity has the biggest influence on the triboelectric charge transfer. There is the following mathematical relation:
Triboelektrisches Signal [Volt]
ci.B. cal A, D v Exp
= dust concentration = measuring signal = constants = flow of the exhaust gas = exponent (constant or function)
The standard one For the continuous dust c o n c e n t r a t i o n measurement
10 Staubart: Micro-Calcilin (3,5 µm)
8 6 4 2 0 0
20
40
60
80
100
120
Staubkonzentration [mg/m³]
In case of drops or aerosols in the exhaust gas the dust has to be measured extractively (e.g. by PFM 97 ED). Conventionally the measurement is done in-situ (e.g. by PFM 97 and 97 W or PFM 02 V).
Technical data control unit dimensions and degree of protection (control unit) probe dimensions and degree of protection (probe) measuring components measuring ranges calibration display media temperature ambient temperature dew point difference flow velocity
Application: monitoring of dust 1 emissions in all industries with dry exhaust gas, especially behind filtering precipitators Design: PFM 97 1 PFM 97 W 1 (17th BImSchV)
The compact one Advantages: combined 1
device (measures additionally gas temperature and velocity or flow) measuring ranges 1 adjustable display of measuring 1 values on LCD display at the control unit operating and status 1 displays on LCD display dust signal in operating 1 and norm state simple assembly without 1 special tools
For the continuous monitoring of filter plants Application: 1 monitoring of filtering precipitators (bag filter, cartridge filter) 1 monitoring of centrifugal separators (cyclones) design of complex 1 monitoring systems
The mobile one Advantages: 4 measuring ranges 1 adjustable compact design (no 1 separate control unit necessary) Effective filter diagnosis 1 (localisation of defective filter elements; determination of damaged membrane valves, not tightly shutting valves ...) excellent price1 perfomance ratio
For the mobile monitoring of several filters Application: monitoring 1
of filtering precipitators, especially pulse-jet-filters
Technical data
Technical data
Technical data
weather-proof aluminium case
no control unit necessary
portable aluminium case
305 x 240 x 300 mm (H x W x D), IP 65
-
370 x 490 x 215 mm (H x W x D), IP 20
combined stainless steel probe with GRP weather protection
stainless steel probe rod at probe head
stainless steel probe rod with pre-amplifier in the probe head
300 x 400 x 1000 mm (W x H x D), IP 65
300 mm (standard), IP 65
30 ... 500 mm (variable probe length), IP 65
dust, gas temperature and flow
dust: measurement with 1 triboelectric sensor
dust: measurement with 1 triboelectric sensor
0 ... 15 mg/m³ (max. 500 mg/m³)
0,3 ... 1.000 mg/m³ (special measuring ranges on request)
0,3 ... 1.000 mg/m³ (special measuring ranges on request)
by gravimetric reference measurements
by gravimetric reference measurements
by gravimetric reference measurements
LCD display
none
LCD display (0 ... 100 %) and digital recorder
max. 280 °C
max. 280 °C
max. 280 °C
-20 ... +50 °C
-20 ... +50 °C
-20 ... +50 °C
min. +5 K
min. +5 K
min. +5 K
from ca. 3 m/s
from ca. 3 m/s
from ca. 3 m/s
5 x 4 ... 20 mA (incl. 2 x dust, temp., flow)
1 x 4 ... 20 mA (dust)
1 x 4 ... 20 mA (dust)
digital signals
6 potential-free contacts
3 potential-free contacts (failure, limit value 1 and 2)
-
power supply
110 VAC, 230 VAC or 24 VDC
110 VAC, 230 VAC or 24 VDC
230 VAC
TUV approval
TI Air, 13th and 27th BImSchV (PFM 97 W: 17th BImSchV)
TI Air
TI Air
analogue signals
Patents
Advantages: system for use at different filters simple installation and 1 handling operating and status 1 displays via LED display of measuring 1 values on LCD and digital recorder storage of measuring 1 values in the recorder and on disc immediate identification 1 of filter defects via online line diagram portable 1
DE 197 29 144 C2 | EU 98112433.2-2204
Filter controller
PFM 02
PFM 02 Ex
PFM 02 C
Filter controller
Filter controller
Filter controller
Selection of measuring point In order to get representative results with regard to the dust concentration of plants, whose emissions have to be monitored continuously, the correct installation of the dust measuring devices is a basic pre-condition. This is also valid for the application of measuring devices within the frame of process monitoring (e.g. qualitative measurement - monitoring of the clean gas dust content of bag filters). The installation point of the probe should fulfil the requirements of local valid guidelines (e.g. EN 132841, VDI 2066 page 1). In case of doubts it is recommended to let determine the measuring point by a responsible measuring institute (in Germany according to §§ 26/28 BImSchG). We recommend to realise at least 5 times the diameter of the exhaust gas channel as entry and exit section.
Richtig!
VDI 2066 Bl. 1 EN 13284-1 5xD
The special one
For the monitoring of filter plants with increased comfort
For the monitoring of filter plants in Ex-zones
Application: monitoring of filtering 1 precipitators (bag filters, cartridge filters) monitoring of centrifugal 1 separators (cyclones) design of complex 1 monitoring systems
VDI 2066 Bl. 1 EN 13284-1 5xD
The comfortable one
Falsch! Richtig! <3xD
2xD
5xD
Advantages: device (no separate control unit necessary) operating and status 1 displays via LCD display of measuring 1 values on LCD fast identification of 1 filter defects by online line diagram display of dust content in 1 mg/m³ after calibration compact 1
Application: monitoring of filter 1 precipitators (bag filters and cartridge filters) monitoring of centrifugal 1 separators (cyclones) marking according to 1 guideline 94/9/EG: II 1/3D IP 65 T 74 °C class: probe rod 1 category 1 / electronic case category 3
The new compact one Advantages: device (no separate control unit necessary) operating and status 1 displays via LCD display of measuring 1 values on LCD fast identification of 1 filter defects by online line diagram display of dust content 1 in mg/m³ after calibration compact 1
For the continuous monitoring of filter plants Application: monitoring 1
of filtering precipitators (bag filters and cartridge filters) monitoring of centrifugal 1 separators (cyclones) design of complex 1 monitoring systems
Advantages: device (no separate control unit necessary) effective filter diagnosis 1 (localisation of defective filter elements; determination of damaged membrane valves, not tightly shutting valves ...) excellent price1 performance ratio compact 1
VDI 2066 Bl. 1 EN 13284-1
Technical data control unit dimensions and degree of protection (control unit) probe immersion depth and degree of protection (probe) measuring principle measuring range calibration display media temperature ambient temperature dew point difference flow velocity analogue signals digital signals
Technical data
Technical data
Technische Daten
control unit in probe head (no separate control unit)
control unit in the probe head (no separate control unit)
control unit in the probe head (no separate control unit)
-
-
-
stainless steel probe rod at the probe head
stainless steel probe rod at the probe head
stainless steel probe rod at the probe head
300 mm (standard), IP 65
300 mm (standard), IP 65
300 mm (Standard), IP 65
dust: measurement with 1 triboelectric sensor
dust: measurement with 1 triboelectric sensor
dust: measurement with 1 triboelectric sensor
0,1 ... 1.000 mg/m³ (special measuring ranges on request)
0,1 ... 1.000 mg/m³ (special measuring ranges on request)
0 ... 100 % (fix adjustment)
By gravimetric reference measurements
by gravimetric reference measurements
by gravimetric reference measurements
LCD display (meas. value in mg/m³ after gravim. calibration)
LCD display (meas. value in mg/m³ after gravim. calibration)
none
max. 280 °C
max. 250 °C
max. 280 °C
-20 ... +50 °C
-20 ... +50 °C
-20 ... +50 °C
min. +5 K
min. +5 K
min. +5 K
from ca. 3 m/s
from ca. 3 m/s
from ca. 3 m/s
1 x 4 ... 20 mA (dust) 3 potential-free outputs (failure, limit value 1 and 2)
1 x 4 ... 20 mA (dust)
3 potential-free outputs (failure, limit value 1 and 2)
1 x 4 ... 20 mA (dust)
3 potential-free outputs (failure/maint., limit value 1 and 2)
power supply
110 VAC, 230 VAC or 24 VDC
24 VDC
24 VDC
TUV approval
TI Air
TI Air
none
Ex-modules for voltage supply as well as analogue and digital signals available
limit value 1 at 60% of the measuring range limit value 2 at 80% of the measuring range
Filter controller
PFM 02
PFM 02 Ex
PFM 02 C
Filter controller
Filter controller
Filter controller
Selection of measuring point In order to get representative results with regard to the dust concentration of plants, whose emissions have to be monitored continuously, the correct installation of the dust measuring devices is a basic pre-condition. This is also valid for the application of measuring devices within the frame of process monitoring (e.g. qualitative measurement - monitoring of the clean gas dust content of bag filters). The installation point of the probe should fulfil the requirements of local valid guidelines (e.g. EN 132841, VDI 2066 page 1). In case of doubts it is recommended to let determine the measuring point by a responsible measuring institute (in Germany according to §§ 26/28 BImSchG). We recommend to realise at least 5 times the diameter of the exhaust gas channel as entry and exit section.
Richtig!
VDI 2066 Bl. 1 EN 13284-1 5xD
The special one
For the monitoring of filter plants with increased comfort
For the monitoring of filter plants in Ex-zones
Application: monitoring of filtering 1 precipitators (bag filters, cartridge filters) monitoring of centrifugal 1 separators (cyclones) design of complex 1 monitoring systems
VDI 2066 Bl. 1 EN 13284-1 5xD
The comfortable one
Falsch! Richtig! <3xD
2xD
5xD
Advantages: device (no separate control unit necessary) operating and status 1 displays via LCD display of measuring 1 values on LCD fast identification of 1 filter defects by online line diagram display of dust content in 1 mg/m³ after calibration compact 1
Application: monitoring of filter 1 precipitators (bag filters and cartridge filters) monitoring of centrifugal 1 separators (cyclones) marking according to 1 guideline 94/9/EG: II 1/3D IP 65 T 74 °C class: probe rod 1 category 1 / electronic case category 3
The new compact one Advantages: device (no separate control unit necessary) operating and status 1 displays via LCD display of measuring 1 values on LCD fast identification of 1 filter defects by online line diagram display of dust content 1 in mg/m³ after calibration compact 1
For the continuous monitoring of filter plants Application: monitoring 1
of filtering precipitators (bag filters and cartridge filters) monitoring of centrifugal 1 separators (cyclones) design of complex 1 monitoring systems
Advantages: device (no separate control unit necessary) effective filter diagnosis 1 (localisation of defective filter elements; determination of damaged membrane valves, not tightly shutting valves ...) excellent price1 performance ratio compact 1
VDI 2066 Bl. 1 EN 13284-1
Technical data control unit dimensions and degree of protection (control unit) probe immersion depth and degree of protection (probe) measuring principle measuring range calibration display media temperature ambient temperature dew point difference flow velocity analogue signals digital signals
Technical data
Technical data
Technische Daten
control unit in probe head (no separate control unit)
control unit in the probe head (no separate control unit)
control unit in the probe head (no separate control unit)
-
-
-
stainless steel probe rod at the probe head
stainless steel probe rod at the probe head
stainless steel probe rod at the probe head
300 mm (standard), IP 65
300 mm (standard), IP 65
300 mm (Standard), IP 65
dust: measurement with 1 triboelectric sensor
dust: measurement with 1 triboelectric sensor
dust: measurement with 1 triboelectric sensor
0,1 ... 1.000 mg/m³ (special measuring ranges on request)
0,1 ... 1.000 mg/m³ (special measuring ranges on request)
0 ... 100 % (fix adjustment)
By gravimetric reference measurements
by gravimetric reference measurements
by gravimetric reference measurements
LCD display (meas. value in mg/m³ after gravim. calibration)
LCD display (meas. value in mg/m³ after gravim. calibration)
none
max. 280 °C
max. 250 °C
max. 280 °C
-20 ... +50 °C
-20 ... +50 °C
-20 ... +50 °C
min. +5 K
min. +5 K
min. +5 K
from ca. 3 m/s
from ca. 3 m/s
from ca. 3 m/s
1 x 4 ... 20 mA (dust) 3 potential-free outputs (failure, limit value 1 and 2)
1 x 4 ... 20 mA (dust)
3 potential-free outputs (failure, limit value 1 and 2)
1 x 4 ... 20 mA (dust)
3 potential-free outputs (failure/maint., limit value 1 and 2)
power supply
110 VAC, 230 VAC or 24 VDC
24 VDC
24 VDC
TUV approval
TI Air
TI Air
none
Ex-modules for voltage supply as well as analogue and digital signals available
limit value 1 at 60% of the measuring range limit value 2 at 80% of the measuring range
Dust measuring devices
PFM 97 and 97 W
PFM 92 C
PFM 92 K
Dust measuring device
Filter controller
Filter controller
Influence of flow The triboelectric signal depends on the mechanical and electrical properties of the dust. Apart from the dust concentration it seems that the gas velocity has the biggest influence on the triboelectric charge transfer. There is the following mathematical relation:
Triboelektrisches Signal [Volt]
ci.B. cal A, D v Exp
= dust concentration = measuring signal = constants = flow of the exhaust gas = exponent (constant or function)
The standard one For the continuous dust c o n c e n t r a t i o n measurement
10 Staubart: Micro-Calcilin (3,5 µm)
8 6 4 2 0 0
20
40
60
80
100
120
Staubkonzentration [mg/m³]
In case of drops or aerosols in the exhaust gas the dust has to be measured extractively (e.g. by PFM 97 ED). Conventionally the measurement is done in-situ (e.g. by PFM 97 and 97 W or PFM 02 V).
Technical data control unit dimensions and degree of protection (control unit) probe dimensions and degree of protection (probe) measuring components measuring ranges calibration display media temperature ambient temperature dew point difference flow velocity
Application: monitoring of dust 1 emissions in all industries with dry exhaust gas, especially behind filtering precipitators Design: PFM 97 1 PFM 97 W 1 (17th BImSchV)
The compact one Advantages: combined 1
device (measures additionally gas temperature and velocity or flow) measuring ranges 1 adjustable display of measuring 1 values on LCD display at the control unit operating and status 1 displays on LCD display dust signal in operating 1 and norm state simple assembly without 1 special tools
For the continuous monitoring of filter plants Application: 1 monitoring of filtering precipitators (bag filter, cartridge filter) 1 monitoring of centrifugal separators (cyclones) design of complex 1 monitoring systems
The mobile one Advantages: 4 measuring ranges 1 adjustable compact design (no 1 separate control unit necessary) Effective filter diagnosis 1 (localisation of defective filter elements; determination of damaged membrane valves, not tightly shutting valves ...) excellent price1 perfomance ratio
For the mobile monitoring of several filters Application: monitoring 1
of filtering precipitators, especially pulse-jet-filters
Technical data
Technical data
Technical data
weather-proof aluminium case
no control unit necessary
portable aluminium case
305 x 240 x 300 mm (H x W x D), IP 65
-
370 x 490 x 215 mm (H x W x D), IP 20
combined stainless steel probe with GRP weather protection
stainless steel probe rod at probe head
stainless steel probe rod with pre-amplifier in the probe head
300 x 400 x 1000 mm (W x H x D), IP 65
300 mm (standard), IP 65
30 ... 500 mm (variable probe length), IP 65
dust, gas temperature and flow
dust: measurement with 1 triboelectric sensor
dust: measurement with 1 triboelectric sensor
0 ... 15 mg/m³ (max. 500 mg/m³)
0,3 ... 1.000 mg/m³ (special measuring ranges on request)
0,3 ... 1.000 mg/m³ (special measuring ranges on request)
by gravimetric reference measurements
by gravimetric reference measurements
by gravimetric reference measurements
LCD display
none
LCD display (0 ... 100 %) and digital recorder
max. 280 °C
max. 280 °C
max. 280 °C
-20 ... +50 °C
-20 ... +50 °C
-20 ... +50 °C
min. +5 K
min. +5 K
min. +5 K
from ca. 3 m/s
from ca. 3 m/s
from ca. 3 m/s
5 x 4 ... 20 mA (incl. 2 x dust, temp., flow)
1 x 4 ... 20 mA (dust)
1 x 4 ... 20 mA (dust)
digital signals
6 potential-free contacts
3 potential-free contacts (failure, limit value 1 and 2)
-
power supply
110 VAC, 230 VAC or 24 VDC
110 VAC, 230 VAC or 24 VDC
230 VAC
TUV approval
TI Air, 13th and 27th BImSchV (PFM 97 W: 17th BImSchV)
TI Air
TI Air
analogue signals
Patents
Advantages: system for use at different filters simple installation and 1 handling operating and status 1 displays via LED display of measuring 1 values on LCD and digital recorder storage of measuring 1 values in the recorder and on disc immediate identification 1 of filter defects via online line diagram portable 1
DE 197 29 144 C2 | EU 98112433.2-2204
Filter controller
PFM 92
PFM 97 ED
PFM 02 V
Filter controller
Dust measuring device
Dust measuring device
Measuring principle Function:
If 2 solid bodies get into contact by friction or touching, an electrical charge transfer will be caused. The arising charge difference is the base for triboelectric filter controllers and dust measuring devices which use the charge exchange between measuring probe and nearby streaming or direct impacting dust particles. Auswerteeinheit Staubsensor The triboelectric signal depends on the mechanical and electrical properties of the dust. Apart from the dust concentration it seems that the gas velocity has the biggest influence on the triboelectric charge transfer.
General advantages: In-situ online measurement of the dust content 1 High measuring accuracy 1 simple assembly without special tools 1 low demand for maintenance (long operating times) 1 no influence of probe pollution on the measuring 1 result no purging air required 1 low operating costs 1
Technical data control unit dimensions and degree of protection (control unit) probe immersion depth and degree of protection (probe) measuring principle
measuring range calibration display media temperature ambient temperature dew point difference flow velocity
The original one For the continuous monitoring of filter plants Application: monitoring of filtering 1 precipitators (bag filters, cartridge filters) monitoring of centrifugal 1 separators (cyclones) tendency measurement 1 of dust concentrations in raw gas
The extractive one Advantages: measuring ranges adjustable display of measuring 1 values on LCD display at the control unit Effective filter diagnosis 1 (localisation of defective filter ele-ments; determination of damaged membrane va l v e s , n o t t i g h t l y shutting valves ...) 4 1
For the continuous dust concentration measurement in wet and sticky gases Application: wet scrubber in waste 1 incinerations land reclamation plants 1 monitoring of exhaust 1 gases in woodprocessing industry (wood dryer, MDFdryer) application in insulant 1 material industry application behind wet 1 FGD
The delicate one Advantages: compact design 1 automatic dilution of the 1 gas flow sucked in self-monitoring of the 1 measuring gas flow automatic back-purging 1 for self-cleaning low demand for mainte1 nance
For the monitoring of filter plants with increased comfort Application: as dust 1
measuring system in combination with the signal of exhaust gas velocity (e.g. FMD 02) monitoring of dust 1 emissions in all industries with dry exhaust gas, especially behind filtering precipitators
Advantages: device (no separate control unit necessary) operating and status 1 displays via LCD display of measuring 1 values on LCD fast identification of 1 filter defects by on-line line diagram display of dust content 1 in mg/m³ after calibration compact 1
Technical data
Technical data
Technical data
weather proof aluminium case
steel-sheet case, mounted on profile frame (incl. blower)
control unit in the probe head (no separate control unit)
210 x 240 x 280 mm (H x W x D), IP 65
600 x 1700 x 500 mm (W x H x D), IP 65
-
stainless steel probe rod with pre-amplifier in the probe head
extractive sampling with GRP weather protection hood
stainless steel probe rod at the probe head
300 mm (standard), IP 65
500 x 750 x 1000 mm (W x H x D), IP 65
300 mm (Standard), IP 65
dust: measurement with 1 triboelectric sensor
dust and internal operating variables
dust: measurement with 1 triboelectric sensor
0,3 ... 1.000 mg/m³ (special measuring ranges on request)
0 ... 15 mg/m³ (max. 500 mg/m³)
0 ... 10 mg/m³ (max. 1.000 mg/m³)
by gravimetric reference measurements
by gravimetric reference measurements
by gravimetric reference measurements
LCD display (0 ... 100 %)
LCD display
LCD display (meas. value in mg/m³ after gravim. calibration)
max. 280 °C (higher temperatures on request)
max. 280 °C
max. 280 °C
-20 ... +50 °C
-20 ... +50 °C
-20 ... +50 °C
min. +5 K
independent from gas humidity
min. +5 K
from ca. 3 m/s
independent from gas flow
from ca. 3 m/s
1 x 4 ... 20 mA (dust)
5 x 4 ... 20 mA (incl. 1 x dust)
2 x 4 ... 20 mA (galvanically isolated)
digital signals
3 potential-free contacts (failure, limit value 1 and 2)
5 potential-free contacts
3 potential-free outputs (failure/maint., limit value 1 and 2)
power supply
110 VAC, 230 VAC or 24 VDC
400 VAC, 50 Hz, 3~
110 VAC, 230 VAC or 24 VDC
TUV approval
TI Air
TI Air, 13th, 17th and 27th BImSchV
TI Air, 13th, 17th and 27th BImSchV
analogue signals
DE 100 22 391 C2 | PCT/DE 01/01674
FMD 02 to combine with PFM 02 V for dust concentration measurement
Gravimetric dust measurement
GMD 06
PRODUCT INFORMATION 3
Gravimetric dust measuring device
Filter controller Dust measuring devices
Measuring principle The GMD 06 is an automatic sampling device for gravimetric dust measurements. These measurements are used as reference measuring method for the calibration of dust measuring devices (in Germany according to DIN VDI 2066). Furthermore it is possible to measure flow. The GMD 06 is able to determine independently all exhaust gas parameters being necessary for the dust measurement (e.g. humidity of the measuring gas, velocity in the stack as well as temperature and pressure). Under consideration of these parameters the GMD 06 extracts automatically a partial gas flow in isocinetic way from the exhaust gas channel. This partial gas flow is sucked via a filter, in which the dust contained is precipitated. A f t e r t h e measurement the filter is treated and weighed. When the filter mass weighed out has been entered manually, the GMD 06 calculates the dust content of the sample in operation and norm state.
Technical data control unit
The manual one For the gravimetric determination of the dust concentration Application: Measurement 1
of the absolute exhaust gas humidity Measurement of the 1 exhaust gas velocity (and D p, temperature) Measurement of the 1 exhaust gas content semi-automatic 1 calculation of the dust concentration after the determined dust mass has been entered
!Advantages:
Compact 1
system consisting of probes and control unit, easy use Variable application 1 possibilities due to probe adjustment to customer conditions On-site diagnosis of the 1 measuring values due to high resolution display 1 Excellent priceperformance ratio
Technical data portable case (control unit integrated) plus additional cases for accessories
probes
1 x humidity measurement, 1 x D p, temperature and flow
display
graphic display
humidity
0 ... 40 Vol%
dynamic pressure
0 ... 1.500 Pa
static pressure
-30 ... 10 kPa
flow (sampling) media temperature ambient temperature dew point difference flow velocity
10 ... 50 l/min max. 400 °C -20 ... +50 °C min. +5 K independent from gas flow
reaction time
< 8 sec
data memory
125 samples
power supply
230 VAC / 50 Hz, 200 W
TUV approval
none
Dr. Födisch Umweltmesstechnik AG
Function:
Dr. Födisch Umweltmesstechnik AG Zwenkauer Straße 159 D - 04420 Markranstädt Tel. Fax. Mail. Web.
+49-34205-755-0 +49-34205-755-40 vertrieb@foedisch.de http://www.foedisch.de
© Dr. Födisch Umweltmesstechnik AG - February 2008 subject to change